Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera

被引:50
作者
Duan, J [1 ]
Macey, DJ [1 ]
Pareek, PN [1 ]
Brezovich, IA [1 ]
机构
[1] Univ Alabama, Dept Radiat Oncol, Birmingham, AL 35233 USA
关键词
HDR brachytherapy; quality assurance; monitoring and verification; pinhole camera;
D O I
10.1118/1.1339882
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We investigated a pinhole imaging system for independent in vivo monitoring and verification of high dose rate (HDR) brachytherapy treatment. The system consists of a high-resolution pinhole collimator, an x-ray fluoroscope, and a standard radiographic screen-film combination. Autofluoroscopy provides real-time images of the in vivo Ir-192 HDR source for monitoring the source location and movement, whereas autoradiography generates a permanent record of source positions on film. Dual-pinhole autoradiographs render stereo-shifted source images that can be used to reconstruct the source dwell positions in three dimensions. The dynamic range and spatial resolution of the system were studied with a polystyrene phantom using a range of source strengths and dwell times. For the range of source activity used in HDR brachytherapy, a 0.5 mm diameter pinhole produced sharp fluoroscopic images of the source within the dynamic range of the fluoroscope. With a source-to-film distance of 35 cm and a 400 speed screen-film combination, the same pinhole yielded well recognizable images of a 281.2 GBq (7.60 Ci) Ir-192 source for dwell times in the typical clinical range of 2 to 400 s. This 0.5 mm diameter pinhole could clearly resolve source positions separated by lateral displacements as small as 1 mm. Using a simple reconstruction algorithm, dwell positions in a phantom were derived from stereo-shifted dual-pinhole images and compared to the known positions. The agreement was better than 1 mm. A preliminary study of a patient undergoing HDR treatment for cervical cancer suggests that the imaging method is clinically feasible. Based on these studies we believe that the pinhole imaging method is capable of providing independent and reliable real-time monitoring and verification for HDR brachytherapy. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:167 / 173
页数:7
相关论文
共 9 条
  • [1] AUTORADIOGRAPHY FOR I-125 SEEDS
    ALBERTI, W
    DIVOUX, S
    POTHMANN, B
    TABOR, P
    HERMANN, KP
    HARDER, D
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (05): : 881 - 884
  • [2] Anger H. O., 1967, INSTRUM NUCL, V1, P485
  • [3] QUANTITATIVE VERIFICATION OF IR-192 PDR AND HDR SOURCE STRUCTURE BY PINHOLE AUTORADIOGRAPHY
    KIROV, AS
    MEIGOONI, AS
    ZHU, Y
    VALICENTI, RK
    WILLIAMSON, JF
    [J]. MEDICAL PHYSICS, 1995, 22 (11) : 1753 - 1757
  • [4] High dose-rate brachytherapy treatment delivery: Report of the AAPM Radiation Therapy Committee Task Group No. 59
    Kubo, HD
    Glasgow, GP
    Pethel, TD
    Thomadsen, BR
    Williamson, JF
    [J]. MEDICAL PHYSICS, 1998, 25 (04) : 375 - 403
  • [5] A Monte Carlo study of verification imaging in high dose rate brachytherapy
    Sheikh-Bagheri, D
    Munro, P
    [J]. MEDICAL PHYSICS, 1998, 25 (04) : 404 - 414
  • [6] The effect of gamma ray penetration on angle-dependent sensitivity for pinhole collimation in nuclear medicine
    Smith, MF
    Jaszczak, RJ
    [J]. MEDICAL PHYSICS, 1997, 24 (11) : 1701 - 1709
  • [7] *US NUCL REG COMM, 1998, PNOIV98058 US NUCL R
  • [8] *US NUCL REG COMM, 1999, PNOIV99012 US NUCL R
  • [9] A system for nonradiographic source localization and real-time planning of intraoperative high dose rate brachytherapy
    Watanabe, Y
    Anderson, LL
    [J]. MEDICAL PHYSICS, 1997, 24 (12) : 2014 - 2023