TBI lung dose comparisons using bilateral and anteroposterior delivery techniques and tissue density corrections

被引:20
作者
Bailey, Daniel W. [1 ,2 ]
Wang, Iris Z. [2 ,3 ]
Lakeman, Tara [3 ]
Hales, Lee D. [2 ]
Singh, Anurag K. [2 ]
Podgorsak, Matthew B. [2 ,3 ]
机构
[1] Northside Hosp Canc Inst, Dept Radiat Oncol, Atlanta, GA 30342 USA
[2] Roswell Pk Canc Inst, Dept Radiat Med, Buffalo, NY 14263 USA
[3] SUNY Buffalo, Dept Physiol & Biophys, Buffalo, NY 14260 USA
关键词
radiation therapy; total body irradiation; TBI; dosimetry; lung dose; heterogeneity correction; tissue density correction; inhomogeneity corrections; TOTAL-BODY IRRADIATION;
D O I
10.1120/jacmp.v16i2.5293
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study compares lung dose distributions for two common techniques of total body photon irradiation (TBI) at extended source-to-surface distance calculated with, and without, tissue density correction (TDC). Lung dose correction factors as a function of lateral thorax separation are approximated for bilateral opposed TBI (supine), similar to those published for anteroposterior-posteroanterior (AP-PA) techniques in AAPM Report 17 (i.e., Task Group 29). 3D treatment plans were created retrospectively for 24 patients treated with bilateral TBI, and for whom CT data had been acquired from the head to the lower leg. These plans included bilateral opposed and AP-PA techniques-each with and without -TDC, using source-toaxis distance of 377 cm and largest possible field size. On average, bilateral TBI requires 40% more monitor units than AP-PA TBI due to increased separation (26% more for 23 MV). Calculation of midline thorax dose without TDC leads to dose underestimation of 17% on average (standard deviation, 4%) for bilateral 6 MV TBI, and 11% on average (standard deviation, 3%) for 23 MV. Lung dose correction factors (CF) are calculated as the ratio of midlung dose (with TDC) to midline thorax dose (without TDC). Bilateral CF generally increases with patient separation, though with high variability due to individual uniqueness of anatomy. Bilateral CF are 5% (standard deviation, 4%) higher than the same corrections calculated for AP-PA TBI in the 6 MV case, and 4% higher (standard deviation, 2%) for 23 MV. The maximum lung dose is much higher with bilateral TBI (up to 40% higher than prescribed, depending on patient anatomy) due to the absence of arm tissue blocking the anterior chest. Dose calculations for bilateral TBI without TDC are incorrect by up to 24% in the thorax for 6 MV and up to 16% for 23 MV. Bilateral lung CF may be calculated as 1.05 times the values published in Table 6 of AAPM Report 17, though a larger patient pool is necessary to better quantify this trend. Bolus or customized shielding will reduce lung maximum dose in the anterior thorax.
引用
收藏
页码:291 / 301
页数:11
相关论文
共 8 条
[1]  
Hui Susanta K, 2004, J Appl Clin Med Phys, V5, P71, DOI 10.1120/jacmp.2022.25311
[2]   Validation of the Eclipse AAA algorithm at extended SSD [J].
Hussain, Amjad ;
Villarreal-Barajas, Eduardo ;
Brown, Derek ;
Dunscombe, Peter .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (03) :90-100
[3]   RADIATION DOSE-VOLUME EFFECTS IN THE LUNG [J].
Marks, Lawrence B. ;
Bentzen, Soren M. ;
Deasy, Joseph O. ;
Kong, Feng-Ming ;
Bradley, Jeffrey D. ;
Vogelius, Ivan S. ;
El Naqa, Issam ;
Hubbs, Jessica L. ;
Lebesque, Joos V. ;
Timmerman, Robert D. ;
Martel, Mary K. ;
Jackson, Andrew .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (03) :S70-S76
[4]   Implementation of a lateral total body irradiation technique with 6 MV photons: The University of Texas Health Science Center in San Antonio experience [J].
Mesa, Francisco ;
Eng, Tony Y. ;
Esquivel, Carlos ;
Fuller, Clifton D. ;
Papanikolaou, Niko ;
Sosa, Modesto .
JOURNAL OF RADIOTHERAPY IN PRACTICE, 2011, 10 (01) :45-54
[5]   Pretransplant pulmonary function tests predict risk of mortality following fractionated total body irradiation and allogeneic peripheral blood stem cell transplant [J].
Singh, Anurag K. ;
Karimpour, Shervin E. ;
Savani, Bipin N. ;
Guion, Peter ;
Hope, Andrew J. ;
Mansueti, John R. ;
Ning, Holly ;
Altemus, Rosemary M. ;
Wu, Colin O. ;
Barrett, A. John .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (02) :520-527
[6]  
Van Dyk J., 1986, The Physical Aspects of Total and Half Body Irradiation, DOI [10.37206/16, DOI 10.37206/16]
[7]   RADIATION PNEUMONITIS FOLLOWING LARGE SINGLE DOSE IRRADIATION - A RE-EVALUATION BASED ON ABSOLUTE DOSE TO LUNG [J].
VANDYK, J ;
KEANE, TJ ;
KAN, S ;
RIDER, WD ;
FRYER, CJH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1981, 7 (04) :461-467
[8]   A simplified technique for delivering total body irradiation (TBI) with improved dose homogeneity [J].
Yao, Rui ;
Bernard, Damian ;
Turian, Julius ;
Abrams, Ross A. ;
Sensakovic, William ;
Fung, Henry C. ;
Chu, James C. H. .
MEDICAL PHYSICS, 2012, 39 (04) :2239-2248