Comparison of 2D and 3D Gamma calculations for an IMRT QA phantom

被引:5
|
作者
Lafratta, R. [1 ,2 ]
Ibbott, G. [1 ,2 ]
Adamovics, J. [3 ,4 ]
Followill, D. [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas Houston, Grad Sch Biomed Sci, Houston, TX 77030 USA
[3] Rider Univ, Lawerenceville, NJ USA
[4] Heuris Pharma LLC, Skillman, NJ 08558 USA
来源
8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE) | 2015年 / 573卷
关键词
RADIATION; DOSIMETRY;
D O I
10.1088/1742-6596/573/1/012055
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Gamma index pass rates were evaluated for an anthropomorphic phantom using both 2D and 3D calculations. The phantom was irradiated with the traditional dosimetry insert loaded with radiochromic film and TLD, and then with a 3D dosimetry insert. A comparison with the calculated dose distribution showed that both the PRESAGE (R) dosimeter and the film and TLD system agreed with the plan to within 5% using 2D gamma index criteria. The 3D gamma index showed a slightly higher pass rate than the 2D gamma index at 3%/3mm, and comparable pass rates using more generous constraints. The acceptable number of pixels passing the tighter constraints (3%/3mm) might be dependent upon the choice between a 2D calculation versus a 3D calculation.
引用
收藏
页数:3
相关论文
共 50 条
  • [21] Characterization of a 2D ionization chamber array for IMRT QA
    Shtraus, N.
    Curran, B.
    Moran, J.
    Litzenberg, D.
    MEDICAL PHYSICS, 2007, 34 (06) : 2600 - 2600
  • [22] Feasibility and Sensitivity of Individual Volume Based 3D Gamma Indices for Composite IMRT QA
    Jin, X.
    Han, C.
    Xie, C.
    MEDICAL PHYSICS, 2017, 44 (06) : 2895 - 2895
  • [23] A simple 3D dose verification method for patient IMRT QA using measured 2D dose distributions and a projection algorithm
    Audet, C
    Poffenbarger, B
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S420 - S420
  • [24] FROM 2D TO 3D DOSE EVALUATION FOR IN VIVO IMRT VERIFICATION
    Camargo, P.
    Rozendaal, R.
    Joep, S.
    Sonke, J. J.
    van Herk, M.
    Mijnheer, B.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S214 - S214
  • [25] A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup
    Guan, Huaiqun
    Hammoud, Rabih
    Yin, Fang-Fang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (04): : 273 - 280
  • [26] 2D IMRT QA passing rate dependency on coronal plane
    Panetta, Joseph V.
    Veltchev, Iavor
    Price, Robert A.
    Ma, C-M Charlie
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 110
  • [27] IMRT QA via MapCHECK a 2D diode detector system
    Cullip, T
    Chang, S
    MEDICAL PHYSICS, 2003, 30 (06) : 1351 - 1351
  • [28] Investigation of biological phantom for 2D and 3D breast EIT images
    Qiao, Guofeng
    Wang, Wei
    Wang, Li
    He, Yi
    Brarner, B.
    Al-Akaidi, M.
    13TH INTERNATIONAL CONFERENCE ON ELECTRICAL BIOIMPEDANCE AND THE 8TH CONFERENCE ON ELECTRICAL IMPEDANCE TOMOGRAPHY 2007, 2007, 17 : 328 - +
  • [29] 2D & 3D Shepp-Logan phantom standards for MRI
    Gach, H. Michael
    Tanase, Costin
    Boada, Femando
    ICSENG 2008: INTERNATIONAL CONFERENCE ON SYSTEMS ENGINEERING, 2008, : 521 - +
  • [30] The influence of decreasing countrates on 2D and 3D acquisitions:: A phantom study
    Van Roy, M
    de Hemricourt, E
    Bartholomé, V
    Alberty, J
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S401 - S401