Using temperature to optimize resolution and reduce analysis times for bioanalytical diastereomer LC-MS/MS separations

被引:7
作者
Cunliffe, Jennifer M. [1 ]
Dreyer, Daniel P. [1 ]
Hayes, Roger N. [1 ]
Clement, Robert P. [1 ]
Shen, Jim X. [1 ]
机构
[1] Merck Res Labs, Summit, NJ 07901 USA
关键词
Bioanalytical; Diastereomer separations; High temperature HPLC; Resolution; PHASE LIQUID-CHROMATOGRAPHY; ULTRA-HIGH-PRESSURE; ELEVATED-TEMPERATURES; MASS-SPECTROMETRY; MOBILE PHASES; RETENTION; VALIDATION; PARTICLES; MOLECULES;
D O I
10.1016/j.jpba.2010.08.017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A rugged and reproducible liquid chromatographic tandem mass spectrometric bioanalytical method was developed for the quantitation of drug stereoisomers in human plasma. Column temperature was shown to be an important variable toward optimizing diastereomer selectivity, resolution and analysis cycle time. Non-linear Van't Hoff plots and changes in peak shape with temperature suggested that selectivity was governed by multiple retention mechanisms. The high temperature chromatography method was validated and used to analyze samples from human clinical trials. Utilization of high temperature chromatography offered alternative selectivity and is a viable approach for difficult separations in regulated bioanalysis. (c) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:179 / 185
页数:7
相关论文
共 37 条
[1]   Some advantages of high temperature for the separation of pharmaceutical compounds with mass spectrometry detection [J].
Albert, M ;
Cretier, G ;
Guillarme, D ;
Heinisch, S ;
Rocca, JL .
JOURNAL OF SEPARATION SCIENCE, 2005, 28 (14) :1803-1811
[2]  
[Anonymous], 2001, FDA Guidance for industry: Bioanalytical method validation
[3]  
ANTIA FD, 1988, J CHROMATOGR, V435, P1
[4]   Key elements of bioanalytical method validation for small molecules [J].
Bansal, Surendra ;
DeStefano, Anthony .
AAPS JOURNAL, 2007, 9 (01) :E109-E114
[5]   Investigation of retention on bare silica using reversed-phase mobile phases at elevated temperatures [J].
Bidlingmeyer, BA ;
Henderson, J .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1060 (1-2) :187-193
[6]   Rationalisation of unusual changes in efficiency and retention with temperature shown for bases in reversed-phase high-performance liquid chromatography at intermediate pH [J].
Buckenmaier, SMC ;
McCalley, DV ;
Euerby, MR .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1060 (1-2) :117-126
[7]   Effect of phase ratio on van't Hoff analysis in reversed-phase liquid chromatography, and phase-ratio-independent estimation of transfer enthalpy [J].
Chester, TL ;
Coym, JW .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1003 (1-2) :101-111
[8]  
Coleman MichaelD., 2005, HUMAN DRUG METABOLIS
[9]   ROLE OF THE TEMPERATURE IN REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING PYROCARBON-CONTAINING ADSORBENTS [J].
COLIN, H ;
DIEZMASA, JC ;
GUIOCHON, G ;
CZAJKOWSKA, T ;
MIEDZIAK, I .
JOURNAL OF CHROMATOGRAPHY, 1978, 167 (DEC) :41-65
[10]   Reversed-phase retention thermodynamics of pure-water mobile phases at ambient and elevated temperature [J].
Coym, JW ;
Dorsey, JG .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1035 (01) :23-29