A generalized least significant change for individuals measured on different DXA systems

被引:51
|
作者
Shepherd, John A. [1 ]
Lu, Ying [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94143 USA
关键词
accuracy; bone mineral density (BMD); cross-calibration; dual-energy X-ray absorptiometry (DXA); GE healthcare; hologic; least significant change (LSC); precision; X-RAY ABSORPTIOMETRY; UNIVERSAL STANDARDIZATION; PRECISION ASSESSMENT; BONE DENSITOMETERS; CROSS-CALIBRATION; DENSITY; PERFORMANCE; WOMEN;
D O I
10.1016/j.jocd.2007.05.002
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In this article, we derive a generalized expression for the least significant change (LSC), which we call the generalized LSC (GLSC), to be used when an individual is measured on 2 different systems. The commonly used LSC is defined as the least amount of change between 2 measurements over time that must be exceeded before a change can be considered true (with 95% confidence). The LSC has clinical applications in monitoring disease progression or treatment effects in bone mineral density (BMD) and bone mineral content. Mathematically, the "ideal" LSC (ILSC) is 2.77 times the precision errors for measures on a single system. When BMD values of an individual are measured by 2 different systems, the LSC will depend not only on the precision errors of both systems but also on the calibration relationship between the systems. Like the ILSC, the GLSC is a simple equation applicable for inter machine comparisons. The GLSC can be defined for any 2 systems with measures obtained from cross-calibration and precision studies using the protocols recommended by the International Society for Clinical Densitometry. We validated the GLSC using 10,000 simulated measurements taken between 2 systems and offer several common uses of the GLSC such as system upgrades within a single manufacturer and replacement of 1 manufacturer by another. We found that when upgrading a Hologic QDR-2000 to a QDR-4500, GLSC was twice as large as the QDR-2000 LSC (0.0432 and 0.0215 g/cm(2) respectively). The GLSC was 2.6 (spine) to 3.6 (total hip) times larger than the LSC when comparing scans between the Hologic Delphi and the GE Lunar Prodigy. We also explore how the magnitude of the correlation coefficient and sample size change the GLSC and show that a correlation coefficient less than 0.95 increases the %GLSC to above 10%, and that increasing study sample sizes beyond 30 in the cross-calibration studies can only decrease the magnitude of the GLSC accuracy by 4%. We conclude that the GLSC, defined using commonly used clinical cross-calibration and precision assessments, is the most accurate method to compare scans between dual-energy X-ray absorptiometry systems.
引用
收藏
页码:249 / 258
页数:10
相关论文
共 24 条
  • [1] Comparing BMD results between two similar DXA systems using the generalized least significant change
    Shepherd, John A.
    Morgan, Sarah L.
    Lu, Ying
    JOURNAL OF CLINICAL DENSITOMETRY, 2008, 11 (02) : 237 - 242
  • [2] Comparing BMD Results Between 2 Similar DXA Systems Using the Generalized Least Significant Change
    Blake, Glen M.
    JOURNAL OF CLINICAL DENSITOMETRY, 2009, 12 (03) : 404 - 406
  • [3] Is a Fixed Value for the Least Significant Change Appropriate?
    Nelson, Lisa
    Gulenchyn, Karen Y.
    Atthey, Mark
    Webber, Colin E.
    JOURNAL OF CLINICAL DENSITOMETRY, 2010, 13 (01) : 18 - 23
  • [4] Effect of Hand Positioning on DXA Total and Regional Bone and Body Composition Parameters, Precision Error, and Least Significant Change
    Thurlow, Shane
    Oldroy, Brian
    Hind, Karen
    JOURNAL OF CLINICAL DENSITOMETRY, 2018, 21 (03) : 375 - 382
  • [5] Cross-calibration, Least Significant Change and Quality Assurance in Multiple Dual-Energy X-ray Absorptiometry Scanner Environments: 2019 ISCD Official Position
    Jankowski, Lawrence G.
    Warner, Sarah
    Gaither, Ken
    Lenchik, Leon
    Fan, Bo
    Lu, Ying
    Shepherd, John
    JOURNAL OF CLINICAL DENSITOMETRY, 2019, 22 (04) : 472 - 483
  • [6] Lack of gender-related differences in least significant changes for DXA body composition analysis in adult subjects
    Saravi, Fernando D.
    Henriquez, Muriel M.
    Visbal, Jorge H. Wilches
    ACTUALIZACIONES EN OSTEOLOGIA, 2022, 18 (02) : 75 - 81
  • [7] Executive Summary of the 2019 ISCD Position Development Conference on Monitoring Treatment, DXA Cross-calibration and Least Significant Change, Spinal Cord Injury, Peri-prosthetic and Orthopedic Bone Health, Transgender Medicine, and Pediatrics
    Shuhart, Christopher R.
    Yeap, Swan Sim
    Anderson, Paul A.
    Jankowski, Lawrence G.
    Lewiecki, E. Michael
    Morse, Leslie R.
    Rosen, Harold N.
    Weber, David R.
    Zemel, Babette S.
    Shepherd, John A.
    JOURNAL OF CLINICAL DENSITOMETRY, 2019, 22 (04) : 453 - 471
  • [8] Dual-Energy X-Ray Absorptiometry Measured Regional Body Composition Least Significant Change: Effect of Region of Interest and Gender in Athletes
    Buehring, Bjoern
    Krueger, Diane
    Libber, Jessie
    Heiderscheit, Bryan
    Sanfilippo, Jennifer
    Johnson, Brian
    Haller, Irina
    Binkley, Neil
    JOURNAL OF CLINICAL DENSITOMETRY, 2014, 17 (01) : 121 - 128
  • [9] The least significant change on bone mineral density scan increased in patients with higher degrees of obesity
    Dorilleau, Claire
    Kanagaratnam, Lukshe
    Charlot, Isabelle
    Hittinger, Ambre
    Bertin, Eric
    Salmon, Jean-Hugues
    Geoffroy, Marion
    AGING CLINICAL AND EXPERIMENTAL RESEARCH, 2024, 36 (01)
  • [10] Estimating Lumbar Spine Least Significant Change for Fewer than Four Vertebrae: The Manitoba BMD Registry
    Rosen, Harold
    Szalat, Auryan
    Leslie, William D.
    JOURNAL OF CLINICAL DENSITOMETRY, 2024, 27 (02)