Impact of Pulse Mode on Dusting Effect for Holmium Laser Lithotripsy: In Vitro Evaluation With Calcium Oxalate Monohydrate Stones

被引:5
作者
Khajeh, Nikta Rezakahn [1 ]
Black, Kristian M.
Daignault-Newton, Stephanie
Hall, Tim L.
Roberts, William W.
Ghani, Khurshid R.
机构
[1] Univ Michigan, Dept Urol, 4432 Med Sci 1,1137 Catherine St, Ann Arbor, MI 48109 USA
关键词
YAG LASER; FRAGMENTATION; DURATION; URETEROSCOPY; TECHNOLOGY; EFFICIENCY; SETTINGS;
D O I
10.1016/j.urology.2021.08.042
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE To assess the distribution of stone fragments (< 0.25-> 2 mm) after in vitro dusting laser lithotripsy with varying pulse modes using canine calcium oxalate monohydrate (COM) stones. Recent work demonstrates that fragments < 0.25 mm are ideal for dusting, and we hypothesized advanced pulse modes might improve this outcome. METHODS A 3D-printed bulb was used as a calyceal model containing a single COM stone. A 230-core fiber (Lumenis) was passed through a ureteroscope (LithoVue, Boston Scientific). Contact laser lithotripsy by a single operator was performed with dusting settings (0.5J x 30Hz; Moses Pulse120H) to deliver lkJ of energy for each trial. Short pulse (SP), long pulse (LP), Moses Distance (MD) and Moses Contact (MC) modes were tested with 5 trials for each parameter. Primary outcome was mass of fragments <0.25, <0.5, <1, and <2 mm. Laser fiber tip degradation was measured using a digital caliper. RESULTS Mass of stone fragments <0.25 mm varied from 34.6%-43.0% depending on the pulse mode, with no statistically significant differences between modes. MC (98.5%) produced a greater mass of fragments <2 mm compared to LP (86.1%; P = .046) but not SP (92.0%). Significantly less fiber tip burnback occurred with MC (0.29 mm) and MD (0.28 mm), compared to SP (0.83 mm; P < .0005). CONCLUSION Regardless of pulse mode, greater than one-third of the mass of COM stone was reduced to fragments <0.25 mm following contact laser lithotripsy. MC produced a greater mass of fragments <2 mm compared to LP and demonstrated less fiber tip burnback compared to SP. Published by Elsevier Inc.
引用
收藏
页码:53 / 57
页数:5
相关论文
共 29 条
  • [21] Quantification of Holmium: Yttrium Aluminum Garnet Optical Tip Degradation
    Mues, Adam C.
    Teichman, Joel M. H.
    Knudsen, Bodo E.
    [J]. JOURNAL OF ENDOUROLOGY, 2009, 23 (09) : 1425 - 1428
  • [22] Repeat Surgery After Ureteroscopic Laser Lithotripsy With Attempted Complete Extraction of Fragments: Long-term Follow-up
    Portis, Andrew J.
    Laliberte, Mark A.
    Heinisch, Andrew
    [J]. UROLOGY, 2015, 85 (06) : 1272 - 1278
  • [23] Optimal Power Settings for Holmium:YAG Lithotripsy
    Sea, Jason
    Jonat, Lee M.
    Chew, Ben H.
    Qiu, Jinze
    Wang, Bingqing
    Hoopman, John
    Milner, Thomas
    Teichman, Joel M. H.
    [J]. JOURNAL OF UROLOGY, 2012, 187 (03) : 914 - 919
  • [24] Holmium:YAG lithotripsy:: Optimal power settings
    Spore, SS
    Teichman, JMH
    Corbin, NS
    Champion, PC
    Williamson, EA
    Glickman, RD
    [J]. JOURNAL OF ENDOUROLOGY, 1999, 13 (08) : 559 - 566
  • [25] Impact of pulse duration on Ho:YAG laser lithotripsy: treatment aspects on the single-pulse level
    Sroka, Ronald
    Pongratz, Thomas
    Scheib, Gabriel
    Khoder, Wael
    Stief, Christian G.
    Herrmann, Thomas
    Nagele, Udo
    Bader, Markus J.
    [J]. WORLD JOURNAL OF UROLOGY, 2015, 33 (04) : 479 - 485
  • [26] Dusting Efficiency of the Moses Holmium Laser: An Automated In Vitro Assessment
    Winship, Brenton
    Wollin, Daniel
    Carlos, Evan
    Li, Jingqiu
    Peters, Chloe
    Simmons, W. Neal
    Preminger, Glenn M.
    Lipkin, Michael
    [J]. JOURNAL OF ENDOUROLOGY, 2018, 32 (12) : 1131 - 1135
  • [27] An in vitro evaluation of laser settings and location in the efficiency of the popcorn effect
    Wollin, Daniel A.
    Tom, Westin R.
    Jiang, Ruiyang
    Simmons, W. Neal
    Preminger, Glenn M.
    Lipkin, Michael E.
    [J]. UROLITHIASIS, 2019, 47 (04) : 377 - 382
  • [28] Variable Pulse Duration From a New Holmium: YAG Laser: The Effect on Stone Comminution, Fiber Tip Degradation, and Retropulsion in a Dusting Model
    Wollin, Daniel A.
    Ackerman, Anika
    Yang, Chen
    Chen, Tony
    Simmons, Walter Neal
    Preminger, Glenn M.
    Lipkin, Michael E.
    [J]. UROLOGY, 2017, 103 : 47 - 51
  • [29] Innovations in shock wave lithotripsy technology: Updates in experimental studies
    Zhou, YF
    Cocks, FH
    Preminger, GM
    Zhong, P