Comparing biomechanical strength of unicortical locking plate versus bicortical compression plate for transverse midshaft metacarpal fracture

被引:3
作者
Shanmugam, Rukmanikanthan [1 ]
Jian, Chong Yin Chyi Chong Shyh [1 ]
Haseeb, Amber [1 ]
Aik, Saw [1 ]
机构
[1] Univ Malaya, Fac Med, Dept Orthopaed Surg NOCERAL, Kuala Lumpur 50603, Malaysia
关键词
compression plate; locking plate; metacarpal fracture; unicortical locking plate; FIXATION; COMPLICATIONS; PHALANGEAL; HAND; SCREWS;
D O I
10.1177/2309499018802511
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: Metacarpal bone fractures constitute 10% of all fractures. Unstable metacarpal fractures require surgical intervention, which poses danger to flexor tendon either due to bicortical drilling or construct of the implant. Unicortical locking plate fixation may be the solution to preventing flexor tendon injury. Studies have compared locking and compression plates. However, in these studies, the biomechanical properties were tested using the static loading method. This study looks into cyclical loading that is more representative of in vivo conditions, particularly for early rehabilitation. We compared the biomechanical strength of the unicortical locking plate and bicortical compression plate system in a transverse metacarpal fracture, tested with cyclical loading and torsion. Method: Twenty pieces of fourth-generation, biomechanical testing grade, left third metacarpal composite bones were used. Resin was used to create the holding block at both ends of the bone. An oscillating saw with 0.8 mm thick saw blade was used to osteotomize the metacarpal sawbones to create a midshaft transverse metacarpal fracture model. Ten pieces were fixed with a 2.0 mm titanium locking plate via unicortical screw purchase and 10 were fixed with a 2.0 mm, four holes, titanium dynamic compression plate, bicortical purchase of screws. They were subjected to cyclic load to failure testing three-point bending and torsion. Results: There were no significant difference in stiffness and cyclic three-point bending to failure between the unicortical locking plate group and the bicortical compression plate group. The bicortical compression plate group is stiffer and has a higher cyclic bending load to failure as compared to the unicortical locking plate group. Conclusion: Unicortical locking plate fixation of metacarpal fracture can be reliably applied clinically to produce a strong and stable construct that allows early mobilization of the joints. This will not only reduce the complication rate of metacarpal plating, but also improve the functional outcome of the hand.
引用
收藏
页数:10
相关论文
共 24 条
[1]   A biomechanical study comparing plate fixation using unicortical and bicortical screws in transverse metacarpal fracture models subjected to cyclic loading [J].
Afshar, R. ;
Fong, T. S. ;
Latifi, M. Hadi ;
Kanthan, S. R. ;
Kamarul, T. .
JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2012, 37E (05) :396-401
[2]   FORCES IN THE NORMAL AND ABNORMAL HAND [J].
AN, KN ;
CHAO, EY ;
COONEY, WP ;
LINSCHEID, RL .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1985, 3 (02) :202-211
[3]  
Barr Cameron, 2013, Hand (N Y), V8, P454, DOI 10.1007/s11552-013-9544-3
[4]   Biomechanical Analysis of the Efficacy of Locking Plates during Cyclic Loading in Metacarpal Fractures [J].
Doht, Stefanie ;
Meffert, Rainer H. ;
Raschke, Michael J. ;
Blunk, Torsten ;
Ochman, Sabine .
SCIENTIFIC WORLD JOURNAL, 2014,
[5]   Plating of metacarpal fractures: Unicortical or bicortical screws? [J].
Dona, E ;
Gillies, RM ;
Gianoutsos, MP ;
Walsh, WR .
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 2004, 29B (03) :218-221
[6]   TENDON-RUPTURE AS A COMPLICATION OF SCREW FIXATION IN FRACTURES IN THE HAND - CASE-REPORT [J].
FAMBROUGH, RA ;
GREEN, DP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1979, 61 (05) :781-782
[7]   COMPARATIVE FATIGUE STRENGTHS AND STABILITIES OF METACARPAL INTERNAL-FIXATION TECHNIQUES [J].
FIROOZBAKHSH, KK ;
MONEIM, MS ;
HOWEY, T ;
CASTANEDA, E ;
PIRELACRUZ, MA .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1993, 18A (06) :1059-1068
[8]  
Fischer KJ, 1999, J HAND SURG-AM, V24A, P928
[9]   Complications of plate fixation in metacarpal fractures [J].
Fusetti, C ;
Meyer, H ;
Borisch, N ;
Stern, R ;
Della Santa, D ;
Papaloïzos, M .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2002, 52 (03) :535-539
[10]   Biomechanical Comparison of Double-Row Locking Plates Versus Single- and Double-Row Non-Locking Plates in a Comminuted Metacarpal Fracture Model [J].
Gajendran, Varun K. ;
Szabo, Robert M. ;
Myo, George K. ;
Curtiss, Shane B. .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2009, 34A (10) :1851-1858