Volumetric Imaging and Reconstruction with Stationary Head CT System using Carbon Nanotube X-ray Source Arrays

被引:1
作者
Luo, Yueting [1 ]
Spronk, Derrek [1 ]
Billingsley, Alex [2 ]
Inscoe, Christina R. [3 ]
Lee, Yueh Z. [4 ]
Zhou, Otto [3 ]
Lu, Jianping [3 ]
机构
[1] Univ N Carolina, Dept Appl Phys Sci, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Dept Biomed Engn, Chapel Hill, NC 27599 USA
[3] Univ N Carolina, Dept Phys & Astron, Chapel Hill, NC 27599 USA
[4] Univ N Carolina, Dept Radiol, Chapel Hill, NC 27599 USA
来源
MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING | 2022年 / 12031卷
关键词
CNT x-ray source; stationary CT; head CT; volumetric reconstruction; iterative reconstruction; COMPUTED-TOMOGRAPHY;
D O I
10.1117/12.2612740
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Purpose: The purpose of this study was to evaluate a novel stationary head CT system (sHCT) enabled by the carbon nanotube (CNT) field emission x-ray source array for volumetric head imaging. A data processing and image reconstruction package was developed and demonstrated for the system. Methods: The experimental sHCT system consisted of three CNT x-ray source arrays placed in parallel imaging planes separately. During imaging, the sources stayed stationary while only the bed advancing through a tunnel formed by the parallel planes. In each imaging cycle, 135 projections were acquired over 232 degrees of view. A CT ACR 464 phantom and a Kyoto head phantom were imaged to evaluate the system performance. Image uniformity, signal to noise ratio, spatial resolution, CT number and detectability were investigated for the proposed system. All images were processed and reconstructed with an iterative reconstruction-based package. Total-variation (TV) regularization methods such as the adaptive steepest descent projection onto convex set (ASDPOCS) algorithm were implemented to reduce noises and artifacts caused by the reduced projection views. Results: Volumetric data with good uniformity, high spatial resolution, and detectability for both high- and low-resolution features were demonstrated for the proposed sHCT scanner. Two scan protocols, step-and-shoot and continuous mode were compared and proved to provide similar image quality while the latter increased the total scan speed. The SIRT-ASDPOCS algorithm effectively suppressed the sparse-view and limited-view artifact and enhance the contrast noise ratio. Conclusion: We demonstrated the feasibility of sHCT for volumetric head imaging using multiple CNT x-ray source arrays. The device provides 3D images with high fidelity. The prototype sHCT system is being installed at the UNC hospital for a patient imaging study.
引用
收藏
页数:9
相关论文
共 16 条
  • [1] A dynamic micro-CT scanner based on a carbon nanotube field emission x-ray source
    Cao, G.
    Lee, Y. Z.
    Peng, R.
    Liu, Z.
    Rajaram, R.
    Calderon-Colon, X.
    An, L.
    Wang, P.
    Phan, T.
    Sultana, S.
    Lalush, D. S.
    Lu, J. P.
    Zhou, O.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (08) : 2323 - 2340
  • [2] Stationary Computed Tomography for Space and other Resource-constrained Environments
    Cramer, Avilash
    Hecla, Jake
    Wu, Dufan
    Lai, Xiaochun
    Boers, Tim
    Yang, Kai
    Moulton, Tim
    Kenyon, Steven
    Arzoumanian, Zaven
    Krull, Wolfgang
    Gendreau, Keith
    Gupta, Rajiv
    [J]. SCIENTIFIC REPORTS, 2018, 8
  • [3] Initial Clinical Evaluation of Gated Stationary Digital Chest Tomosynthesis
    Lee, Yueh Z.
    Gunnell, Taylor
    Inscoe, Christy R.
    Puett, Connor
    Lu, Jianping
    Zhou, Otto
    [J]. MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [4] Adaptive-weighted total variation minimization for sparse data toward low-dose x-ray computed tomography image reconstruction
    Liu, Yan
    Ma, Jianhua
    Fan, Yi
    Liang, Zhengrong
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (23) : 7923 - 7956
  • [5] Simulation on system configuration for stationary head CT using linear carbon nanotube x-ray source arrays
    Luo, Yueting
    Spronk, Derrek
    Lee, Yueh Z.
    Zhou, Otto
    Lu, Jianping
    [J]. JOURNAL OF MEDICAL IMAGING, 2021, 8 (05)
  • [6] Stationary Head CT with Linear CNT X-ray Source Arrays: Image Quality Assessment through Simulation
    Luo, Yueting
    Spronk, Derrek
    Inscoe, Christina R.
    Lee, Yueh
    Zhou, Otto
    Lu, Jianping
    [J]. MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [7] Misawa M., 2003, Kerntechnik, V68, P85
  • [8] Multisource X-Ray and CT: Lessons Learned and Future Outlook
    Neculaes, Vasile Bogdan
    Edic, Peter M.
    Frontera, Mark
    Caiafa, Antonio
    Wang, Ge
    De Man, Bruno
    [J]. IEEE ACCESS, 2014, 2 : 1568 - 1585
  • [9] An update on carbon nanotube-enabled X-ray sources for biomedical imaging
    Puett, Connor
    Inscoe, Christina
    Hartman, Allison
    Calliste, Jabari
    Franceschi, Dora K.
    Lu, Jianping
    Zhou, Otto
    Lee, Yueh Z.
    [J]. WILEY INTERDISCIPLINARY REVIEWS-NANOMEDICINE AND NANOBIOTECHNOLOGY, 2018, 10 (01)
  • [10] High resolution stationary digital breast tomosynthesis using distributed carbon nanotube x-ray source array
    Qian, Xin
    Tucker, Andrew
    Gidcumb, Emily
    Shan, Jing
    Yang, Guang
    Calderon-Colon, Xiomara
    Sultana, Shabana
    Lu, Jianping
    Zhou, Otto
    Spronk, Derrek
    Sprenger, Frank
    Zhang, Yiheng
    Kennedy, Don
    Farbizio, Tom
    Jing, Zhenxue
    [J]. MEDICAL PHYSICS, 2012, 39 (04) : 2090 - 2099