An analytical approach to acceptance criteria for quality assurance of intensity modulated radiotherapy

被引:16
作者
Carver, Antony [1 ]
Gilmore, Martyn [1 ]
Riley, Stephen [1 ]
Uzan, Julien [1 ]
Mayles, Philip [1 ]
机构
[1] Clatterbridge Ctr Oncol, Dept Phys, Bebington CH64 1SS, Merseyside, England
关键词
IMRT quality assurance; PROSTATE-CANCER; IMRT;
D O I
10.1016/j.radonc.2011.08.045
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Detectability and impact of potential treatment machine errors on IMRT treatments were evaluated. The ability of the gamma index to detect deliberately introduced errors was assessed and their clinical impact was assessed using Tumour Control Probability (TCP) and Normal Tissue Complication Probability. TCP was only marginally affected by 2 mm errors in MLC position. Dose delivery errors had greater impact but were not detected as effectively using the gamma index. Acceptance criteria should include mean dose as well as gamma to help identify errors in the delivered dose. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 100 (2011) 453-455
引用
收藏
页码:453 / 455
页数:3
相关论文
共 20 条
[1]  
[Anonymous], 2009, J APPL CLIN MED PHYS
[2]   Escalated-dose versus standard-dose conformal radiotherapy in prostate cancer: first results from the MRC RT01 randomised controlled trial [J].
Dearnaley, David P. ;
Sydes, Matthew R. ;
Graham, John D. ;
Aird, Edwin G. ;
Bottomley, David ;
Cowan, Richard A. ;
Huddart, Robert A. ;
Jose, Chakiath C. ;
Matthews, John H. L. ;
Millar, Jeremy ;
Moore, A. Rollo ;
Morgan, Rachel C. ;
Russell, J. Martin ;
Scrase, Christopher D. ;
Stephens, Richard J. ;
Syndikus, Isabel ;
Parmar, Mahesh K. B. .
LANCET ONCOLOGY, 2007, 8 (06) :475-487
[3]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[4]   Evaluation of a biplanar diode array dosimeter for quality assurance of step-and-shoot IMRT [J].
Feygelman, Vladimir ;
Forster, Kenneth ;
Opp, Daniel ;
Nilsson, Gorgen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (04) :64-78
[5]   Dose-volume effects for normal tissues in external radiotherapy: Pelvis [J].
Fiorino, Claudio ;
Valdagni, Riccardo ;
Rancati, Tiziana ;
Sanguineti, Giuseppe .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) :153-167
[6]  
Fredh A, 2010, RADIOTHER ONCOL, V96, pS73
[7]  
James H, 2008, 96 IPEM
[8]   On the insensitivity of single field planar dosimetry to IMRT inaccuracies [J].
Kruse, Jon J. .
MEDICAL PHYSICS, 2010, 37 (06) :2516-2524
[9]   HISTOGRAM REDUCTION METHOD FOR CALCULATING COMPLICATION PROBABILITIES FOR 3-DIMENSIONAL TREATMENT PLANNING EVALUATIONS [J].
KUTCHER, GJ ;
BURMAN, C ;
BREWSTER, L ;
GOITEIN, M ;
MOHAN, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :137-146
[10]   Evaluation of the gamma dose distribution comparison method [J].
Low, DA ;
Dempsey, JF .
MEDICAL PHYSICS, 2003, 30 (09) :2455-2464