Aqueous shunts in the treatment of glaucoma

被引:14
作者
Mosaed, Sameh [1 ]
Minckler, Donald S. [1 ]
机构
[1] Univ Calif Irvine, Gavin Herbert Eye Inst, Irvine, CA 92717 USA
关键词
aqueous shunts; glaucoma drainage device; glaucoma implant; tube shunt; surgical glaucoma; REFRACTORY PEDIATRIC GLAUCOMA; OPEN-ANGLE GLAUCOMA; AHMED GLAUCOMA; DRAINAGE DEVICES; TERM OUTCOMES; SCLERAL FLAP; FOLLOW-UP; SILICONE; SURGERY; VALVE;
D O I
10.1586/ERD.10.32
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The main goal of surgical glaucoma treatment is to reduce the intraocular pressure. The standard procedure over the past 40 years has been trabeculectomy, first described in 1968. This procedure comes with a very high risk profile, and results in moderate long-term results at best. Tube shunt devices were introduced as an alternate means to divert fluid out of the glaucomatous eye, attempting to improve upon the poor long-term outcomes of trabeculectomy, particularly in refractory cases that had failed attempts at prior glaucoma surgeries. They are currently available in a handful of different designs and have evolved from representing a second- or third-line treatment in refractory cases, to their current standing of accepted first-line surgical intervention for many surgeons. The designs of the aqueous shunts are simple and installation involves straightforward principles of ophthalmic surgery. Recent additions to the literature, as well as the development of the Ex-PRESS Miniature Shunt, are changing the way shunts are employed in the treatment algorithm for surgical glaucoma. The outcomes of these procedures are encouraging, and support their use as first-line surgical options for the treatment of adult and juvenile glaucoma.
引用
收藏
页码:661 / 666
页数:6
相关论文
共 28 条
  • [1] Chen TC, 2005, J PEDIAT OPHTH STRAB, V42, P274
  • [2] Pars plana insertion of glaucoma drainage devices for refractory glaucoma
    de Guzman, MHPU
    Valencia, A
    Farinelli, AC
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2006, 34 (02) : 102 - 107
  • [3] The Ex-PRESS glaucoma shunt versus trabeculectomy in open-angle glaucoma: a prospective randomized study
    de Jong, Leo A. M. S.
    [J]. ADVANCES IN THERAPY, 2009, 26 (03) : 336 - 345
  • [4] The Ahmed valve in refractory pediatric glaucoma
    Englert, JA
    Freedman, SF
    Cox, TA
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1999, 127 (01) : 34 - 42
  • [5] Friedman DS, 2004, ARCH OPHTHALMOL-CHIC, V122, P532
  • [6] Three-Year Follow-up of the Tube Versus Trabeculectomy Study
    Gedde, Steven J.
    Schiffman, Joyce C.
    Feuer, William J.
    Herndon, Leon W.
    Brandt, James D.
    Budenz, Donald L.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (05) : 670 - 684
  • [7] Is the Ex-PRESS Glaucoma Shunt Magnetic Resonance Imaging Safe?
    Geffen, Noa
    Trope, Graham E.
    Alasbali, Tariq
    Salonen, David
    Crowley, Adrian P.
    Buys, Yvonne M.
    [J]. JOURNAL OF GLAUCOMA, 2010, 19 (02) : 116 - 118
  • [8] Glaucoma drainage devices: A systematic literature review and current controversies
    Hong, CH
    Arosemena, A
    Zurakowski, D
    Ayyala, RS
    [J]. SURVEY OF OPHTHALMOLOGY, 2005, 50 (01) : 48 - 60
  • [9] Comparison of polypropylene and silicone Ahmed Glaucoma Valves
    Ishida, Kyoko
    Netland, Peter A.
    Costa, Vital P.
    Shiroma, Lineu
    Khan, Baseer
    Ahmed, Iqbal Ike K.
    [J]. OPHTHALMOLOGY, 2006, 113 (08) : 1320 - 1326
  • [10] Ex-PRESS Miniature Glaucoma Device Implanted Under a Scleral Flap Alone or Combined With Phacoemulsification Cataract Surgery
    Kanner, Elliott M.
    Netland, Peter A.
    Sarkisian, Steven R.
    Du, Haiming