On-chip definition of picolitre sample injection plugs for miniaturised liquid chromatography

被引:35
|
作者
O'Neill, AP
O'Brien, P
Alderman, J
Hoffman, D
McEnery, M
Murrihy, J
Glennon, JD
机构
[1] Natl Univ Ireland Univ Coll Cork, Natl Microelect Res Ctr, Cork, Ireland
[2] Natl Univ Ireland Univ Coll Cork, Dept Chem, Cork, Ireland
关键词
injection methods; chip technology; instrumentation;
D O I
10.1016/S0021-9673(01)00999-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The fabrication of components for a miniaturised liquid chromatography system on silicon has recently been reported by our research group [J. Cap. Electrophoresis Microchip Technol. 6 (1999) 33; Analyst 125 (2000) 25]. To date, inlet and outlet connection ports, separation micro-channels (20-200 mum in width, 0.5-10 mum in depth, 15-60 mum in length), and an intersection for picolitre injection have been etched on a silicon wafer and then sealed with a Pyrex cover plate on which platinum electrodes for on-chip detection have been patterned. The platinum electrodes have been used for the amperometric detection of phenol, using 20 nl off-chip injection. In this work we present our latest results obtained with on-chip pressure driven picolitre injection, designed to realize the full capabilities of this micro-LC system. The injection volume is dependent on the micro-channel depth, width, and also on the intersection length, allowing injection in the low nanolitre to picolitre range. (C) 2001 Elsevier Science BV All rights reserved.
引用
收藏
页码:259 / 263
页数:5