Determination of lead by hydride generation atom trapping flame atomic absorption spectrometry

被引:24
作者
Ertas, Nusret [2 ]
Arslan, Zikri [1 ]
Tyson, Julian F. [3 ]
机构
[1] Jackson State Univ, Dept Chem, Jackson, MS 39217 USA
[2] Gazi Univ, Fac Pharm, Dept Analyt Chem, TR-06339 Ankara, Turkey
[3] Univ Massachusetts, Dept Chem, Amherst, MA 01003 USA
关键词
D O I
10.1039/b712126a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An atom trapping approach is described for the determination of lead by hydride generation. ame atomic absorption spectrometry. Lead hydride (PbH4), generated on-line by reacting lead in hydrochloric acid -potassium ferricyanide medium with sodium borohydride (NaBH4), was trapped on the interior walls of a slotted T-tube under highly oxidizing flame conditions. Atomization was achieved by aspirating 50 mu L of methyl isobutyl ketone (MIBK) to the flame. Optimization of the experimental parameters for the generation and collection of lead hydride was described to achieve the highest peak height sensitivity. The detection limits were 0.075, 0.047 and 0.028 mu g L (-1) for 2.6, 5.2 and 7.8 mL of blank solution (n = 13), respectively. Calibration was linear up to 3.0 mg L (-1) for a 30-s trapping period. The relative standard deviation was between 2 and 4%. The method was successfully applied to the determination of Pb in NIST certified reference materials, San Joaquin Soil (SRM 2709) and Apple Leaves (SRM 1515).
引用
收藏
页码:223 / 228
页数:6
相关论文
共 50 条