Line-scanning fiber bundle endomicroscopy with a virtual detector slit

被引:30
|
作者
Hughes, Michael [1 ]
Yang, Guang-Zhong [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, Inst Global Hlth Innovat, Hamlyn Ctr Robot Surg, London SW7 2AZ, England
来源
BIOMEDICAL OPTICS EXPRESS | 2016年 / 7卷 / 06期
基金
英国工程与自然科学研究理事会;
关键词
HIGH-RESOLUTION MICROENDOSCOPY; CONFOCAL ENDOMICROSCOPY; LOW-COST; MICROSCOPY; ESOPHAGUS;
D O I
10.1364/BOE.7.002257
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Coherent fiber bundles can be used to relay the image plane from the distal tip of an endomicroscope to an external confocal microscopy system. The frame rate is therefore determined by the speed of the microscope's laser scanning system which, at 10-20 Hz, may be undesirably low for in vivo clinical applications. Line-scanning allows an increase in the frame rate by an order of magnitude in exchange for some loss of optical sectioning, but the width of the detector slit cannot easily be adapted to suit different imaging conditions. The rolling shutter of a CMOS camera can be used as a virtual detector slit for a bench-top line-scanning confocal microscope, and here we extend this idea to endomicroscopy. By synchronizing the camera rolling shutter with a scanning laser line we achieve confocal imaging with an electronically variable detector slit. This architecture allows us to acquire every other frame with the detector slit offset by a known distance, and we show that subtracting this second image leads to improved optical sectioning. (C) 2016 Optical Society of America
引用
收藏
页码:2257 / 2268
页数:12
相关论文
共 50 条
  • [1] Reduced motion artifacts and speed improvements in enhanced line-scanning fiber bundle endomicroscopy
    Thrapp, Andrew D.
    Hughes, Michael R.
    JOURNAL OF BIOMEDICAL OPTICS, 2021, 26 (05)
  • [2] High speed, line-scanning, fiber bundle fluorescence confocal endomicroscopy for improved mosaicking
    Hughes, Michael
    Yang, Guang-Zhong
    BIOMEDICAL OPTICS EXPRESS, 2015, 6 (04): : 1241 - 1251
  • [3] Line scanning, fiber bundle fluorescence HiLo endomicroscopy with confocal slit detection
    Zhang, Haojie
    Vyas, Khushi
    Yang, Guang-Zhong
    JOURNAL OF BIOMEDICAL OPTICS, 2019, 24 (11)
  • [4] Image scanning lensless fiber-bundle endomicroscopy
    Weinberg, Gil
    Weiss, Uri
    Katz, Ori
    OPTICS EXPRESS, 2023, 31 (22) : 37050 - 37057
  • [5] Fiber Bundle Fluorescence Endomicroscopy
    Tsai, Tsung-Han
    Zhou, Chao
    Fujimoto, James G.
    2010 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO) AND QUANTUM ELECTRONICS AND LASER SCIENCE CONFERENCE (QELS), 2010,
  • [6] Line-scanning confocal microscopic imaging based on virtual structured modulation
    Zhao Jia-wang
    Zhang Yun-hai
    Wang Fa-min
    Miao Xin
    Shi Xin
    CHINESE OPTICS, 2021, 14 (02) : 431 - 445
  • [7] Line-scanning laser ophthalmoscope
    Hammer, Daniel X.
    Ferguson, R. Daniel
    Ustun, Teoman E.
    Bigelow, Chad E.
    Iftimia, Nicusor V.
    Webb, Robert H.
    JOURNAL OF BIOMEDICAL OPTICS, 2006, 11 (04)
  • [8] LASER LINE-SCANNING SENSORS
    MATTHEWS, WF
    JUNG, RF
    OPTICAL ENGINEERING, 1975, 14 (02) : 116 - 119
  • [9] Line-scanning, stage scanning confocal microscope
    Carucci, John A.
    Stevenson, Mary
    Gareau, Daniel
    OPTICAL BIOPSY XIV: TOWARD REAL-TIME SPECTROSCOPIC IMAGING AND DIAGNOSIS, 2016, 9703
  • [10] STAGE-SCANNING LINE-SCANNING CONFOCAL MICROSCOPE
    Gareau, Daniel
    Meehan, Shane
    Carucci, John
    LASERS IN SURGERY AND MEDICINE, 2017, 49 (04) : 408 - 409