Accuracy assessment of the Tensoval duo control according to the British and European Hypertension Societies' standards

被引:15
作者
De Greeff, Annemarie [1 ]
Arora, Jasleen [1 ]
Hervey, Simon [1 ]
Liu, Bing [1 ]
Shennan, Andrew H. [1 ]
机构
[1] Kings Coll London, St Thomas Hosp, Div Reprod & Endocrinol, Maternal & Fetal Res Unit, London SE1 7EH, England
关键词
accuracy; adults; auscultation; automated; blood pressure measurement; British Hypertension Society; International Protocol; validation;
D O I
10.1097/MBP.0b013e3282f3fb2e
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Noninvasive blood pressure (BP) measurement is dependent on either auscultation or oscillometry. The Tensoval duo control device uses auscultatory and/or oscillometric technology to determine BP. We evaluated the accuracy of this device in adults according to the British Hypertension Society (BHS) protocol and the International Protocol of the European Hypertension Society. Methods Ethical approval was obtained and participants gave written informed consent. Eighty-five participants who fulfilled the protocol criteria were recruited. Nine sequential same arm measurements were taken from each participant by two trained observers, comparing the device to mercury sphygmomanometery. Data analysis was performed according to the respective protocol guidelines. Results The device achieved an A grade for both systolic and diastolic pressures with 68, 91 and 98% of systolic and 73, 91 and 98% of diastolic differences within the <= 5, <= 10 and <= 15 mmHg categories of the BHS protocol. It maintained an A/A grade throughout the low, medium and high-pressure analysis. It also achieved the requirements of the Association for the Advancement of Medical Instrumentation (ANSI/AAMI SP10 protocol) with a mean difference (SD) of -2.4 (6.0) mmHg for systolic and 0.9 (6.3) mmHg for diastolic pressure. It also passed all the phases of the International Protocol. Conclusion The Tensoval duo control can be recommended for clinical and home care use in an adult population according to the criteria of the BHS protocol and the International Protocol. It is the first fully automated BP device for self-measurement to maintain an A/A grade throughout the low, medium and high-pressure ranges.
引用
收藏
页码:111 / 116
页数:6
相关论文
共 14 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Is the accuracy of blood pressure measuring devices underestimated at increasing blood pressure levels? [J].
Braam, RL ;
Thien, T .
BLOOD PRESSURE MONITORING, 2005, 10 (05) :283-289
[3]   Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol [J].
Coleman, A ;
Freeman, P ;
Steel, S ;
Shennan, A .
BLOOD PRESSURE MONITORING, 2006, 11 (01) :27-32
[4]   Validation of the Microlife BP3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol [J].
Cuckson, AC ;
Reinders, A ;
Shabeeh, H ;
Shennan, AH .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :319-324
[5]  
*DEP HLTH INF CTR, 2006, HLTH SURV ENGL 2005
[6]   Limitations of current validation protocols for home blood pressure monitors for individual patients [J].
Gerin, W ;
Schwartz, AR ;
Schwartz, JE ;
Pickering, TG ;
Davidson, KW ;
Bress, J ;
O'Brien, E ;
Atkins, N .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :313-318
[7]   ACCURACY AND PERFORMANCE OF THE TERUMO ES-H51, A NEW PORTABLE BLOOD-PRESSURE MONITOR [J].
IMAI, Y ;
HASHIMOTO, J ;
MINAMI, N ;
MUNAKATA, M ;
WATANABE, N ;
SAKUMA, H ;
SEKINO, H ;
ABE, K .
AMERICAN JOURNAL OF HYPERTENSION, 1994, 7 (03) :255-260
[8]   Validation of semi-automatic device UA-704 for self-measurement of blood pressure [J].
Kobalava, ZD ;
Kotovskaya, YV ;
Rodionov, E .
BLOOD PRESSURE MONITORING, 2005, 10 (04) :223-225
[9]  
Modesti, 1996, Blood Press Monit, V1, P63
[10]   A website for blood pressure measuring devices: dableducational.com [J].
O'Brien, E .
BLOOD PRESSURE MONITORING, 2003, 8 (04) :177-180