Comparison of 3D and 2D Breast Density Estimation from Synthetic Ultrasound Tomography Images and Digital Mammograms of Anthropomorphic Software Breast Phantoms

被引:2
作者
Bakic, Predrag R. [1 ]
Li, Cuiping [2 ]
West, Erik [2 ]
Sak, Mark [2 ]
Gavenonis, Sara C. [1 ]
Duric, Nebojsa [2 ]
Maidment, Andrew D. A. [1 ]
机构
[1] Univ Penn, Dept Radiol, 3400 Spruce St, Philadelphia, PA 19104 USA
[2] Wayne State Univ, Karmanos Canc Inst, Detroit, MI 48201 USA
来源
MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING | 2011年 / 7961卷
关键词
Digital mammography; ultrasound tomography; anthropomorphic breast phantom; breast density analysis; validation; SIMULATION; MODEL;
D O I
10.1117/12.878932
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Breast density descriptors were estimated from ultrasound tomography (UST) and digital mammogram (DM) images of 46 anthropomorphic software breast phantoms. Each phantom simulated a 450 ml or 700 ml breast with volumetric percent density (PD) values between 10% and 50%. The UST based volumetric breast density (VBD) estimates were calculated by thresholding the reconstructed UST images. Percent density (PD) values from DM images were estimated interactively by a clinical breast radiologist using Cumulus software. Such obtained UST VBD and Cumulus PD estimates were compared with the ground truth VBD values available from phantoms. The UST VBD values showed a high correlation with the ground truth, as evidenced by the Pearson correlation coefficient of r=0.93. The Cumulus PD values also showed a high correlation with the ground truth (r=0.84), as well as with the UST VBD values (r=0.78). The consistency in measuring the UST VBD and Cumulus PD values was analyzed using the standard error of the estimation by linear regression (sigma(E)). The sigma(E) value for Cumulus PD was 1.5 times higher compared to the UST VBD (6.54 vs. 4.21). The sigma(E) calculated from two repeated Cumulus estimation sessions (sigma(E)=4.66) was comparable with the UST. Potential sources of the observed errors in density measurement are the use of global thresholding and (for Cumulus) the human observer variability. This preliminary study of simulated phantom UST images showed promise for non-invasive estimation of breast density.
引用
收藏
页数:11
相关论文
共 19 条
[1]   Simulation of digital mammogram acquisition [J].
Bakic, P ;
Brzakovic, D .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :866-877
[2]   Mammogram synthesis using a 3D simulation. 1. Breast tissue model and image acquisition simulation [J].
Bakic, PR ;
Albert, M ;
Brzakovic, D ;
Maidment, ADA .
MEDICAL PHYSICS, 2002, 29 (09) :2131-2139
[3]   Mammogram synthesis using a 3D simulation. II. Evaluation of synthetic mammogram texture [J].
Bakic, PR ;
Albert, M ;
Brzakovic, D ;
Maidment, ADA .
MEDICAL PHYSICS, 2002, 29 (09) :2140-2151
[4]  
Bakic PR, 2010, INT WORKSH DIG MAMM
[5]  
Bakic PR, 2009, RADIOLOGY
[6]  
Bakic PR, 2010, SPIE MED IMAGING
[7]   Development of a physical 3D anthropomorphic breast phantom [J].
Carton, Ann-Katherine ;
Bakic, Predrag ;
Ullberg, Christer ;
Derand, Helen ;
Maidment, Andrew D. A. .
MEDICAL PHYSICS, 2011, 38 (02) :891-896
[8]  
Conant EF, 2010, INT WORKSH DIG MAMM
[9]  
Duric N, 2010, SPIE MED IMAGING
[10]   Detection of breast cancer with ultrasound tomography: First results with the Computed Ultrasound Risk Evaluation (CURE) prototype [J].
Duric, Nebojsa ;
Littrup, Peter ;
Poulo, Lou ;
Babkin, Alex ;
Pevzner, Roman ;
Holsapple, Earle ;
Rama, Olsi ;
Glide, Carri .
MEDICAL PHYSICS, 2007, 34 (02) :773-785