摘要
While aortic valve root compliance and leaflet coaptation have significant influence on valve closure, their implications have not yet been fully evaluated. The present study developed a full fluid-structure interaction (FSI) model that is able to cope with arbitrary coaptation between the leaflets of the aortic valve during the closing phase. Two simplifications were also evaluated for the simulation of the closing phase only. One employs an FSI model with a rigid root and the other uses a "dry" (without flow) model. Numerical tests were performed to verify the model. New metrics were defined to process the results in terms of leaflet coaptation area and contact pressure. The axial displacement of the leaflets, closure time and coaptation parameters were similar in the two FSI models, whereas the dry model, with imposed uniform load on the leaflets, produced larger coaptation area and contact pressure, larger axial displacement and faster closure time compared with the FSI model. The differences were up to 30% in the coaptation area, 55% in the contact pressure and 170% in the closure time. Consequently, an FSI model should be used to accurately resolve the kinematics of the aortic valve and leaflet coaptation details during the end-closing stage.
引用
收藏
页码:173 / 182
页数:10
相关论文
共 31 条
  • [1] Fluid-structure interaction and multi-body contact: Application to aortic valves
    Astorino, Matteo
    Gerbeau, Jean-Frederic
    Pantz, Olivier
    Traore, Karim-Frederic
    [J]. COMPUTER METHODS IN APPLIED MECHANICS AND ENGINEERING, 2009, 198 (45-46) : 3603 - 3612
  • [2] An approach to the simulation of fluid-structure interaction in the aortic valve
    Carmody, CJ
    Burriesci, G
    Howard, IC
    Patterson, EA
    [J]. JOURNAL OF BIOMECHANICS, 2006, 39 (01) : 158 - 169
  • [3] Chandran Krishnan B, 2010, Cardiovasc Eng Technol, V1, P18
  • [4] Dynamic finite element analysis of the aortic root from MRI-derived parameters
    Conti, Carlo A.
    Votta, Emiliano
    Della Corte, Alessandro
    Del Viscovo, Luca
    Bancone, Ciro
    Cotrufo, Maurizio
    Redaelli, Alberto
    [J]. MEDICAL ENGINEERING & PHYSICS, 2010, 32 (02) : 212 - 221
  • [5] A computational fluid-structure interaction analysis of a fiber-reinforced stentless aortic valve
    De Hart, J
    Baaijens, FPT
    Peters, GWM
    Schreurs, PJG
    [J]. JOURNAL OF BIOMECHANICS, 2003, 36 (05) : 699 - 712
  • [6] Opening and closing characteristics of the aortic valve after valve-sparing procedures using a new aortic root conduit
    De Paulis, R
    De Matteis, GM
    Nardi, P
    Scaffa, R
    Buratta, MM
    Chiariello, L
    [J]. ANNALS OF THORACIC SURGERY, 2001, 72 (02) : 487 - 494
  • [7] Simulation of fluid structure interaction with the interface artificial compressibility method
    Degroote, Joris
    Swillens, Abigail
    Bruggeman, Peter
    Haelterman, Robby
    Segers, Patrick
    Vierendeels, Jan
    [J]. INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2010, 26 (3-4) : 276 - 289
  • [8] Dumont K., 2004, Computer Methods in Biomechanics and Biomedical Engineering, V7, P139, DOI 10.1080/10255840410001715222
  • [9] Stress variations in the human aortic root and valve: The role of anatomic asymmetry
    Grande, KJ
    Cochran, RP
    Reinhall, PG
    Kunzelman, KS
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 1998, 26 (04) : 534 - 545
  • [10] Re-creation of sinuses is important for sparing the aortic valve: A finite element study
    Grande-Allen, KJ
    Cochran, RP
    Reinhall, PG
    Kunzelman, KS
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 119 (04) : 753 - 762