Localization of needle tip with color Doppler during pericardiocentesis: In vitro validation and initial clinical application

被引:30
作者
Armstrong, G
Cardon, L
Vilkomerson, D
Lipson, D
Wong, J
Rodriguez, LL
Thomas, JD
Griffin, BP
机构
[1] Cleveland Clin Fdn, Cleveland, OH 44195 USA
[2] EchoCath Inc, Princeton, NJ USA
[3] Medispectra, Lexington, MA USA
基金
美国国家航空航天局;
关键词
D O I
10.1067/mje.2001.106680
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study evaluates a new device that uses color Doppler ultrasonography to enable real-time image guidance of the aspirating needle, which has not been possible until now. The ColorMark device (EchoCath Inc, Princeton, NJ) induces high-frequency, low-amplitude vibrations in the needle to enable localization with color Doppler. We studied this technique in 25 consecutive patients undergoing pericardiocentesis, and in vitro, in a urethane phantom with which the accuracy of color Doppler localization of the needle tip was compared with that obtained by direct measurement. Tip localization was excellent in vitro; errors axial to the ultrasound beam (velocity Doppler -0.13 +/- 0.90 mm, power Doppler -0.05 +/- 1.7 mm) were less than lateral errors (velocity -0.36 +/- 1.8 mm, power -0.02 +/- 2.8 mm). In 18 of 25 patients, the needle was identified and guided into the pericardial space with the ColorMark technique, and it allowed successful, uncomplicated drainage of fluid. Initial failures were the result of incorrect settings on the echocardiographic machine and inappropriate combinations of the needle puncture site and imaging window. This study demonstrates a novel color Doppler technique that is highly accurate at localizing a needle tip. The technique is feasible for guiding pericardiocentesis. Further clinical validation of this technique is required.
引用
收藏
页码:29 / 37
页数:9
相关论文
共 16 条
  • [1] THE PUMP MANEUVER - AN ATRAUMATIC ADJUNCT TO ENHANCE US NEEDLE TIP LOCALIZATION
    BISCEGLIA, M
    MATALON, TAS
    SILVER, B
    [J]. RADIOLOGY, 1990, 176 (03) : 867 - 868
  • [2] CALLAHAN JA, 1983, J THORAC CARDIOV SUR, V85, P877
  • [3] CALLAHAN JA, 1984, AM J CARDIOL, V54, P246, DOI 10.1016/0002-9149(84)90354-0
  • [4] Pericardiocentesis guided by two-dimensional echocardiography
    Callahan, JA
    Seward, JB
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 1997, 14 (05): : 497 - 504
  • [5] CARDIAC-TAMPONADE - PERICARDIOCENTESIS DIRECTED BY TWO-DIMENSIONAL ECHOCARDIOGRAPHY
    CALLAHAN, JA
    SEWARD, JB
    TAJIK, AJ
    [J]. MAYO CLINIC PROCEEDINGS, 1985, 60 (05) : 344 - 347
  • [6] DENG JL, 1989, CHUNG HUA FANG SHE H, V23, P340
  • [7] ULTRASOUND-ASSISTED CANNULATION OF THE INTERNAL JUGULAR-VEIN - A PROSPECTIVE COMPARISON TO THE EXTERNAL LANDMARK-GUIDED TECHNIQUE
    DENYS, BG
    URETSKY, BF
    REDDY, PS
    [J]. CIRCULATION, 1993, 87 (05) : 1557 - 1562
  • [8] JOBES DR, 1995, NEW ENGL J MED, V332, P1579
  • [9] Konishi T, 1982, J Cardiogr, V12, P181
  • [10] PERCUTANEOUS PERICARDIAL CATHETER DRAINAGE - REPORT OF 42 CONSECUTIVE CASES
    KOPECKY, SL
    CALLAHAN, JA
    TAJIK, AJ
    SEWARD, JB
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1986, 58 (07) : 633 - 635