A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars

被引:23
作者
Piasecki, Lucila [1 ]
dos Reis, Paulo Jose [2 ]
Jussiani, Eduardo Inocente [3 ]
Andrello, Avacir Casanova [3 ]
机构
[1] SUNY Buffalo, Dept Periodont & Endodont, 240 Squire Hall, Buffalo, NY 14214 USA
[2] State Univ Cent West, Dept Phys, Guarapuava, Parana, Brazil
[3] Univ Estadual Londrina, Phys Dept, Appl Nucl Phys Lab, Londrina, Parana, Brazil
关键词
Apex ID; CanalPro; electronic apex locator; micro-computed tomography; Root ZX; working length; WORKING LENGTH DETERMINATION; IN-VIVO; ROOT CANALS; MESIAL ROOT; FORAMEN; TEETH;
D O I
10.1016/j.joen.2018.09.001
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: This study evaluated the accuracy of 3 electronic apex locators (EALs) (CanalPro [ColteneEndo, Cuyahoga Falls, OH], Apex ID [SybronEndo, Glendora, CA], and Root ZX Mini [J Morita, Tokyo, Japan]) in curved mesial canals of extracted mandibular molars using micro-computed tomographic (micro-CT) scanning. Methods: The root canal length and the actual working length of 58 canals were measured using the visual method and 3-dimensional micro-CT reconstructions. The measurements of the EALs at marks "APEX/0.0" and "0.5 mark" were recorded as the electronic root canal length and the electronic working length, respectively. The absolute mean values and the percentages of distribution of the electronic measurements were compared with the actual lengths of the canals. The presence of isthmuses, accessory canals, foramen position, and the distance between the apical constriction and the apical foramen were recorded from the microCT scans, and their correlation to the accuracy of the EALs was determined. Results: All electronic measurements showed high agreement with their respective gold standard, except the electronic root canal length of the Apex ID (P < .05). No difference in the percentage of precise measurements (within the +/- 0.5 mm) was found among the devices (chi(2) and z tests, P > .05). Of the anatomic parameters evaluated, the presence of a lateral foramen negatively affected the 0.0 mark of the Apex ID (Wilcoxon test, P < .05). Conclusions: The Root ZX Mini and CanalPro were precise for both root canal length and working length determination in mesial curved canals of mandibular molars, whereas the Apex ID was accurate for the working length when using the 0.5 mark.
引用
收藏
页码:1872 / 1877
页数:6
相关论文
共 35 条
[1]  
Aguiar BA, 2017, INT J DENT, V2017, DOI 10.1155/2017/5893790
[2]   Not All Electronic Foramen Locators Are Accurate in Teeth with Enlarged Apical Foramina: An In Vitro Comparison of 5 Brands [J].
Akisue, Eduardo ;
Gratieri, Simone Domingues ;
Barletta, Fernando Branco ;
Caldeira, Celso Luiz ;
Grazziotin-Soares, Renata ;
Gavini, Giulio .
JOURNAL OF ENDODONTICS, 2014, 40 (01) :109-112
[3]   Evaluation of the accuracy of nine electronic apex locators by using Micro-CT [J].
Connert, T. ;
Judenhofer, M. S. ;
Huelber-J, M. ;
Schell, S. ;
Mannheim, J. G. ;
Pichler, B. J. ;
Loest, C. ;
ElAyouti, A. .
INTERNATIONAL ENDODONTIC JOURNAL, 2018, 51 (02) :223-232
[4]   Accuracy of 2 Endodontic Rotary Motors with Integrated Apex Locator [J].
Cruz, Alessandra Timponi Goes ;
Wichnieski, Caroline ;
Carneiro, Everdan ;
da Silva Neto, Ulisses Xavier ;
Gambarini, Gianluca ;
Piasecki, Lucila .
JOURNAL OF ENDODONTICS, 2017, 43 (10) :1716-1719
[5]   A 3-DIMENSIONAL STUDY OF CANAL CURVATURES IN THE MESIAL ROOTS OF MANDIBULAR MOLARS [J].
CUNNINGHAM, CJ ;
SENIA, ES .
JOURNAL OF ENDODONTICS, 1992, 18 (06) :294-300
[6]   Influence of Preflaring on the Accuracy of Length Determination With Four Electronic Apex Locators [J].
de Camargo, Ericson Janolio ;
Zapata, Ronald Ordinola ;
Medeiros, Paulo Leal ;
Bramante, Clovis Monteiro ;
Bernardineli, Norberti ;
Garcia, Roberto Brandao ;
de Moraes, Ivaldo Gomes ;
Hungaro Duarte, Marco Antonio .
JOURNAL OF ENDODONTICS, 2009, 35 (09) :1300-1302
[7]   Ex Vivo Evaluation of the Accuracy of Electronic Foramen Locators in Root Canals with an Obstructed Apical Foramen [J].
de Vasconcelos, Bruno Carvalho ;
Verissimo Chaves, Rebeca Dibe ;
Vivacqua-Gomes, Nilton ;
de Miranda Candeiro, George Taccio ;
Bernardes, Ricardo Affonso ;
Vivan, Rodrigi Ricci ;
Hungaro Duarte, Marco Antonio .
JOURNAL OF ENDODONTICS, 2015, 41 (09) :1551-1554
[8]   Investigation of Apex Locators and Related Morphological Factors [J].
Ding, Jiangfeng ;
Gutmann, James L. ;
Fan, Bing ;
Lu, Yao ;
Chen, Hao .
JOURNAL OF ENDODONTICS, 2010, 36 (08) :1399-1403
[9]   Comparison of In Vivo and In Vitro Readings When Testing the Accuracy of the Root ZX Apex Locator [J].
Duran-Sindreu, Fernando ;
Stoeber, Eva ;
Mercade, Montse ;
Vera, Jorge ;
Garcia, Marc ;
Bueno, Rufino ;
Roig, Miguel .
JOURNAL OF ENDODONTICS, 2012, 38 (02) :236-239
[10]   A simple mounting model for consistent determination of the accuracy and repeatability of apex locators [J].
ElAyouti, A ;
Löst, C .