A scintillation detector of protons and neutrons, products of a dd reaction

被引:4
作者
Kornilov, VN
Maximov, VV
机构
[1] Novosibirsk State Univ, Novosibirsk 630090, Russia
[2] Russian Acad Sci, Siberian Div, Budker Inst Nucl Phys, Novosibirsk 630090, Russia
关键词
Physical Chemistry; Deuterium; Time Resolution; Vacuum Chamber; Atomic Beam;
D O I
10.1023/A:1017571221230
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A detector of 3.02-MeV protons and 2.45-MeV neutrons is designed for experiments with the injection of deuterium atomic beams in a Gas-Dynamic Trap facility. An organic scintillator and a phototube with a fine-mesh dynode system are used in the sensor. The detector is operated in the mode of single-particle counting; the time resolution is similar to 20 ns. Discrimination between different particles is effected by using a controlled damper that shuts off the proton flow and also by analyzing pulse-height spectra. The sensor is inserted in a vacuum chamber; the rotating mechanism of the sensor and location of the ports in the facility enables measurements along the plasma column.
引用
收藏
页码:219 / 223
页数:5
相关论文
共 7 条
  • [1] ANIKEEV AV, 1999, FIZ PLAZMY MOSCOW, V25, P499
  • [2] *CERN, 1999, PAW CERN PROGR LIBR
  • [3] Huddlestone R. H., 1965, Plasma Diagnostic Techniques
  • [4] MIRROR-BASED NEUTRON SOURCES FOR FUSION TECHNOLOGY STUDIES
    IVANOV, AA
    RYUTOV, DD
    [J]. NUCLEAR SCIENCE AND ENGINEERING, 1990, 106 (03) : 235 - 242
  • [5] KVASHNIN AN, 1999, PRIB TEKH EKSP, P81
  • [6] Mirnov V. V., 1980, VOPROSY ATOMNOJ NAUK, P57
  • [7] PUCHEROV NN, 1975, TABLITSY TORMOZNOI S