Fatigue Testing of a New Locking Plate for Hip Fractures

被引:10
|
作者
Hunt, Stephen [1 ]
Martin, Rod [1 ]
Woolridge, Bryan [1 ]
机构
[1] Mem Univ Newfoundland, Dept Orthoped Surg, St John, NF A1C 5S7, Canada
关键词
Hip fracture; Fatigue; Femoral neck; Trauma; Biomechnical; Locking plate; FEMORAL-NECK FRACTURES; ORTHOPEDIC DEVICES; FIXATION; METAANALYSIS; FAILURE; SCREW;
D O I
10.5405/jmbe.826
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Femoral neck shortening and varus deformity can have a significant effect on functional outcomes in patients with femoral neck fractures, particularly in young trauma patients. This study proposes a locking plate construct to address femoral neck fractures. The biomechanical properties of the locking plate are investigated with a specific focus on fatigue performance. The challenges associated with the development of a standardized pre-clinical testing model are also discussed. The development of a testing strategy and three mechanical tests are described. Each of the testing phases compare the proposed plate construct against the currently accepted gold standard of three cancellous screws. The first phase of biomechanical testing involves cadaveric specimens, which are statically loaded until ultimate failure of the construct. The second phase cyclically loads instrumented synthetic bone substitute samples with hardware installed without compression at the fracture site. The third phase cyclically loads samples with hardware installed with compression at the fracture site. Static testing results indicate that the proposed locking plate construct can withstand significantly higher static loads before failure than can the three cancellous screw technique. Fatigue testing results show that without compression across the fracture site, the samples instrumented with the locking plate construct had a shorter fatigue life than samples instrumented with 3 cancellous screws. With compression across the fracture site, the fatigue performance of the proposed locking plate is equivalent to that of 3 cancellous screws. The results indicate that the proposed locking plate construct for hip fractures has superior static strength and equivalent fatigue life compared to those of 3 cancellous screws. Biomechanical testing results show that compression across the fracture site is critical to the fatigue performance of both constructs. The development of a preclinical testing strategy for trauma products requires an understanding of the intended application of the product, expected modes of failure, and a thorough understanding of the product's service requirements.
引用
收藏
页码:117 / 122
页数:6
相关论文
共 50 条
  • [21] Interprosthetic femoral fractures treated with locking plate
    Nabil Ebraheim
    Trevor Carroll
    Muhammad Z. Moral
    Justin Lea
    Adam Hirschfeld
    Jiayong Liu
    International Orthopaedics, 2014, 38 : 2183 - 2189
  • [22] Biomechanical analysis of a novel locking plate with smooth pegs versus a conventional locking plate with threaded screws for proximal humerus fractures
    Yamamoto, Nobuyuki
    Hongo, Michio
    Berglund, Lawrence J.
    Sperling, John W.
    Cofield, Robert H.
    An, Kai-Nan
    Steinmann, Scott P.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2013, 22 (04) : 445 - 450
  • [23] Surgical outcomes of non-periprosthetic distal femur fragility fractures treated with a locking plate
    Liu, Jiayong
    Hein, David
    Huffman, Christopher
    Rao, Brian M.
    Cooper, Jonathan
    Ebraheim, Nabil A.
    ANNALS OF JOINT, 2022, 7
  • [24] A comparison of proximal femoral locking plate versus 95-degree angled blade plate in the treatment of reverse intertrochanteric fractures
    Azboy, Ibrahim
    Demirtas, Abdullah
    Gem, Mehmet
    Cakir, Idris Ahmet
    Tutak, Yilmaz
    EKLEM HASTALIKLARI VE CERRAHISI-JOINT DISEASES AND RELATED SURGERY, 2014, 25 (01): : 15 - 20
  • [25] Locking versus non-locking plate fixation in the management of mandibular fractures: a meta-analysis
    Chrcanovic, B. R.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2014, 43 (10) : 1243 - 1250
  • [26] Minimally invasive plate osteosynthesis with locking plate for metastatic humeral fractures
    Matsumura, Tomohiro
    Saito, Tomohiro
    Akiyama, Toru
    Takeshita, Katsushi
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2022, 27 (06) : 1309 - 1314
  • [27] The first 100 patients treated with a new anatomical pre-contoured locking plate for clavicular midshaft fractures
    Reisch, Tania
    Camenzind, Roland S.
    Fuhrer, Reto
    Riede, Ulf
    Helmy, Naeder
    BMC MUSCULOSKELETAL DISORDERS, 2019, 20 (1)
  • [28] Extracortical plate fixation with new plate inserts and cerclage wires for the treatment of periprosthetic hip fractures
    Johannes D. Bastian
    Andre Butscher
    Gianni Bigolin
    Matthias A. Zumstein
    Hubert P. Nötzli
    International Orthopaedics, 2014, 38 : 489 - 494
  • [29] Extracortical plate fixation with new plate inserts and cerclage wires for the treatment of periprosthetic hip fractures
    Bastian, Johannes D.
    Butscher, Andre
    Bigolin, Gianni
    Zumstein, Matthias A.
    Noetzli, Hubert P.
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (03) : 489 - 494
  • [30] Complications of the locking plate for displaced proximal humeral fractures
    Lu Yi
    Wang Man-yi
    Zhu Yi-ming
    Jiang Chun-yan
    CHINESE MEDICAL JOURNAL, 2010, 123 (19) : 2671 - 2675