Biomechanical comparison of anatomical plating systems for comminuted distal humeral fractures

被引:22
|
作者
Varady, Patrick A. [1 ,2 ]
von Rueden, Christian [1 ,2 ,3 ]
Greinwald, Markus [1 ,2 ]
Hungerer, Sven [1 ,2 ,3 ]
Paetzold, Robert [1 ,2 ,3 ]
Augat, Peter [1 ,2 ]
机构
[1] Paracelsus Med Univ, Inst Biomech, Salzburg, Austria
[2] Trauma Ctr Murnau, Inst Biomech, Prof Kuntscher Str 8, D-82418 Murnau, Germany
[3] Trauma Ctr Murnau, Dept Trauma Surg, Prof Kuntscher Str 8, D-82418 Murnau, Germany
关键词
Anatomical poly-axial plate; Distal humeral fracture; Failuremechanism; Fatigue limit; Systemstiffness; FIXATION; PARALLEL; LOCKING; CONFIGURATIONS; FORCES; JOINT;
D O I
10.1007/s00264-017-3444-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Six different mono-axial and poly-axial distal humeral plating systems with an anatomical plate design were compared. The aim of the biomechanical tests was to examine differences regarding system stiffness, median fatigue limit, and failure mechanisms. Method Different configurations of two double plate fixation systems by two manufacturers for the treatment of complex distal humeral fractures (AO/OTA type C2.3) were biomechanically tested in a physiologically relevant setup. Result The 180A degrees Stryker configuration presented itself as the system with the highest stiffness, being significantly stiffer (p < 0.001) than every system other than the poly-axial 180A degrees aap system (p = 0.378). For the median fatigue limit the 180A degrees Stryker and poly-axial aap systems were ranked first and second. The failure mechanism for all 90A degrees systems was a fatigue breakage of the posterolateral plate. The 180A degrees aap systems demonstrated breakage of the most distal screws of the lateral plate. The 180A degrees Stryker system demonstrated screw breakage on both the medial and lateral plates. Discussion Breakage of the posterolateral plate as a failure mechanism for the 90A degrees systems was expected. The 180A degrees systems demonstrated a higher stiffness compared to the 90A degrees constructs for the axial loading. In conclusion, both poly-axial anatomical plating systems provide sufficient stability in this scenario, and the 180A degrees configurations demonstrated superior stiffness.
引用
收藏
页码:1709 / 1714
页数:6
相关论文
共 50 条
  • [1] Biomechanical comparison of anatomical plating systems for comminuted distal humeral fractures
    Patrick A. Varady
    Christian von Rüden
    Markus Greinwald
    Sven Hungerer
    Robert Pätzold
    Peter Augat
    International Orthopaedics, 2017, 41 : 1709 - 1714
  • [2] Dual Nonlocked Plating as an Alternative to Locked Plating for Comminuted Distal Fibula Fractures: A Biomechanical Comparison Study
    Randall, Rachel M.
    Nagle, Tara
    Steckler, Andrew
    Billow, Damien
    Berkowitz, Mark J.
    JOURNAL OF FOOT & ANKLE SURGERY, 2019, 58 (05) : 916 - 919
  • [3] A NEW ANATOMICAL PLATE FOR EXTRA-ARTICULAR DISTAL HUMERAL FRACTURES: BIOMECHANICAL STUDY
    Mutlu, Harun
    Polat, Abdulkadir
    Cetin, Mehmet Umit
    Mutlu, Serhat
    Demir, Teyfik
    Parmaksizoglu, Atilla Sancar
    ACTA ORTOPEDICA BRASILEIRA, 2022, 30 (01):
  • [4] Biomechanical and anatomical considerations for dual plating of distal femur fractures: a systematic literature review
    DeKeyser, Graham J.
    Hakim, Anne J.
    O'Neill, Dillon C.
    Schlickewei, Carsten W.
    Marchand, Lucas S.
    Haller, Justin M.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (10) : 2597 - 2609
  • [5] Dorsal Distraction Plating for Highly Comminuted Distal Radius Fractures
    Lee, Daniel J.
    Elfar, John C.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (02): : 355 - 357
  • [6] Biomechanical comparison of screw osteosyntheses and anatomical plating for coronoid shear fractures of the ulna
    Rausch, Valentin
    Jettkant, Birger
    Lotzien, Sebastian
    Rosteius, Thomas
    Mempel, Eileen
    Schildhauer, Thomas A.
    Seybold, Dominik
    Gessmann, Jan
    Koenigshausen, Matthias
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (09) : 1509 - 1515
  • [7] Distal Ulna Fractures: A Biomechanical Comparison of Locking Versus Nonlocking Plating Constructs
    Collins, Melissa
    Hart, Adam
    Hines, Jerod
    Steffen, Thomas
    Harvey, Edward J.
    Martineau, Paul A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2014, 28 (08) : 470 - 475
  • [8] A biomechanical comparison of two plating techniques in lateral clavicle fractures
    Suter, Cyrill
    von Rohr, Martina
    Majewski, Martin
    Duerselen, Lutz
    Warnecke, Daniela
    Schild, Natalie
    Nowakowski, Andrej Maria
    CLINICAL BIOMECHANICS, 2019, 67 : 78 - 84
  • [9] Biomechanical Comparison of Fixed- versus Variable-Angle Locking Screws for Distal Humerus Comminuted Fractures
    Nourbakhsh, Ali
    Hirschfeld, Adam G.
    Dhulipala, Sravan
    Hutton, William
    Ganey, Timothy
    Lozada, Luis
    Schlatterer, Daniel
    Lourie, Gary Mark
    CLINICS IN ORTHOPEDIC SURGERY, 2019, 11 (03) : 302 - 308
  • [10] Biomechanical evaluation of a modified proximal humeral locking plate application for distal extra-articular diaphyseal humeral fractures
    Lim, Joon-Ryul
    Yoon, Tae-Hwan
    Choi, Yun-Rak
    Lee, Hwan-Mo
    Chun, Yong-Min
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (09) : 1877 - 1883