Wide energy range efficiency calibration for a lanthanum bromide scintillation detector

被引:18
作者
Favalli, A. [1 ]
Mehner, H. -C. [1 ]
Simonelli, F. [2 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Inst Protect & Secur Citizen, I-21020 Ispra, VA, Italy
[2] Commiss European Communities, Joint Res Ctr, Inst Hlth & Consumer Protect, I-21020 Ispra, VA, Italy
关键词
lanthanum bromide scintillation detector; efficiency;
D O I
10.1016/j.radmeas.2007.11.062
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In recent years a new type of scintillation detector based on the LaBr3:Ce crystal has been developed. This detector offers a substantial advantage over the standard Nal(Tl) detector. In comparison the energy resolution (about 3% at 662keV) is better and the efficiency for high-energy gamma rays is higher. The detector crystal has other advantages such as a high scintillation light output with a fast decay time. Due to the favorable characteristics and ease of use, the lanthanum bromide detector is likely to gain acceptance quickly in a variety of applications. We use the detector for activation analysis in conjunction with a 14-MeV neutron generator. This includes the detection of prompt gamma rays from neutron inelastic scattering and neutron capture. The elements of our interest include oxygen, hydrogen, nitrogen, carbon, fluorine and chlorine which emit characteristic gamma energies in the range of 1-10MeV. The paper presents the absolute full-energy peak efficiency calibration of a 38.1 mm x 38.1 mm (1.5 in x 1.5 in) LaBr3:Ce crystal for gamma-ray energies up to 5 MeV. For the present calibration the detector/source distance was set to 250 mm. (C) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:506 / 509
页数:4
相关论文
共 5 条
[1]   66Ga:: a standard for high-energy calibration of Ge detectors [J].
Baglin, CM ;
Browne, E ;
Norman, EB ;
Molnár, GL ;
Belgya, T ;
Révay, Z ;
Szelecsényi, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 481 (1-3) :365-377
[2]   Design and characterisation of a pulsed neutron interrogation facility [J].
Favalli, A. ;
Pedersen, B. .
RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) :74-77
[3]  
FAVALLI A, 2007, 29 ESARDA ANN M AIX
[4]  
Milbrath BD, 2005, IEEE NUCL SCI CONF R, P283
[5]  
ZIEGLER JF, 2002, STOPPING POWER RANGE