Integrated High Frequency Linear Transducer Array for Needle Biopsy Probe

被引:0
|
作者
Wodnicki, Robert [1 ]
Cummins, Thomas [2 ]
Jung, Hayong [1 ]
Shung, K. Kirk [1 ]
机构
[1] USC, Dept Biomed Engn, Los Angeles, CA 90007 USA
[2] USC, Dept Elect Engn, Los Angeles, CA USA
关键词
ultrasound; flex circuit; synthetic aperture imaging; high frequency;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We are developing a high frequency ultrasound imaging array for a needle biopsy probe. The goal of this work is to integrate a 64 channel high frequency linear array fabricated using PMN-PT transducer material alongside a high voltage CMOS ASIC. The ASIC is designed to multiplex the signals from the array to a single cable for synthetic aperture imaging. In this paper we discuss some of the challenges in this development and present initial results. The 64 channel transducer array was first assembled to a single layer flex circuit using gold bumps and silver-loaded epoxy. Pulse/echo data was obtained for this first prototype assembly. In addition to the gold bumped assembly, we have used silver coated glass beads and silver epoxy for assembly, and present initial results for this assembly as well.
引用
收藏
页数:3
相关论文
共 50 条
  • [1] Three-dimensional interventional photoacoustic imaging for biopsy needle guidance with a linear array transducer
    Wang, Hang
    Liu, Songde
    Wang, Tong
    Zhang, Chenxi
    Feng, Ting
    Tian, Chao
    JOURNAL OF BIOPHOTONICS, 2019, 12 (12)
  • [2] Emergency Ultrasound-guided Pericardiocentesis Using a High-frequency Linear Array Transducer to Improve Needle Tip Visualization
    Fields, J. Matthew
    Paziana, Karolina
    Vuljaj, Nikola
    Fischer, Jonathan
    Ku, Bon S.
    ACADEMIC EMERGENCY MEDICINE, 2012, 19 (05) : e16 - e16
  • [3] Design and fabrication of high frequency BNT film based linear array transducer
    Liu, Hongwei
    Ren, Wei
    Zhao, Jinyan
    Zhao, Huifeng
    Wu, Xiaoqing
    Shi, Peng
    Zhou, Qifa
    Shung, K. Kirk
    CERAMICS INTERNATIONAL, 2015, 41 : S631 - S637
  • [4] Integrated thermoacoustic and ultrasound imaging based on the combination of a hollow concave transducer array and a linear transducer array
    Wang, Xue
    Huang, Lin
    Chi, Zihui
    Jiang, Huabei
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (11):
  • [5] Mechanical failure of a curved linear array echoendoscope: loose fixation of the high frequency transducer
    Kaneko, Junichi
    Tamakoshi, Hiroki
    Niwa, Tomoyuki
    Takinami, Masaki
    Nishino, Masafumi
    Takahashi, Yurimi
    Yamada, Takanori
    ENDOSCOPY, 2024, 56 : E651 - E652
  • [6] Design and fabrication of a 20 MHz high frequency linear array ultrasonic transducer for medical use
    Lee, Wonseok
    Roh, Yongrae
    JOURNAL OF THE ACOUSTICAL SOCIETY OF KOREA, 2016, 35 (06): : 466 - 472
  • [7] Research on high frequency arc transducer array directivity
    Liu Zhen-jun
    PROCEEDINGS OF THE 2015 INTERNATIONAL INDUSTRIAL INFORMATICS AND COMPUTER ENGINEERING CONFERENCE, 2015, : 1574 - 1578
  • [8] Focused high frequency needle transducer for ultrasonic imaging and trapping
    Hsu, Hsiu-Sheng
    Zheng, Fan
    Li, Ying
    Lee, Changyang
    Zhou, Qifa
    Shung, K. Kirk
    APPLIED PHYSICS LETTERS, 2012, 101 (02)
  • [9] High-frequency Pulsed-wave Ultrasound Doppler System for Biomedical Applications with PMN-PT Needle Transducer and 30 MHz Linear Array
    Xu, Xiaochen
    Zhou, Qifa
    Ameri, Hossein
    Gottlieb, Emanuel J.
    Lai, Bruce
    Yen, Jesse T.
    Cannata, Jonathan M.
    Han, Pengdi
    Humayun, Mark S.
    Shung, K. Kirk
    2006 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-5, PROCEEDINGS, 2006, : 2234 - +
  • [10] Linear-array-based photoacoustic imaging of human microcirculation with a range of high frequency transducer probes
    Zafar, Haroon
    Breathnach, Aedan
    Subhash, Hrebesh M.
    Leahy, Martin J.
    JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (05)