Advanced template matching method for estimation of intervertebral kinematics of lumbar spine

被引:32
作者
Cerciello, T. [1 ]
Romano, M. [1 ]
Bifulco, P. [1 ]
Cesarelli, M. [1 ]
Allen, R. [2 ]
机构
[1] Univ Naples Federico II, Dept Biomed Elect & Telecommun Engn, I-80125 Naples, Italy
[2] Univ Southampton, Inst Sound & Vibrat Res, Southampton SO17 1BJ, Hants, England
关键词
Intervertebral kinematics; Fluoroscopic sequences; Vertebra tracking; Image template matching; INSTANTANEOUS CENTERS; FINITE CENTER; ROTATION; MOTION; INSTABILITY; VERTEBRAE; RECOGNITION; PLANE; PAIN;
D O I
10.1016/j.medengphy.2011.06.009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Diagnosis of low back pain and other degenerative spinal pathologies can be extremely difficult and, so far, there are not accepted standards. In general, such pathologies are associated with alteration of mechanical properties of spine and, in particular, with the instability of spinal motion. Intervertebral kinematics can be a valuable, objective method to assess the functionality of spinal segments. Fluoroscopic imaging system can provide continuous screening of lumbar tracts during patient's motion, with an acceptable low X-ray dose. Estimation of intervertebral kinematics relies on accurate recognition of vertebrae positions throughout the fluoroscopic sequence: specific vertebrae features are identified and tracked either by manual selection or by automated methods. This study presents a new method of vertebra tracking, based on image template matching of the contour of the vertebral body for an accurate intervertebral kinematics analysis. An image gradient operator was utilized to obtain the vertebral contours: it operates after an edge-preserving smoothing filter designed to reduce low dose X-ray image noise. Once a template is defined for each vertebra, this is used to determine the best vertebral location in each image throughout the fluoroscopic sequence. Accuracy of the proposed method was tested using images of a calibration model. Average error achieved for the intervertebral angle is of the order of 0.4 degrees and approximately 2 mm for the intervertebral centre of rotation. Five fluoroscopic lumbar sequences of healthy volunteers undergoing passive flexion extension motion were processed. The intervertebral kinematics was compared with other methods (automated and manual) by an estimation of measurement error. Results showed that the current method provides a better representation of the evolution over time of kinematic parameters. In particular, root mean square differences between the current method and a manual selection procedure performed by an experienced and trained clinician resulted 1.3 degrees for the intervertebral angles and 0.9 mm for the intervertebral trajectory. The proposed method provides an effective, automated and objective technique for estimation of intervertebral kinematics of lumbar spine. (C) 2011 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1293 / 1302
页数:10
相关论文
共 50 条
[41]   Characterizing Lumbar Spine Kinematics and Kinetics During Simulated Low-Speed Rear Impact Collisions [J].
Fewster, Kayla M. ;
Zehr, Jackie D. ;
Gooyers, Chad E. ;
Parkinson, Robert J. ;
Callaghan, Jack P. .
JOURNAL OF APPLIED BIOMECHANICS, 2022, 38 (03) :155-163
[42]   Dynamic Changes in Lumbar Spine Kinematics During Gait May Explain Improvement in Back Pain and Disability in Patients With Hip-Spine Syndrome [J].
Chen, Stephen R. ;
LeVasseur, Clarissa M. ;
Como, Christopher J. ;
Couch, Brandon K. ;
Talentino, Spencer E. ;
Klatt, Brian A. ;
O'Malley, Michael J. ;
Donaldson, William F. ;
Lee, Joon Y. ;
Shaw, Jeremy D. ;
Anderst, William J. .
SPINE, 2023, 48 (12) :867-873
[43]   Template Matching and Simplification Method for Building Features Based on Shape Cognition [J].
Yan, Xiongfeng ;
Ai, Tinghua ;
Zhang, Xiang .
ISPRS INTERNATIONAL JOURNAL OF GEO-INFORMATION, 2017, 6 (08)
[44]   A Novel Chamfer Template Matching Method Using Variational Mean Field [J].
Duc Thanh Nguyen .
2014 IEEE CONFERENCE ON COMPUTER VISION AND PATTERN RECOGNITION (CVPR), 2014, :2425-2432
[45]   Group target composition recognition method based on unfixed template matching [J].
Zhang C.-F. ;
Yang H.-W. ;
Hu W.-D. .
Xi Tong Gong Cheng Yu Dian Zi Ji Shu/Systems Engineering and Electronics, 2010, 32 (07) :1468-1470
[46]   A fast template matching method based on improved ring projection transformation and local dynamic time warping [J].
Wang, Yulin ;
Liu, Wen ;
Li, Fei ;
Li, Heng ;
Zha, Wenbin ;
He, Jianliang ;
Ma, Guocai ;
Duan, Yugang .
OPTIK, 2020, 216
[47]   Intervertebral disc deformation in the lower lumbar spine during object lifting measured in vivo using indwelling bone pins [J].
Schmid, Stefan ;
Kramers-de Quervain, Ines ;
Baumgartner, Walter .
JOURNAL OF BIOMECHANICS, 2024, 176
[48]   Effect of Increasing Implant Height on Lumbar Spine Kinematics and Foraminal Size Using the ProDisc-L Prosthesis [J].
Gaffey, John L. ;
Ghanayem, Alexander J. ;
Voronov, Michael L. ;
Havey, Robert M. ;
Carandang, Gerard ;
Abjornson, Celeste ;
Patwardhan, Avinash G. .
SPINE, 2010, 35 (19) :1777-1782
[49]   Capturing Three-Dimensional In Vivo Lumbar Intervertebral Joint Kinematics Using Dynamic Stereo-X-Ray Imaging [J].
Aiyangar, Ameet K. ;
Zheng, Liying ;
Tashman, Scott ;
Anderst, William J. ;
Zhang, Xudong .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2014, 136 (01)
[50]   Hierarchical Template Matching for 3D Myocardial Tracking and Cardiac Strain Estimation [J].
Bhalodiya, Jayendra M. ;
Palit, Arnab ;
Ferrante, Enzo ;
Tiwari, Manoj K. ;
Bhudia, Sunil K. ;
Arvanitis, Theodoros N. ;
Williams, Mark A. .
SCIENTIFIC REPORTS, 2019, 9 (1)