AC Measurements and Multiplexed Detection Of Microparticles Using Parallel Channel Coulter Counter

被引:0
|
作者
Jagtiani, Ashish V. [1 ]
Zhe, Jiang [1 ]
机构
[1] Univ Akron, Dept Mech Engn, Akron, OH 44325 USA
来源
2010 18TH BIENNIAL UNIVERSITY/GOVERNMENT/INDUSTRY MICRO-NANO SYMPOSIUM | 2010年
关键词
Multichannel; Coulter-counter; particle detection; high-throughput; multiplexed detection; PARTICLES;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A micromachined 4-channel Coulter counting device is reported to achieve high throughput counting of microparticles. The design, fabrication and testing of this device is presented. Multiplexed detection was implemented by amplitude modulation of signals from individual channels to obtain a single combined response for all four channels, requiring only one set of detection electronics. Testing results using 30 mu m polystyrene particles demonstrated the throughput of the multiplexed device was improved 300% by simultaneously detecting particles through its parallel 4 channels. The multiplexed detection principle can be extended to a larger number of channels is to further improve the throughput.