The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance

被引:23
作者
Espinoza, A. [1 ]
Petasecca, M. [1 ]
Fuduli, I. [1 ]
Howie, A. [2 ]
Bucci, J. [2 ]
Corde, S. [3 ]
Jackson, M. [3 ]
Lerch, M. L. F. [1 ]
Rosenfeld, A. B. [1 ]
机构
[1] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
[2] St George Hosp, Canc Care Ctr, Kogarah, NSW 2217, Australia
[3] Prince Wales Hosp, Dept Radiat Oncol, Randwick, NSW 2031, Australia
基金
澳大利亚研究理事会;
关键词
brachytherapy; quality assurance; magic plate; gamma index; diode array; VERIFICATION; DOSIMETRY; DETECTOR;
D O I
10.1118/1.4905233
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: High dose rate (HDR) brachytherapy is a treatment method that is used increasingly worldwide. The development of a sound quality assurance program for the verification of treatment deliveries can be challenging due to the high source activity utilized and the need for precise measurements of dwell positions and times. This paper describes the application of a novel phantom, based on a 2D 11x11 diode array detection system, named "magic phantom" (MPh), to accurately measure plan dwell positions and times, compare them directly to the treatment plan, determine errors in treatment delivery, and calculate absorbed dose. Methods: The magic phantom system was CT scanned and a 20 catheter plan was generated to simulate a nonspecific treatment scenario. This plan was delivered to the MPh and, using a custom developed software suite, the dwell positions and times were measured and compared to the plan. The original plan was also modified, with changes not disclosed to the primary authors, and measured again using the device and software to determine the modifications. A new metric, the "position-time gamma index," was developed to quantify the quality of a treatment delivery when compared to the treatment plan. The MPh was evaluated to determine the minimum measurable dwell time and step size. The incorporation of the TG-43U1 formalism directly into the software allows for dose calculations to be made based on the measured plan. The estimated dose distributions calculated by the software were compared to the treatment plan and to calibrated EBT3 film, using the 2D gamma analysis method. Results: For the original plan, the magic phantom system was capable of measuring all dwell points and dwell times and the majority were found to be within 0.93 mm and 0.25 s, respectively, from the plan. By measuring the altered plan and comparing it to the unmodified treatment plan, the use of the position-time gamma index showed that all modifications made could be readily detected. The MPh was able to measure dwell times down to 0.067 +/- 0.001 s and planned dwell positions separated by 1 mm. The dose calculation carried out by the MPh software was found to be in agreement with values calculated by the treatment planning system within 0.75%. Using the 2D gamma index, the dose map of the MPh plane and measured EBT3 were found to have a pass rate of over 95% when compared to the original plan. Conclusions: The application of this magic phantom quality assurance system to HDR brachytherapy has demonstrated promising ability to perform the verification of treatment plans, based upon the measured dwell positions and times. The introduction of the quantitative position-time gamma index allows for direct comparison of measured parameters against the plan and could be used prior to patient treatment to ensure accurate delivery. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:663 / 673
页数:11
相关论文
共 29 条
  • [1] Characterization of an Innovative p-type Epitaxial Diode for Dosimetry in Modern External Beam Radiotherapy
    Aldosari, A. H.
    Espinoza, A.
    Robinson, D.
    Fuduli, I.
    Porumb, C.
    Alshaikh, S.
    Carolan, M.
    Lerch, M. L. F.
    Perevertaylo, V.
    Rosenfeld, A. B.
    Petasecca, M.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (06) : 4705 - 4712
  • [2] [Anonymous], 2013, [No title captured], Patent No. 7972259
  • [3] [Anonymous], ONC MASTERPLAN V 3 2
  • [4] Batic M., 2009, NUCL SCI S NSS MLC 2
  • [5] Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams
    Brown, Thomas A. D.
    Hogstrom, Kenneth R.
    Alvarez, Diane
    Matthews, Kenneth L., II
    Ham, Kyungmin
    Dugas, Joseph P.
    [J]. MEDICAL PHYSICS, 2012, 39 (12) : 7412 - 7417
  • [6] Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera
    Duan, J
    Macey, DJ
    Pareek, PN
    Brezovich, IA
    [J]. MEDICAL PHYSICS, 2001, 28 (02) : 167 - 173
  • [7] The feasibility study and characterization of a two-dimensional diode array in "magic phantom" for high dose rate brachytherapy quality assurance
    Espinoza, A.
    Beeksma, B.
    Petasecca, M.
    Fuduli, I.
    Porumb, C.
    Cutajar, D.
    Corde, S.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    [J]. MEDICAL PHYSICS, 2013, 40 (11)
  • [8] Multichannel Data Acquisition System comparison for Quality Assurance in external beam radiation therapy
    Fuduli, I.
    Newall, M. K.
    Espinoza, A. A.
    Porumb, C. S.
    Carolan, M.
    Lerch, M. L. F.
    Metcalfe, P.
    Rosenfeld, A. B.
    Petasecca, M.
    [J]. RADIATION MEASUREMENTS, 2014, 71 : 338 - 341
  • [9] A comparative analysis of multichannel Data Acquisition Systems for quality assurance in external beam radiation therapy
    Fuduli, I.
    Porumb, C.
    Espinoza, A. A.
    Aldosari, A. H.
    Carolan, M.
    Lerch, M. L. F.
    Metcalfe, P.
    Rosenfeld, A. B.
    Petasecca, M.
    [J]. JOURNAL OF INSTRUMENTATION, 2014, 9
  • [10] A dosimetric study on the Ir-192 high dose rate Flexisource
    Granero, D.
    Perez-Calatayud, J.
    Casal, E.
    Ballester, F.
    Venselaar, J.
    [J]. MEDICAL PHYSICS, 2006, 33 (12) : 4578 - 4582