Dose perturbations due to contrast medium and air in MammoSite® treatment:: An experimental and Monte Carlo study

被引:23
作者
Cheng, CW
Mitra, R
Li, XA
Das, IJ
机构
[1] Arizona Oncol Associates, Tucson, AZ 85712 USA
[2] Alton Ochsner Med Fdn & Ochsner Clin, New Orleans, LA 70121 USA
[3] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
[4] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
关键词
MammoSite; breast; dose perturbation; Monte Carlo simulation;
D O I
10.1118/1.1943827
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In the management of early breast cancer, a partial breast irradiation technique called MammoSite(R) (Proxima Therapeutic Inc., Alpharetta, GA) has been advocated in recent years. In MammoSite, a balloon implanted at the surgical cavity during tumor excision is filled with a radio-opaque solution, and radiation is delivered via a high dose rate brachytherapy source situated at the center of the balloon. Frequently air may be introduced during placement of the balloon and/or injection of the contrast solution into the balloon. The purpose of this work is to quantify as well as to understand dose perturbations due to the presence of a high-Z contrast medium and/or an air bubble with measurements and Monte Carlo calculations. In addition, the measured dose distribution is compared with that obtained from a commercial treatment planning system (Nucletron PLATO System). For a balloon diameter of 42 rum, the dose variation as a function of distance from the balloon surface is measured for various concentrations of a radio-opaque solution (in the range 5%-25% by volume) with a small volume parallel plate ion chamber and a micro-diode detector placed perpendicular to the balloon axis. Monte Carlo simulations are performed to provide a basic understanding of the interaction mechanism and the magnitude of dose perturbation at the interface near balloon surface. Our results show that the radio-opaque concentration produces dose perturbation up to 6%. The dose perturbation occurs mostly within the distances <1 mm from the balloon surface. The Plato system that does not include heterogeneity correction may be sufficient for dose planning at distances >= 10 rum from the balloon surface for the iodine concentrations used in the MammoSite procedures. The dose enhancement effect near the balloon surface (<1 mm) due to the higher iodine concentration is not correctly predicted by the Plato system. The dose near the balloon surface may be increased by 0.5% per cm(3) of air. Monte Carlo simulation suggests that the interface effect (enhanced dose near surface) is primarily due to Compton electrons of short range (<0.5 mm). For more accurate dosimetry in MammoSite delivery, the dose perturbation due to the presence of a radio-opaque contrast medium and air bubbles should be considered in a brachytherapy planning system. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:2279 / 2287
页数:9
相关论文
共 48 条
  • [1] Arthur Douglas W, 2002, Brachytherapy, V1, P184, DOI 10.1016/S1538-4721(02)00099-5
  • [2] Optimization of mammosite therapy
    Astrahan, MA
    Jozsef, G
    Streeter, OE
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 58 (01): : 220 - 232
  • [3] Belkacemi Y, 2003, Cancer Radiother, V7 Suppl 1, p129s
  • [4] Briesmeister J. F., 2000, LA13709 LOS AL NAT L
  • [5] Monte Carlo dose calculations in homogeneous media and at interfaces: A comparison between GEPTS, EGSnrc, MCNP, and measurements
    Chibani, O
    Li, XA
    [J]. MEDICAL PHYSICS, 2002, 29 (05) : 835 - 847
  • [6] BACKSCATTER DOSE PERTURBATION IN KILOVOLTAGE PHOTON BEAMS AT HIGH ATOMIC-NUMBER INTERFACES
    DAS, IJ
    CHOPRA, KL
    [J]. MEDICAL PHYSICS, 1995, 22 (06) : 767 - 773
  • [7] Forward dose perturbation at high atomic number interfaces in kilovoltage x-ray beams
    Das, IJ
    [J]. MEDICAL PHYSICS, 1997, 24 (11) : 1781 - 1787
  • [8] BACKSCATTER DOSE PERTURBATION AT HIGH ATOMIC-NUMBER INTERFACES IN MEGAVOLTAGE PHOTON BEAMS
    DAS, IJ
    KAHN, FM
    [J]. MEDICAL PHYSICS, 1989, 16 (03) : 367 - 375
  • [9] Dose perturbations at high-Z interfaces in kilovoltage photon beams:: comparison with Monte Carlo simulations and measurements
    Das, IJ
    Moskvin, VP
    Kassaee, A
    Tabata, T
    Verhaegen, F
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2002, 64 (03) : 173 - 179
  • [10] Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source
    Daskalov, GM
    Loffler, E
    Williamson, JF
    [J]. MEDICAL PHYSICS, 1998, 25 (11) : 2200 - 2208