3D Reconstruction of Coronary Veins from a Single X-Ray Fluoroscopic Image and Pre-operative MR

被引:0
作者
Panayiotou, Maria [1 ]
Toth, Daniel [1 ,3 ]
Adem, Tamer [1 ]
Mountney, Peter [4 ]
Brost, Alexander [5 ]
Behar, Jonathan M. [1 ,2 ]
Rinaldi, C. Aldo [2 ]
Housden, R. James [1 ]
Rhode, Kawal S. [1 ]
机构
[1] Kings Coll London, Div Imaging Sci & Biomed Engn, London, England
[2] Guys & St Thomas Hosp NHS Fdn Trust, Dept Cardiol, London, England
[3] Siemens Healthcare Ltd, London, England
[4] Siemens Healthineers, Med Imaging Technol, Princeton, NJ USA
[5] Siemens Healthcare GmbH, Forchheim, Germany
来源
STATISTICAL ATLASES AND COMPUTATIONAL MODELS OF THE HEART: IMAGING AND MODELLING CHALLENGES, 2016 | 2017年 / 10124卷
基金
英国工程与自然科学研究理事会; “创新英国”项目;
关键词
Coronary veins; 3D reconstruction; X-ray fluoroscopy; CARDIAC-RESYNCHRONIZATION THERAPY; ARTERIES; VISUALIZATION; ANGIOGRAPHY;
D O I
10.1007/978-3-319-52718-5_8
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Cardiac resynchronization therapy (CRT) is an effective treatment for patients with congestive heart failure and ventricular dyssynchrony. Despite the overall efficacy of CRT, approximately 30% of patients receiving CRT do not improve. One of the main technical problems related to the CRT procedure is inadequate visualisation in Xray fluoroscopy of the venous anatomy in relation to accurate cardiac chamber visualisation. This paper proposes a novel approach for 3D reconstruction of coronary veins from a single contrast enhanced intra-operative fluoroscopy image. For this application, the method uses back-projection geometry and a Euclidean distance/angle-based cost function. The algorithm is validated on a phantom and five patient datasets, comprising six view-angle orientations for the phantom dataset and two view-angle orientations for each of the patient datasets. Median(interquartile range) 3D-reconstruction accuracies of 1.41(0.55-3.00) mm and 3.28(2.10-4.89) mm were established for the phantom and patient data, respectively. The technique can facilitate careful advancement of the cannulating guide over a guidewire or a diagnostic catheter positioned in the coronary sinus, and consequently, improve the chances of response to CRT.
引用
收藏
页码:66 / 75
页数:10
相关论文
共 15 条
[1]  
[Anonymous], 2001, Robotica, DOI DOI 10.1017/S0263574700223217
[2]   Reconstruction of coronary arteries from a single rotational X-ray projection sequence [J].
Blondel, C ;
Malandain, G ;
Vaillant, R ;
Ayache, N .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (05) :653-663
[3]  
Chen S.Y.J., 2000, METHOD APPARATUS 3 D
[4]   Visualization of the cardiac venous system using cardiac magnetic resonance [J].
Chiribiri, Amedeo ;
Kelle, Sebastian ;
Goetze, Stephan ;
Kriatselis, Charalampos ;
Thouet, Thomas ;
Tangeharoen, Tarinee ;
Paetsch, Ingo ;
Schnackenburg, Bernhard ;
Fleck, Eckart ;
Nagel, Eike .
AMERICAN JOURNAL OF CARDIOLOGY, 2008, 101 (03) :407-412
[5]   Reconstruction of coronary arteries from X-ray angiography: A review [J].
Cimen, Serkan ;
Gooya, Ali ;
Grass, Michael ;
Frangi, Alejandro F. .
MEDICAL IMAGE ANALYSIS, 2016, 32 :46-68
[6]   Cardiac MRI to Investigate Myocardial Scar and Coronary Venous Anatomy Using a Slow Infusion of Dimeglumine Gadobenate in Patients Undergoing Assessment for Cardiac Resynchronization Therapy [J].
Duckett, Simon G. ;
Chiribiri, Amedeo ;
Ginks, Matthew R. ;
Sinclair, Stephen ;
Knowles, Benjamin R. ;
Botnar, Rene ;
Carr-White, Gerry S. ;
Rinaldi, Christopher A. ;
Nagel, Eike ;
Razavi, Reza ;
Schaeffter, Tobias .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 33 (01) :87-95
[7]   SPATIAL RECONSTRUCTION OF CORONARY-ARTERIES FROM ANGIOGRAPHIC IMAGES [J].
GUGGENHEIM, N ;
DORIOT, PA ;
DORSAZ, PA ;
DESCOUTS, P ;
RUTISHAUSER, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (01) :99-110
[8]  
Jolly Marie-Pierre, 2012, Statistical Atlases and Computational Models of the Heart. Imaging and Modelling Challenges. Second International Workshop, STACOM 2011 Held in Conjunction with MICCAI 2011. Revised Selected Papers, P98, DOI 10.1007/978-3-642-28326-0_10
[9]   Noninvasive visualization of the cardiac venous system using multislice computed tomography [J].
Jongbloed, MRM ;
Lamb, HJ ;
Bax, JJ ;
Schuijf, JD ;
de Roos, A ;
van der Wall, EE ;
Schalij, MJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (05) :749-753
[10]   3D coronary reconstruction from routine single-plane coronary angiograms: Clinical validation and quantitative analysis of the right coronary artery in 100 patients [J].
Messenger, JC ;
Chen, SYJ ;
Carroll, JD ;
Burchenal, JEB ;
Kioussopoulos, K ;
Groves, BM .
INTERNATIONAL JOURNAL OF CARDIAC IMAGING, 2000, 16 (06) :413-427