Comparison of the I-Gel and the Laryngeal Mask Airway Proseal during General Anesthesia: A Systematic Review and Meta-Analysis

被引:37
|
作者
Park, Sun Kyung [1 ]
Choi, Geun Joo [2 ]
Choi, Yun Suk [1 ]
Ahn, Eun Jin [3 ]
Kang, Hyun [2 ]
机构
[1] Jeju Natl Univ, Dept Anesthesiol & Pain Med, Coll Med, Jeju, South Korea
[2] Chung Ang Univ, Coll Med, Dept Anesthesiol & Pain Med, Seoul 156756, South Korea
[3] SungKyunKwan Univ Hosp, Dept Anesthesiol & Pain Med, Seoul, South Korea
来源
PLOS ONE | 2015年 / 10卷 / 03期
关键词
SUPRAGLOTTIC AIRWAY; LMA; CROSSOVER; PRESSURE;
D O I
10.1371/journal.pone.0119469
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objectives Conflicting results have been reported for the i-gel and the laryngeal mask airway proseal (LMA-P) during general anesthesia. The objective of the current investigation was to compare the efficacy and safety of the i-gel vs. the LMA-P during general anesthesia. Methods Two authors performed searches of MEDLINE, EMBASE, CENTRAL, and Google Scholar to identify randomized clinical trials that compared the LMA-P with the i-gel during general anesthesia. A meta -analysis was performed using both random and fixed-effect models. Publication bias was evaluated using Begg's funnel plot and Egger's linear regression test. Results Twelve randomized clinical trials met the eligibility criteria. There were no significant differences in insertion success rate at the first attempt (risk ratio [RR] 1.01, 95% confidence interval [CI] 0.97, 1.06), ease of insertion (RR 1.14, 95% CI 0.93, 1.39), oropharyngeal leak pressure (OLP) (MD -1.98, 95% CI -5.41, 1.45), quality of fiberoptic view (RR 1.00, 95% CI 0.91, 1.10) and success rate of gastric tube insertion (RR 1.07, 95% CI 0.98, 1.18) between the i-gel and the LMA-P, respectively. The i-gel had a shorter insertion time than the LMA-P (MD -3.99, 95% CI -7.13, -0.84) and a lower incidence of blood staining on the device (RR 0.26, 95% CI 0.14, 0.49), sore throat (RR 0.28, 95% CI 0.15, 0.50) and dysphagia (RR 0.27, 95% CI 0.10, 0.74). Conclusions Both devices were comparable in ease of insertion to insert and both had sufficient OLP to provide a reliable airway. Only a few minor complications were reported. The i-gel was found to have fewer complications (blood staining, sore throat, dysphagia) than the LMA-P and offers certain advantages over the LMA-P in adults under general anesthesia.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] A randomised crossover trial comparing the i-gel supraglottic airway and classic laryngeal mask airway
    Janakiraman, C.
    Chethan, D. B.
    Wilkes, A. R.
    Stacey, M. R.
    Goodwin, N.
    ANAESTHESIA, 2009, 64 (06) : 674 - 678
  • [32] The clinical effectiveness of the streamlined liner of pharyngeal airway (SLIPA (TM)) compared with the laryngeal mask airway ProSeal (TM) during general anesthesia
    Choi, Yun Mi
    Cha, Su Man
    Kang, Hyun
    Baek, Chong Wha
    Jung, Yong Hun
    Woo, Young Cheol
    Kim, Jin Yun
    Koo, Gill Hoi
    Park, Sun Gyoo
    KOREAN JOURNAL OF ANESTHESIOLOGY, 2010, 58 (05) : 450 - 457
  • [33] Laryngeal Mask Airway as an Appropriate Option in Pediatric Laparoscopic Inguinal Hernia Repair: A Systematic Review and Meta-Analysis
    Yang, Cheng-Jui
    Cheng, Po -Lung
    Huang, Yan-Jiun
    Huang, Fu-Huan
    JOURNAL OF PEDIATRIC SURGERY, 2024, 59 (04) : 660 - 666
  • [34] A randomized equivalence trial comparing the i-gel and laryngeal mask airway Supreme in children
    Jagannathan, Narasimhan
    Sommers, Katherine
    Sohn, Lisa E.
    Sawardekar, Amod
    Shah, Ravi D.
    Mukherji, Isabella I.
    Miller, Steven
    Voronov, Polina
    Seraphin, Sally
    PEDIATRIC ANESTHESIA, 2013, 23 (02) : 127 - 133
  • [35] Comparison of laryngeal mask airway and endotracheal tube in general anesthesia in children
    Dong, Wei
    Zhang, Wei
    Er, Jianxu
    Liu, Jiapeng
    Han, Jiange
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2023, 26 (06)
  • [36] A comparison of the i-gel™ and the PRO-Breathe® laryngeal mask during pressure support ventilation in children
    Drake-Brockman, T. F. E.
    Ledowski, T.
    Hegarty, M.
    Gessner, M.
    von Ungern-Sternberg, B. S.
    ANAESTHESIA, 2015, 70 (12) : 1412 - 1417
  • [37] Intraocular pressure and haemodynamic responses to insertion of the i-gel, laryngeal mask airway or endotracheal tube
    Ismail, Salah A.
    Bisher, Neama A.
    Kandil, Hazem W.
    Mowafi, Hany A.
    Atawia, Hayam A.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2011, 28 (06) : 443 - 448
  • [38] A randomized controlled study to compare oropharyngeal leak pressure between I-gel™ and laryngeal mask airway supreme™ in children in lateral position under general anesthesia
    Kundal, Raksha
    Puri, Kriti
    Agrawal, Garima
    Singh, Ranju
    Pandey, Maitree
    AIN SHAMS JOURNAL OF ANESTHESIOLOGY, 2023, 15 (01)
  • [39] A Comparison of the Performance of the I-gel™ vs. the LMA-S™ during Anesthesia: A Meta-Analysis of Randomized Controlled Trials
    Chen, Xiaoguang
    Jiao, Jinghua
    Cong, Xuefeng
    Liu, Lei
    Wu, Xiaomei
    PLOS ONE, 2013, 8 (08):
  • [40] Contrast of oropharyngeal leak pressure and clinical performance of I-gel™ and LMA ProSeal™ in patients: A meta-analysis
    Tan, Yuan
    Jiang, Jingyao
    Wang, Rurong
    PLOS ONE, 2022, 17 (12):