Multiplexed detection of cardiac biomarkers in serum with nanowire arrays using readout ASIC

被引:49
作者
Zhang, Guo-Jun [1 ,2 ]
Chai, Kevin Tshun Chuan [2 ]
Luo, Henry Zhan Hong [2 ]
Huang, Joon Min [2 ]
Tay, Ignatius Guang Kai [2 ]
Lim, Andy Eu-Jin [2 ]
Je, Minkyu [2 ]
机构
[1] Hubei Univ Chinese Med, Sch Lab Med, Wuhan 430065, Peoples R China
[2] ASTAR, Inst Microelect, Singapore 117685, Singapore
关键词
Silicon nanowire array; Biosensors; Cardiac biomarkers; Serum; Multiplexed detection; Application specific integrated circuit; SILICON NANOWIRE; ELECTRICAL DETECTION; DNA HYBRIDIZATION; INTERFACE; DIAGNOSIS; MARKERS;
D O I
10.1016/j.bios.2012.02.052
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Early detection of cardiac biomarkers for diagnosis of heart attack is the key to saving lives. Conventional method of detection like the enzyme-linked immunosorbent assay (ELISA) is time consuming and low in sensitivity. Here, we present a label-free detection system consisting of an array of silicon nanowire sensors and an interface readout application specific integrated circuit (ASIC). This system provides a rapid solution that is highly sensitive and is able to perform direct simultaneous-multiplexed detection of cardiac biomarkers in serum. Nanowire sensor arrays were demonstrated to have the required selectivity and sensitivity to perform multiplexed detection of 100 fg/ml troponin T, creatine kinase MM, and creatine kinase MB in serum. A good correlation between measurements from a probe station and the readout ASIC was obtained. Our detection system is expected to address the existing limitations in cardiac health management that are currently imposed by the conventional testing platform, and opens up possibilities in the development of a miniaturized device for point-of-care diagnostic applications. (C) 2012 Elsevier BM. All rights reserved.
引用
收藏
页码:218 / 223
页数:6
相关论文
共 28 条
[1]  
Bennati, 2009, ISSCC, P348
[2]   Quantitative real-time measurements of DNA hybridization with alkylated nonoxidized silicon nanowires in electrolyte solution [J].
Bunimovich, Yuri L. ;
Shin, Young Shik ;
Yeo, Woon-Seok ;
Amori, Michael ;
Kwong, Gabriel ;
Heath, James R. .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2006, 128 (50) :16323-16331
[3]   A 64-Channel Readout ASIC for Nanowire Biosensor Array with Electrical Calibration Scheme [J].
Chai, Kevin T. C. ;
Choe, Kunil ;
Bernal, Olivier D. ;
Gopalakrishnan, Pradeep K. ;
Zhang, Guo-Jun ;
Kang, Tae Goo ;
Je, Minkyu .
2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, :3491-3494
[4]   Label-Free Electrical Detection of Cardiac Biomarker with Complementary Metal-Oxide Semiconductor-Compatible Silicon Nanowire Sensor Arrays [J].
Chua, Jay Huiyi ;
Chee, Ru-Ern ;
Agarwal, Ajay ;
Wong, She Mein ;
Zhang, Guo-Jun .
ANALYTICAL CHEMISTRY, 2009, 81 (15) :6266-6271
[5]   Nanowire nanosensors for highly sensitive and selective detection of biological and chemical species [J].
Cui, Y ;
Wei, QQ ;
Park, HK ;
Lieber, CM .
SCIENCE, 2001, 293 (5533) :1289-1292
[6]  
Espostil C.S.D., 2006, ISSCC, P69
[7]   A low-cost interface to high-value resistive sensors varying over a wide range [J].
Flammini, A ;
Marioli, D ;
Taroni, A .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2004, 53 (04) :1052-1056
[8]   Silicon nanowire arrays for label-free detection of DNA [J].
Gao, Zhiqiang ;
Agarwal, Ajay ;
Trigg, Alastair D. ;
Singh, Navab ;
Fang, Cheng ;
Tung, Chih-Hang ;
Fan, Yi ;
Buddharaju, Kavitha D. ;
Kong, Jinming .
ANALYTICAL CHEMISTRY, 2007, 79 (09) :3291-3297
[9]   A 160 dB equivalent dynamic range auto-scaling interface for resistive gas sensors arrays [J].
Grassi, Marco ;
Malcovati, Piero ;
Baschirotto, Andrea .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2007, 42 (03) :518-528
[10]   Direct ultrasensitive electrical detection of DNA and DNA sequence variations using nanowire nanosensors [J].
Hahm, J ;
Lieber, CM .
NANO LETTERS, 2004, 4 (01) :51-54