SiPM dynamic range for CEPC scintillator-based electromagnetic calorimeter

被引:0
|
作者
Niu, Y. [1 ,2 ,3 ]
Shi, Y. [1 ,2 ,3 ]
Zhao, H. [4 ]
Zhang, Y. [1 ,2 ,3 ]
Ruan, M. [4 ]
Liu, J. [1 ,2 ,3 ]
机构
[1] State Key Lab Particle Detect & Elect, Beijing 100049, Peoples R China
[2] State Key Lab Particle Detect & Elect, Hefei 230026, Peoples R China
[3] Univ Sci & Technol China, Dept Modern Phys, Hefei 230026, Peoples R China
[4] Chinese Acad Sci, Inst High Energy Phys, Beijing 100049, Peoples R China
来源
JOURNAL OF INSTRUMENTATION | 2021年 / 16卷 / 12期
基金
中国国家自然科学基金;
关键词
Calorimeter methods; Calorimeters; Scintillators; scintillation and light emission processes (solid; gas and liquid scintillators);
D O I
10.1088/1748-0221/16/12/T12008
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A high-granularity scintillator calorimeter readout with silicon photomultipliers (SiPMs) is an electromagnetic calorimeter (ECAL) candidate for experiments at the Circular Electron Positron Collider (CEPC). A critical design parameter of this ECAL candidate is the dynamic range of the SiPMs. This study investigates the SiPM dynamic range required for the CEPC scintillator ECAL. A model is developed on the basis of the operation principles of SiPMs to describe the response of an SiPM to light pulses within one recovery period by considering the cross-talk effect, photon detection efficiency, and number of pixels. The response curve of a 10000-pixel SiPM predicted by the model is consistent with the measured curve within 2% for an incident light pulse of up to 12000 photons. The intrinsic fluctuations of the SiPM response naturally exist in this model, and the correction of the saturation effect in the SiPM response is investigated. Monte Carlo (MC) simulation shows that the algorithm can restore the response linearity of an SiPM for an incident light pulse in which the number of photons is up to around six times the number of SiPM pixels. The model and correction program are implemented for full simulation of the ZH production Z -> nu nu H -> gamma gamma channel to evaluate the impact of the SiPM dynamic range of the CEPC scintillator ECAL on the reconstructed Higgs boson mass and the sensitivity to the Higgs signal in this channel. The results show that the CEPC scintillator ECAL equipped with no less than 4000 SiPM pixels and operated with a light yield of 20 photon-electrons per channel for a single minimum ionizing particle can meet the requirements for Higgs boson precision measurement in the di-photon channel at the CEPC.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] CALICE scintillator-SiPM calorimeter prototypes: R&D highlights and beamtests
    Liu, Jianbei
    Liu, Yong
    Ootani, Wataru
    Takeshita, Tohru
    Yang, Haijun
    Zhang, Yunlong
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2025, 1072
  • [32] Geant4 simulation of a scintillator-lead shashlik calorimeter with a SiPM readout
    Berra, A.
    NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2011, 34 (05): : 35 - 45
  • [33] Advances on inorganic scintillator-based optic fiber dosimeters
    Ding, Liang
    Wu, Qiong
    Wang, Qun
    Li, Yamei
    Perks, Richard M.
    Zhao, Liang
    EJNMMI PHYSICS, 2020, 7 (01)
  • [34] Advances on inorganic scintillator-based optic fiber dosimeters
    Liang Ding
    Qiong Wu
    Qun Wang
    Yamei Li
    Richard M. Perks
    Liang Zhao
    EJNMMI Physics, 7
  • [35] The Readout Electronics Research Design and Development for CEPC Scintillators Electromagnetic Calorimeter Prototype
    Zhao, Shensen
    Liu, Shubin
    Shen, Zhongtao
    Niu, Yazhou
    Xue, Qi'ao
    Feng, Changqing
    An, Qi
    Zhang, Yunlong
    Liu, Jianbei
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2019, 66 (07) : 1107 - 1114
  • [36] Simulation of a scintillator-based Compton telescope with micropattern readout
    Varner, Robert. L.
    Beene, James. R.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 1248 - 1254
  • [37] Scintillator-based detectors for dark matter searches I
    Kim, S. K.
    Kim, H. J.
    Kim, Y. D.
    NEW JOURNAL OF PHYSICS, 2010, 12
  • [38] Lens design challenges for scintillator-based neutron imaging
    Geppert-Kleinrath, V.
    Tafoya, L.
    Danly, C.
    Fatherley, V.
    Mendoza, E.
    Vaughan, J.
    Wilde, C.
    Volegov, P.
    RADIATION DETECTORS IN MEDICINE, INDUSTRY, AND NATIONAL SECURITY XIX, 2018, 10763
  • [39] A scintillator-based particle detector for CRYRING@ESR
    Hahn, Christoph
    Menz, Esther
    Pfaefflein, Philip
    Weber, Guenter
    Stoehlker, Thomas
    X-RAY SPECTROMETRY, 2020, 49 (02) : 338 - 341
  • [40] Determining the muon mass using a scintillator-based detector
    Woo, Neal
    Essick, John
    AMERICAN JOURNAL OF PHYSICS, 2017, 85 (08) : 611 - 618