Long-term safety and absorption assessment of a novel bioresorbable nitrided iron scaffold in porcine coronary artery

被引:32
作者
Zheng, Jian-Feng [1 ]
Xi, Zi-Wei [1 ]
Li, Yang [1 ]
Li, Jia-Nan [7 ]
Qiu, Hong [1 ]
Hu, Xiao-Ying [1 ]
Luo, Tong [1 ]
Wu, Chao [1 ]
Wang, Xin [2 ]
Song, Lai-Feng [3 ]
Li, Li [3 ]
Qi, Hai-Ping [4 ]
Zhang, Gui [4 ]
Qin, Li [4 ]
Zhang, Wan-Qian [4 ]
Shi, Xiao-Li [5 ]
Wang, Shu-Han [6 ]
Zhang, De-Yuan [4 ]
Xu, Bo [1 ]
Gao, Run-Lin [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Fuwai Hosp, Natl Ctr Cardiovasc Dis, Dept Cardiol, Beijing, Peoples R China
[2] Chinese Acad Med Sci & Peking Union Med Coll, Fuwai Hosp, Beijing Key Lab Preclin Res & Evaluat Cardiovasc, Natl Ctr Cardiovasc Dis,Anim Expt Ctr, Beijing, Peoples R China
[3] Chinese Acad Med Sci & Peking Union Med Coll, Fuwai Hosp, Natl Ctr Cardiovasc Dis, Dept Pathol, Beijing, Peoples R China
[4] Biotyx Med Shenzhen Co Ltd, R&D Ctr, Shenzhen, Peoples R China
[5] Lifetech Sci Shenzhen Co Ltd, Natl & Local Joint Engn Lab Intervent Med Biotech, Shenzhen, Peoples R China
[6] Shen Zhen Testing Ctr Med Devices, Shenzhen, Peoples R China
[7] Capital Med Univ, Beijing Tiantan Hosp, Dept Cardiol & Macrovasc Dis, Beijing, Peoples R China
基金
国家重点研发计划;
关键词
Bioresorbable scaffold; Nitrided iron bioresorbable coronary scaffold; Preclinical study; Completely bioresorbable; ABSORBABLE METAL SCAFFOLD; SUSTAINED SAFETY; BIOSOLVE-II; STENT; BIOCOMPATIBILITY; PERFORMANCE; THROMBOSIS; OUTCOMES;
D O I
10.1016/j.bioactmat.2022.01.005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aimed to investigate the long-term biocompatibility, safety, and degradation of the ultrathin nitrided iron bioresorbable scaffold (BRS) in vivo, encompassing the whole process of bioresorption in porcine coronary arteries. Fifty-two nitrided iron scaffolds (strut thickness of 70 mu m) and 28 Vision Co-Cr stents were randomly implanted into coronary arteries of healthy mini-swine. The efficacy and safety of the nitrided iron scaffold were comparable with those of the Vision stentwithin 52 weeks after implantation. In addition, the long-term biocompatibility, safety, and bioresorption of the nitrided iron scaffold were evaluated by coronary angiography, optical coherence tomography, micro-computed tomography, scanning electron microscopy, energy dispersive spectrometry and histopathological evaluations at 4, 12, 26, 52 weeks and even at 7 years after implantation. In particular, a large number of struts were almost completely absorbed in situ at 7 years follow-up, which were first illustrated in this study. The lymphatic drainage pathway might serve as the potential clearance way of iron and its corrosion products.
引用
收藏
页码:496 / 505
页数:10
相关论文
共 32 条
  • [11] Lide D.R., 2003, HDB CHEM PHYS, V84th, P688
  • [12] In vivo degradation and endothelialization of an iron bioresorbable scaffold
    Lin, Wenjiao
    Zhang, Hongjie
    Zhang, Wanqian
    Qi, Haiping
    Zhang, Gui
    Qian, Jie
    Li, Xin
    Qin, Li
    Li, Haifeng
    Wang, Xiang
    Qiu, Hong
    Shi, Xiaoli
    Zheng, Wei
    Zhang, Deyuan
    Gao, Runlin
    Ding, Jiandong
    [J]. BIOACTIVE MATERIALS, 2021, 6 (04) : 1028 - 1039
  • [13] Long-term in vivo corrosion behavior, biocompatibility and bioresorption mechanism of a bioresorbable nitrided iron scaffold
    Lin, Wenjiao
    Qin, Li
    Qi, Haiping
    Zhang, Deyuan
    Zhang, Gui
    Gao, Runlin
    Qiu, Hong
    Xia, Ying
    Cao, Ping
    Wang, Xiang
    Zheng, Wei
    [J]. ACTA BIOMATERIALIA, 2017, 54 : 454 - 468
  • [14] Cytotoxicity and its test methodology for a bioabsorbable nitrided iron stent
    Lin, Wenjiao
    Zhang, Gui
    Cao, Ping
    Zhang, Deyuan
    Zheng, Yufeng
    Wu, Rangxiu
    Qin, Li
    Wang, Geqi
    Wen, Taoyuan
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2015, 103 (04) : 764 - 776
  • [15] Initial findings of impact of strut width on stent coverage and apposition of sirolimus-eluting stents assessed by optical coherence tomography
    Nakatani, Shimpei
    Nishino, Masami
    Taniike, Masayuki
    Makino, Nobuhiko
    Kato, Hiroyasu
    Egami, Yasuyuki
    Shutta, Ryu
    Tanouchi, Jun
    Yamada, Yoshio
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (05) : 776 - 781
  • [16] Coronary Responses and Differential Mechanisms of Late Stent Thrombosis Attributed to First-Generation Sirolimus- and Paclitaxel-Eluting Stents
    Nakazawa, Gaku
    Finn, Aloke V.
    Vorpahl, Marc
    Ladich, Elena R.
    Kolodgie, Frank D.
    Virmani, Renu
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (04) : 390 - 398
  • [17] The Evolving Role of Inflammatory Biomarkers in Risk Assessment After Stent Implantation
    Niccoli, Giampaolo
    Montone, Rocco A.
    Ferrante, Giuseppe
    Crea, Filippo
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 56 (22) : 1783 - 1793
  • [18] Impact of Locally Administered Carboxydextran-Coated Super-Paramagnetic Iron Nanoparticles on Cellular Immune Function
    Pedro, Luisa
    Harmer, Quentin
    Mayes, Eric
    Shields, Jacqueline D.
    [J]. SMALL, 2019, 15 (20)
  • [19] A novel approach to temporary stenting:: degradable cardiovascular stents produced from corrodible metal -: results 6-18 months after implantation into New Zealand white rabbits
    Peuster, M
    Wohlsein, P
    Brügmann, M
    Ehlerding, M
    Seidler, K
    Fink, C
    Brauer, H
    Fischer, A
    Hausdorf, G
    [J]. HEART, 2001, 86 (05) : 563 - 569
  • [20] Long-term biocompatibility of a corrodible peripheral iron stent in the porcine descending aorta
    Peuster, Matthias
    Hesse, Carola
    Schloo, Tirza
    Fink, Christoph
    Beerbaum, Philipp
    von Schnakenburg, Christian
    [J]. BIOMATERIALS, 2006, 27 (28) : 4955 - 4962