Fatigue Testing of a New Locking Plate for Hip Fractures

被引:10
作者
Hunt, Stephen [1 ]
Martin, Rod [1 ]
Woolridge, Bryan [1 ]
机构
[1] Mem Univ Newfoundland, Dept Orthoped Surg, St John, NF A1C 5S7, Canada
关键词
Hip fracture; Fatigue; Femoral neck; Trauma; Biomechnical; Locking plate; FEMORAL-NECK FRACTURES; ORTHOPEDIC DEVICES; FIXATION; METAANALYSIS; FAILURE; SCREW;
D O I
10.5405/jmbe.826
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Femoral neck shortening and varus deformity can have a significant effect on functional outcomes in patients with femoral neck fractures, particularly in young trauma patients. This study proposes a locking plate construct to address femoral neck fractures. The biomechanical properties of the locking plate are investigated with a specific focus on fatigue performance. The challenges associated with the development of a standardized pre-clinical testing model are also discussed. The development of a testing strategy and three mechanical tests are described. Each of the testing phases compare the proposed plate construct against the currently accepted gold standard of three cancellous screws. The first phase of biomechanical testing involves cadaveric specimens, which are statically loaded until ultimate failure of the construct. The second phase cyclically loads instrumented synthetic bone substitute samples with hardware installed without compression at the fracture site. The third phase cyclically loads samples with hardware installed with compression at the fracture site. Static testing results indicate that the proposed locking plate construct can withstand significantly higher static loads before failure than can the three cancellous screw technique. Fatigue testing results show that without compression across the fracture site, the samples instrumented with the locking plate construct had a shorter fatigue life than samples instrumented with 3 cancellous screws. With compression across the fracture site, the fatigue performance of the proposed locking plate is equivalent to that of 3 cancellous screws. The results indicate that the proposed locking plate construct for hip fractures has superior static strength and equivalent fatigue life compared to those of 3 cancellous screws. Biomechanical testing results show that compression across the fracture site is critical to the fatigue performance of both constructs. The development of a preclinical testing strategy for trauma products requires an understanding of the intended application of the product, expected modes of failure, and a thorough understanding of the product's service requirements.
引用
收藏
页码:117 / 122
页数:6
相关论文
共 50 条
  • [41] Effect of distal locking type on radiological results of proximal femoral nails used in the surgical treatment of hip fractures
    Davut, Serkan
    CUKUROVA MEDICAL JOURNAL, 2022, 47 (03): : 1350 - 1358
  • [42] Fatigue failure of an orthopedic implant - A locking compression plate
    Kanchanomai, C.
    Phiphobmongkol, V.
    Muanjan, P.
    ENGINEERING FAILURE ANALYSIS, 2008, 15 (05) : 521 - 530
  • [43] Unilateral locking plate versus unilateral locking plate combined with compression bolt for Schatzker I-IV tibial plateau fractures: a comparative study
    Wang, Zhongzheng
    Zheng, Zhanle
    Wang, Yuchuan
    Zhu, Yanbin
    Tan, Zhanchao
    Chen, Wei
    Hou, Zhiyong
    Zhang, Yingze
    INTERNATIONAL ORTHOPAEDICS, 2022, 46 (05) : 1133 - 1143
  • [44] Minimally invasive plate osteosynthesis for distal radius fractures with a palmar locking plate
    Takada, N.
    Otsuka, T.
    Yamada, K.
    Suzuki, H.
    Hasuo, T.
    Kondo, A.
    Fukuta, M.
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2012, 38 (06) : 627 - 632
  • [45] Minimally invasive plate osteosynthesis for distal radius fractures with a palmar locking plate
    N. Takada
    T. Otsuka
    K. Yamada
    H. Suzuki
    T. Hasuo
    A. Kondo
    M. Fukuta
    European Journal of Trauma and Emergency Surgery, 2012, 38 : 627 - 632
  • [46] Hip fractures: current review of treatment and management
    Kim, Daniel C.
    Honeycutt, Michael W.
    Riehl, John T.
    CURRENT ORTHOPAEDIC PRACTICE, 2019, 30 (04): : 385 - 394
  • [47] Comparative analysis of locking plate versus hook plate osteosynthesis of Neer type IIB lateral clavicle fractures
    Erdle, Benjamin
    Izadpanah, Kaywan
    Jaeger, Martin
    Jensen, Patrizia
    Konstantinidis, Lukas
    Zwingmann, Joern
    Suedkamp, Norbert P.
    Maier, Dirk
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2017, 137 (05) : 651 - 662
  • [48] Distal locking in short hip nails: Cause or prevention of peri-implant fractures?
    Skala-Rosenbaum, Jiri
    Dzupa, Valer
    Bartoska, Radek
    Dousa, Pavel
    Waldauf, Petr
    Krbec, Martin
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (04): : 887 - 892
  • [49] A biomechanical comparison of composite femurs and cadaver femurs used in experiments on operated hip fractures
    Basso, Trude
    Klaksvik, Jomar
    Syversen, Unni
    Foss, Olav A.
    JOURNAL OF BIOMECHANICS, 2014, 47 (16) : 3898 - 3902
  • [50] Arthroscopic locking plate removal after proximal humeral fractures
    Christine Voigt
    Andreas Geisler
    Helmut Lill
    Archives of Orthopaedic and Trauma Surgery, 2010, 130 : 391 - 395