Contrast enhanced imaging with a stationary digital breast tomosynthesis system

被引:1
作者
Puett, Connor [1 ]
Calliste, Jabari [1 ]
Wu, Gongting [1 ]
Inscoe, Christina R. [1 ]
Lee, Yueh Z. [1 ]
Zhou, Otto [1 ]
Lu, Jianping [1 ]
机构
[1] Univ North Carolina Chapel Hill, Chapel Hill, NC 27599 USA
来源
MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING | 2017年 / 10132卷
关键词
digital breast tomosynthesis; DBT; contrast-enhanced breast imaging; CE-DBT; scatter correction; carbon nanotubes; CNT; SCATTER CORRECTION; RECONSTRUCTION; MAMMOGRAPHY;
D O I
10.1117/12.2254348
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Digital breast tomosynthesis (DBT) captures some depth information and thereby improves the conspicuity of breast lesions, compared to standard mammography. Using contrast during DBT may also help distinguish malignant from benign sites. However, adequate visualization of the low iodine signal requires a subtraction step to remove background signal and increase lesion contrast. Additionally, attention to factors that limit contrast, including scatter, noise, and artifact, are important during the image acquisition and post-acquisition processing steps. Stationary DBT (sDBT) is an emerging technology that offers a higher spatial and temporal resolution than conventional DBT. This phantom-based study explored contrast-enhanced sDBT (CE sDBT) across a range of clinically-appropriate iodine concentrations, lesion sizes, and breast thicknesses. The protocol included an effective scatter correction method and an iterative reconstruction technique that is unique to the sDBT system. The study demonstrated the ability of this CE sDBT system to collect projection images adequate for both temporal subtraction (TS) and dual-energy subtraction (DES). Additionally, the reconstruction approach preserved the improved contrast-to-noise ratio (CNR) achieved in the subtraction step. Finally, scatter correction increased the iodine signal and CNR of iodine-containing regions in projection views and reconstructed image slices during both TS and DES. These findings support the ongoing study of sDBT as a potentially useful tool for contrast-enhanced breast imaging and also highlight the significant effect that scatter has on image quality during DBT.
引用
收藏
页数:9
相关论文
共 23 条
[1]   Comparison of MRI and digital breast tomosynthesis in the preoperative evaluation of multifocal breast cancer [J].
Batohi, Bhavna ;
Vinci, Valeria ;
Peacock, Clare ;
Michell, Michael ;
Iqbal, Asif ;
Evans, David ;
Morel, Juliet ;
Satchithananda, Keshthra ;
Wasan, Reema ;
Rahim, Rumana .
BREAST CANCER RESEARCH, 2015, 17
[2]   A carbon nanotube field emission cathode with high current density and long-term stability [J].
Calderon-Colon, Xiomara ;
Geng, Huaizhi ;
Gao, Bo ;
An, Lei ;
Cao, Guohua ;
Zhou, Otto .
NANOTECHNOLOGY, 2009, 20 (32)
[3]  
Calliste J., 2015, P SOC PHOTO-OPT INS, V9412
[4]  
Calliste J., Med Phys
[5]   Dual-energy contrast-enhanced digital breast tomosynthesis - a feasibility study [J].
Carton, A-K ;
Gavenonis, S. C. ;
Currivan, J. A. ;
Conant, E. F. ;
Schnall, M. D. ;
Maidment, A. D. A. .
BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (988) :344-350
[6]   Screening for Dense Breasts: Digital Breast Tomosynthesis [J].
Destounis, Stamatia V. ;
Morgan, Renee ;
Arieno, Andrea .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (02) :261-264
[7]  
Hartman A. E., 2016, P SOC PHOTO-OPT INS, V9783
[8]  
Hill M. L., 2016, P IWDM, V9699, P645
[9]   Demonstration of a Scatter Correction Technique in Digital Breast Tomosynthesis [J].
Inscoe, Christy R. ;
Tucker, Andrew W. ;
Zhou, Otto Z. ;
Lu, Jianping .
MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
[10]   Contrast-enhanced digital mammography: Initial clinical experience [J].
Jong, RA ;
Yaffe, MJ ;
Skarpathiotakis, M ;
Shumak, RS ;
Danjoux, NM ;
Gunesekara, A ;
Plewes, DB .
RADIOLOGY, 2003, 228 (03) :842-850