A preliminary study of the novel application of normoxic polymer gel dosimeters for the measurement of CTDI on diagnostic x-ray CT scanners

被引:46
|
作者
Hill, B
Venning, AJ
Baldock, C
机构
[1] Canberra Hosp, Biomed Engn Serv, Med Phys Sect, Woden, ACT 2606, Australia
[2] Univ Sydney, Inst Med Phys, Woden, ACT 2606, Australia
[3] Queensland Univ Technol, Sch Phys & Chem Sci, Brisbane, Qld 4001, Australia
[4] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
关键词
computer tomography; normoxic polymer gel dosimetry; dosimeter; CT; MRI; doseprofile;
D O I
10.1118/1.1925181
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Computer tomography dose index (CTDI) is a measurement undertaken during acceptance testing and subsequent quality assurance measurements of diagnostic x-ray CT scanners for the determination of patient dose. Normoxic polymer gel dosimeters have been used for the first time to measure dose and subsequently CTDI during acceptance testing of a CT scanner and compared with the conventional ionization chamber measurement for a range of imaging protocols. The normoxic polymer gel dosimeter was additionally used to simultaneously determine slice-width dose profiles and CTDI in the transaxial plane, the measurements of which are usually determined with thermoluminescent dosimetry or film. The resulting CTDI for all slice widths calculated from the normoxic polymer gel dosimeter were within corresponding ionization chamber CTDI values. Slice-width dose-profiles full-width half-maximum values from the normoxic polymer gel dosimeter were compared to the slice sensitivity profiles and were within the tolerances of the manufacturer. Normoxic polymer gel dosimeters have been shown to be a useful device for determining CTDI and dose distributions for CT equipment, and provide additional information not possible with just the use of an ionization chamber. (c) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:1589 / 1597
页数:9
相关论文
共 50 条
  • [21] A novel application of normoxic polymer gel dosimeters for near real-time dose measurement using cone-beam computed tomography
    Wu, Tung-Hsin
    Tsai, Chia-Jung
    Zhang, Geoffrey
    Yu, Chun-Yen
    Liang, Ji-An
    Wu, Jay
    Lee, Jason Js.
    Ho, Yung-Jen
    Huang, Tzung-Chi
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 633 : S279 - S281
  • [22] An x-ray CT polymer gel dosimetry prototype: II. Gel characterization and clinical application
    Johnston, H.
    Hilts, M.
    Carrick, J.
    Jirasek, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (10): : 3155 - 3175
  • [23] Cosolvent-free polymer gel dosimeters with improved dose sensitivity and resolution for x-ray CT dose response
    Chain, J. N. M.
    Jirasek, A.
    Schreiner, L. J.
    McAuley, K. B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (07): : 2091 - 2102
  • [24] Commissioning a Multislice CT Scanner for X-Ray CT Polymer Gel Dosimetry
    Johnston, H.
    Hilts, M.
    Jirasek, A.
    MEDICAL PHYSICS, 2014, 41 (06) : 238 - 238
  • [25] Characterization of amorphous thermoluminescence dosimeters for patient dose measurement in X-ray diagnostic procedures
    Ramli, N. N. H.
    Salleh, H.
    Mahdiraji, G. A.
    Zulkifli, M. I.
    Hashim, S.
    Bradley, D. A.
    Noor, N. M.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 116 : 130 - 134
  • [26] Image reconstruction for interrupted-beam x-ray CT on diagnostic clinical scanners
    Muckley, Matthew J.
    Chen, Baiyu
    Vahle, Thomas
    O'Donnell, Thomas
    Knoll, Florian
    Sodickson, Aaron D.
    Sodickson, Daniel K.
    Otazo, Ricardo
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (15):
  • [27] Reconstructing 3D x-ray CT images of polymer gel dosimeters using the zero-scan method
    Kakakhel, M. B.
    Kairn, T.
    Kenny, J.
    Trapp, J. V.
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [28] Technical considerations for implementation of X-ray CT polymer gel dosimetry
    Hilts, M
    Jirasek, A
    Duzenli, C
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (08): : 1727 - 1745
  • [29] Introduction of a deformable x-ray CT polymer gel dosimetry system
    Maynard, E.
    Heath, E.
    Hilts, M.
    Jirasek, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (07):
  • [30] Incorporating multislice imaging into x-ray CT polymer gel dosimetry
    Johnston, H.
    Hilts, M.
    Jirasek, A.
    MEDICAL PHYSICS, 2015, 42 (04) : 1666 - 1677