The second international in-vivo monitoring intercomparison program for whole body counting facilities by Canadian and United States agencies

被引:6
作者
Kramer, GH
Loesch, RM
Olsen, PC
机构
[1] Radiat Protect Bur, Human Monitoring Lab, Ottawa, ON K1A 1C1, Canada
[2] US DOE, Off Hlth, Washington, DC 20585 USA
[3] Pacific NW Natl Lab, Richland, WA 99352 USA
来源
HEALTH PHYSICS | 2001年 / 80卷 / 03期
关键词
calibration; reference man; phantom; whole body counting;
D O I
10.1097/00004032-200103000-00003
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The Canadian National Calibration Reference Centre for In-Vivo Monitoring and the United States Department of Energy collaborated to offer a second international in vivo intercomparison program to whole body counting facilities in 1996. This program used a Reference Female phantom shell filled with radioactive tissue-substitute polyurethane to simulate a uniform fission-product distribution in soft tissues. The nuclides used were Cs-137 and Co-60. The phantom also contained K-40 homogeneously distributed in an amount similar to a Reference Female to produce a representative Compton background in the resulting spectra. Participants were asked to identify the nuclides and report activities for all except K-40. They were also asked to measure the precision of counting and supply the MDA for Cs-137 and Co-60. The bias results were in the range of -30% to +80% with most facilities falling inside the range of -25% to +50% (Canadian and U.S. acceptable performance criteria). Results indicated that there was no measurable size dependency for this phantom. All reported precisions were less than 5% but NaI detector based systems seemed to have a systematic uncertainty in addition to Poisson variability. Contrarily, this was not found for Ge detector based systems. MDA data was scattered (14-3,500 Bq for Cs-137 and 9-460 Bq for Co-60) and only suggested that lengthening the counting time improves MDA. Health Phys. 80(3):214-224; 2001.
引用
收藏
页码:214 / 224
页数:11
相关论文
共 6 条
  • [1] *ANSI, 1996, N13301996 ANSI HPS
  • [2] *AT EN CONTR BOARD, 1997, S107 ANSI AT EN CONT
  • [3] BERK KN, 1998, DATA ANAL MICROSOFT
  • [4] THE CANADIAN NATIONAL CALIBRATION REFERENCE CENTER FOR BIOASSAY AND IN-VIVO MONITORING - A PROGRAM SUMMARY
    KRAMER, GH
    ZAMORA, ML
    [J]. HEALTH PHYSICS, 1994, 67 (02): : 192 - 196
  • [5] THE BRMD BOMAB PHANTOM FAMILY
    KRAMER, GH
    BURNS, L
    NOEL, L
    [J]. HEALTH PHYSICS, 1991, 61 (06): : 895 - 902
  • [6] The 1993 intercomparison of the measurement of in vivo radioactivity
    Kramer, GH
    Loesch, RM
    Olsen, PC
    [J]. RADIATION PROTECTION DOSIMETRY, 1999, 86 (03) : 197 - 205