Interlaboratory Validation for a Real-Time PCR Salmonella Detection Method Using the ABI 7500 FAST Real-Time PCR System

被引:17
作者
Cheng, Chorng-Ming [1 ]
Doran, Tara [2 ]
Lin, Wen [3 ]
Chen, Kai-Shun [1 ]
Williams-Hill, Donna [1 ]
Pamboukian, Ruiqing [3 ]
机构
[1] US FDA, Off Regulatory Affairs, Pacific Reg Lab Southwest, Irvine, CA 92612 USA
[2] US FDA, Ctr Food Safety & Appl Nutr, College Pk, MD 20740 USA
[3] US FDA, Off Regulatory Affairs, Off Regulatory Sci, Rockville, MD 20857 USA
关键词
POLYMERASE-CHAIN-REACTION; GENE; AMPLIFICATION; TYPHIMURIUM; FOODS;
D O I
10.4315/0362-028X.JFP-14-244
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Sixteen FERN (Food Emergency Response Network) member laboratories collaborated in this study to verify extension of the real-time PCR Salmonella detection method originally designed for the single-tube Cepheid SmartCycler II and validated against the Salmonella method of the U.S. Food and Drug Administration Bacteriological Analytical Manual to the Applied Biosystems (ABI) 7500 FAST Real-Time PCR system multiwell plate platform. Four foods were selected for this study: chili powder, soft cheese, fish, and tomatoes; these foods represent products that are commonly analyzed for the presence of Salmonella for regulatory purposes. Each food consisted of six uninoculated control samples, six samples inoculated with low Salmonella levels (target 1 to 5 CFU/25 g), and six samples inoculated with high levels (target 10 to 50 CFU/25 g). All samples were tested for Salmonella using the 24-h quantitative PCR (qPCR) method for detecting Salmonella, which utilizes modified buffered peptone water as the sole enrichment medium and an internal control for the qPCR. Each of these 18 samples was individually analyzed for Salmonella by the collaborating laboratories using both the ABI 7500 FAST system (alternative method) and the SmartCycler LE system (reference method). Statistical analysis of the data revealed no significant difference (P >= 0.05) between these two qPCR platforms except for the chili powder samples. The differences noted with chili powder (P = 0.0455) were attributed to the enhanced sensitivity of the ABI 7500 FAST system compared with the SmartCycler II system. The detection limit of both qPCR methods was 0.02 to 0.15 CFU/g. These results provide a solid basis for extending the 24-h qPCR Salmonella method to the ABI 7500 FAST system for high-throughput detection of Salmonella in foods:
引用
收藏
页码:1119 / 1124
页数:6
相关论文
共 50 条
  • [41] Real-time PCR, a method fit for detection and quantification of Erwinia amylovora
    Dreo, Tanja
    Pirc, Manca
    Ravnikar, Maja
    TREES-STRUCTURE AND FUNCTION, 2012, 26 (01): : 165 - 178
  • [42] A real-time PCR for the detection and characterisation of Aspergillus species
    Fricke, Stephan
    Fricke, Christian
    Oelkrug, Christopher
    Blatz, Rosemarie
    Schoenfelder, Uta
    Niederwieser, Dietger
    Hilger, Nadja
    Ruhnke, Markus
    Rodloff, Arne C.
    MYCOSES, 2012, 55 (05) : 416 - 425
  • [43] Citation Classic The Beginnings of Real-Time PCR
    Williams, P. Mickey
    CLINICAL CHEMISTRY, 2009, 55 (04) : 833 - 834
  • [44] Real-time reverse-transcriptase PCR for Salmonella Typhimurium detection from lettuce and tomatoes
    Miller, Nathan D.
    Davidson, P. M.
    D'Souza, Doris H.
    LWT-FOOD SCIENCE AND TECHNOLOGY, 2011, 44 (04) : 1088 - 1097
  • [45] Real-time PCR in Food Science: PCR Diagnostics
    Rodriguez-Lazaro, David
    Cook, Nigel
    Hernandez, Marta
    CURRENT ISSUES IN MOLECULAR BIOLOGY, 2013, 15 (02) : 39 - 43
  • [46] Handheld real-time PCR device
    Ahrberg, Christian D.
    Ilic, Bojan Robert
    Manz, Andreas
    Neuzil, Pavel
    LAB ON A CHIP, 2016, 16 (03) : 586 - 592
  • [47] IN HOUSE VALIDATION OF REAL-TIME PCR METHOD FOR DETECTION OF CLOSTRIDIUM BOTULINUM IN FOOD AND FEED MATRIXES
    Grenda, Tomasz
    Kwiatek, Krzysztof
    BULLETIN OF THE VETERINARY INSTITUTE IN PULAWY, 2010, 54 (04) : 557 - 562
  • [48] Development and validation of a real-time PCR assay for the detection of anguillid herpesvirus 1
    van Beurden, S. J.
    Voorbergen-Laarman, M. A.
    Roozenburg, I.
    van Tellingen, J.
    Haenen, O. L. M.
    Engelsma, M. Y.
    JOURNAL OF FISH DISEASES, 2016, 39 (01) : 95 - 104
  • [49] Real-time PCR method for detection of short DNA using a deoxyuridine probe and application for detection of fomivirsen
    Harikai, Naoki
    Tanaka, Yuko
    Miyashita, Satoshi
    Zaima, Kazumasa
    Shinomiya, Kazufusa
    BIOTECHNIQUES, 2022, 73 (06) : 281 - 287
  • [50] Real-time PCR assay for detection of Trichinella in meat
    Alonso, Mercedes
    Herrero, Beatriz
    Vieites, Juan M.
    Espineira, Montserrat
    FOOD CONTROL, 2011, 22 (08) : 1333 - 1338