A New Frontier in Temporal Bone Imaging: Photon-Counting Detector CT Demonstrates Superior Visualization of Critical Anatomic Structures at Reduced Radiation Dose

被引:82
作者
Benson, J. C. [1 ]
Rajendran, K. [1 ]
Lane, J. I. [1 ]
Diehn, F. E. [1 ]
Weber, N. M. [1 ]
Thorne, J. E. [1 ]
Larson, N. B. [2 ]
Fletcher, J. G. [1 ]
McCollough, C. H. [1 ]
Leng, S. [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Quantitat Hlth Sci, Rochester, MN USA
关键词
D O I
10.3174/ajnr.A7452
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Photon-counting detector CT is a new technology with a limiting spatial resolution of < 150 mu m. In vivo comparisons between photon-counting detector CT and conventional energy-integrating detector CT are needed to determine the clinical impact of photon counting-detector CT in temporal bone imaging. MATERIALS AND METHODS: Prospectively recruited patients underwent temporal bone CT examinations on an investigational photon-counting detector CT system after clinically indicated temporal bone energy-integrating detector CT. Photon-counting detector CT images were obtained at an average 31% lower dose compared with those obtained on the energy-integrating detector CT scanner. Reconstructed images were evaluated in axial, coronal, and Poschl planes using the smallest available section thickness on each system (0.4 mm on energy-integrating detector CT; 0.2 mm on photon-counting detector CT). Two blinded neuroradiologists compared images side-by-side and scored them using a 5-point Likert scale. A post hoc reassignment of readers? scores was performed so that the scores reflected photon-counting detector CT performance relative to energy-integrating detector CT. RESULTS: Thirteen patients were enrolled, resulting in 26 image sets (left and right sides). The average patient age was 63.6 [SD, 13.4] years; 7 were women. Images from the photon-counting detector CT scanner were significantly preferred by the readers in all reconstructed planes (P <.001). Photon-counting detector CT was rated superior for the evaluation of all individual anatomic structures, with the oval window (4.79) and incudostapedial joint (4.75) receiving the highest scores on a Likert scale of 1-5. CONCLUSIONS: Temporal bone CT images obtained on a photon-counting detector CT scanner were rated as having superior spatial resolution and better critical structure visualization than those obtained on a conventional energy-integrating detector scanner, even with a substantial dose reduction.
引用
收藏
页码:579 / 584
页数:6
相关论文
共 19 条
[1]   Photon-counting x-ray detectors for CT [J].
Danielsson, Mats ;
Persson, Mats ;
Sjolin, Martin .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03)
[2]   A semi-automated quantitative comparison of metal artifact reduction in photon-counting computed tomography by energy-selective thresholding [J].
Do, T. D. ;
Sawall, S. ;
Heinze, S. ;
Reiner, T. ;
Ziener, C. H. ;
Stiller, W. ;
Schlemmer, H. P. ;
Kachelriess, M. ;
Kauczor, H. U. ;
Skornitzke, S. .
SCIENTIFIC REPORTS, 2020, 10 (01)
[3]   Computed tomography with a full FOV photon-counting detector in a clinical setting, the first experience [J].
Ferda, Jiri ;
Vendis, Tomas ;
Flohr, Thomas ;
Schmidt, Bernhard ;
Henning, Andre ;
Ulzheimer, Stefan ;
Pecen, Ladislav ;
Ferdova, Eva ;
Baxa, Jan ;
Mirka, Hynek .
EUROPEAN JOURNAL OF RADIOLOGY, 2021, 137
[4]   Novel ultrahigh resolution data acquisition and image reconstruction for multi-detector row CT [J].
Flohr, T. G. ;
Stierstorfer, K. ;
Suess, C. ;
Primak, A. N. ;
McCollough, C. H. .
MEDICAL PHYSICS, 2007, 34 (05) :1712-1723
[5]   Photon-counting CT review [J].
Flohr, Thomas ;
Petersilka, Martin ;
Henning, Andre ;
Ulzheimer, Stefan ;
Ferda, Jiri ;
Schmidt, Bernhard .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 :126-136
[6]   Image Quality Assessment for Clinical Cadmium Telluride-Based Photon-Counting Computed Tomography Detector in Cadaveric Wrist Imaging [J].
Grunz, Jan-Peter ;
Huflage, Henner ;
Heidenreich, Julius Frederik ;
Erguen, Suleyman ;
Petersilka, Martin ;
Allmendinger, Thomas ;
Bley, Thorsten Alexander ;
Petritsch, Bernhard .
INVESTIGATIVE RADIOLOGY, 2021, 56 (12) :785-790
[7]   Cross Sectional Imaging of the Ear and Temporal Bone [J].
Juliano A.F. .
Head and Neck Pathology, 2018, 12 (3) :302-320
[8]   Radiation dose and image conspicuity comparison between conventional 120 kVp and 150 kVp with spectral beam shaping for temporal bone CT [J].
Kim, Chang Rae ;
Jeon, Ji Young .
EUROPEAN JOURNAL OF RADIOLOGY, 2018, 102 :68-73
[9]   150-μm Spatial Resolution Using Photon-Counting Detector Computed Tomography Technology Technical Performance and First Patient Images [J].
Leng, Shuai ;
Rajendran, Kishore ;
Gong, Hao ;
Zhou, Wei ;
Halaweish, Ahmed F. ;
Henning, Andre ;
Kappler, Steffen ;
Baer, Matthias ;
Fletcher, Joel G. ;
McCollough, Cynthia H. .
INVESTIGATIVE RADIOLOGY, 2018, 53 (11) :655-662
[10]   Dose-efficient ultrahigh-resolution scan mode using a photon counting detector computed tomography system [J].
Leng, Shuai ;
Yu, Zhicong ;
Halaweish, Ahmed ;
Kappler, Steffen ;
Hahn, Katharina ;
Henning, Andre ;
Li, Zhoubo ;
Lane, John ;
Levin, David L. ;
Jorgensen, Steven ;
Ritman, Erik ;
McCollough, Cynthia .
JOURNAL OF MEDICAL IMAGING, 2016, 3 (04)