Determination of recombination and polarity correction factors, kS and kP, for small cylindrical ionization chambers PTW 31021 and PTW 31022 in pulsed filtered and unfiltered beams

被引:11
作者
Bruggmoser, Gregor [1 ]
Saum, Rainer [1 ]
Kranzer, Rafael [2 ]
机构
[1] Univ Klinikum Freiburg, Klin Strahlenheilkunde, Robert Koch Str 3, D-79106 Freiburg, Germany
[2] PTW Freiburg, Lorracher Str 7, D-79115 Freiburg, Germany
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2018年 / 28卷 / 03期
关键词
Ionization chamber; Saturation loss; Recombination; Polarity effect; HIGH-ENERGY PHOTON; ION RECOMBINATION; ELECTRON-BEAMS; PLANE-PARALLEL; DOSIMETRY;
D O I
10.1016/j.zemedi.2017.09.010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this technical communication is to provide correction factors for recombination and polarity effect for two new ionization chambers PTW PinPoint 3D (type 31022) and PTW Semiflex 3D (type 31021). The correction factors provided are for the (based on the) German DIN 6800-2 dosimetry protocol and the AAPM TG51 protocol. The measurements were made in filtered and unfiltered high-energy photon beams in a water equivalent phantom at maximum depth of the PDD and a field size on the surface of 10 cm x 10 cm. The design of the new chamber types leads to an ion collection efficiency and a polarity effect that are well within the specifications requested by pertinent dosimetry protocols including the addendum of TG-51. It was confirmed that the recombination effect of both chambers mainly depends on dose per pulse and is independent of the filtration of the photon beam.
引用
收藏
页码:247 / 253
页数:7
相关论文
共 19 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]  
[Anonymous], 2008, 6800 DIN
[3]   Recombination factors for the cylindrical FC65-G ionization chamber in pulsed photon beams and the plane-parallel Roos ionization chamber in pulsed electron beams [J].
Berg, M ;
Noerrevang, O .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (23) :5309-5318
[4]  
Boag JW., 1966, IONIZATION CHAMBERS, V2, P1
[5]   Determination of the recombination correction factor kS for some specific plane-parallel and cylindrical ionization chambers in pulsed photon and electron beams [J].
Bruggmoser, G. ;
Saum, R. ;
Schmachtenberg, A. ;
Schmid, F. ;
Schuele, E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (02) :N35-N50
[6]   Ion recombination corrections for the NACP parallel-plate chamber in a pulsed electron beam [J].
Burns, DT ;
McEwen, MR .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2033-2045
[7]   COMMENTS ON ION RECOMBINATION CORRECTIONS FOR PLANE-PARALLEL AND THIMBLE CHAMBERS IN ELECTRON AND PHOTON RADIATION [J].
BURNS, JE ;
BURNS, DT .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (12) :1986-1988
[8]   MEASUREMENT OF SATURATION CORRECTION FACTORS OF THIMBLE-TYPE IONIZATION CHAMBERS IN PULSED PHOTON BEAMS [J].
DERIKUM, K ;
ROOS, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) :755-763
[9]  
Derikum K., 2002, ALLGEMEINEN KORREKTI, P57
[10]  
ICRU, 1982, ICRU Report 34