Accuracy assessment of the Tensoval duo control according to the British and European Hypertension Societies' standards

被引:15
作者
De Greeff, Annemarie [1 ]
Arora, Jasleen [1 ]
Hervey, Simon [1 ]
Liu, Bing [1 ]
Shennan, Andrew H. [1 ]
机构
[1] Kings Coll London, St Thomas Hosp, Div Reprod & Endocrinol, Maternal & Fetal Res Unit, London SE1 7EH, England
关键词
accuracy; adults; auscultation; automated; blood pressure measurement; British Hypertension Society; International Protocol; validation;
D O I
10.1097/MBP.0b013e3282f3fb2e
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Noninvasive blood pressure (BP) measurement is dependent on either auscultation or oscillometry. The Tensoval duo control device uses auscultatory and/or oscillometric technology to determine BP. We evaluated the accuracy of this device in adults according to the British Hypertension Society (BHS) protocol and the International Protocol of the European Hypertension Society. Methods Ethical approval was obtained and participants gave written informed consent. Eighty-five participants who fulfilled the protocol criteria were recruited. Nine sequential same arm measurements were taken from each participant by two trained observers, comparing the device to mercury sphygmomanometery. Data analysis was performed according to the respective protocol guidelines. Results The device achieved an A grade for both systolic and diastolic pressures with 68, 91 and 98% of systolic and 73, 91 and 98% of diastolic differences within the <= 5, <= 10 and <= 15 mmHg categories of the BHS protocol. It maintained an A/A grade throughout the low, medium and high-pressure analysis. It also achieved the requirements of the Association for the Advancement of Medical Instrumentation (ANSI/AAMI SP10 protocol) with a mean difference (SD) of -2.4 (6.0) mmHg for systolic and 0.9 (6.3) mmHg for diastolic pressure. It also passed all the phases of the International Protocol. Conclusion The Tensoval duo control can be recommended for clinical and home care use in an adult population according to the criteria of the BHS protocol and the International Protocol. It is the first fully automated BP device for self-measurement to maintain an A/A grade throughout the low, medium and high-pressure ranges.
引用
收藏
页码:111 / 116
页数:6
相关论文
共 14 条
[11]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[12]   THE BRITISH-HYPERTENSION-SOCIETY PROTOCOL FOR THE EVALUATION OF AUTOMATED AND SEMIAUTOMATED BLOOD-PRESSURE MEASURING DEVICES WITH SPECIAL REFERENCE TO AMBULATORY SYSTEMS [J].
OBRIEN, E ;
PETRIE, J ;
LITTLER, W ;
DESWIET, M ;
PADFIELD, PL ;
OMALLEY, K ;
JAMIESON, M ;
ALTMAN, D ;
BLAND, M ;
ATKINS, N .
JOURNAL OF HYPERTENSION, 1990, 8 (07) :607-619
[13]   Validation of the DINAMAP ProCare blood pressure device according to the international protocol in an adult population [J].
Reinders, Annemarie ;
Reggiori, Fayrouge ;
Shennan, Andrew H. .
BLOOD PRESSURE MONITORING, 2006, 11 (05) :293-296
[14]   Lack of reproducibility in pregnancy of Korotkoff phase IV as measured by mercury sphygmomanometry [J].
Shennan, A ;
Gupta, M ;
Halligan, A ;
Taylor, DJ ;
deSwiet, M .
LANCET, 1996, 347 (8995) :139-142