Numerical and experimental investigation of kerf depth effect on high-frequency phased array transducer

被引:8
|
作者
Zhang, J. Y. [1 ,2 ,3 ]
Xu, W. J. [2 ]
Carlier, J. [2 ]
Ji, X. M. [1 ,3 ]
Queste, S. [4 ]
Nongaillard, B. [2 ]
Huang, Y. P. [1 ,3 ]
机构
[1] Fudan Univ, Dept Microelect, ASIC, Shanghai 200433, Peoples R China
[2] Univ Valenciennes, CNRS, UMR 8520, IEMN,Dept OAE, F-59313 Valenciennes, France
[3] Fudan Univ, Dept Microelect, Syst State Key Lab, Shanghai 200433, Peoples R China
[4] Univ Franche Comte, CNRS, UMR 6174, Inst FEMTO ST, F-25044 Besancon, France
基金
美国国家科学基金会;
关键词
High-frequency; Single crystal lithium niobate (LiNbO3); Linear arrays; Micromachining;
D O I
10.1016/j.ultras.2011.08.006
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Background: High-frequency ultrasonic transducer arrays are essential for high resolution imaging in clinical analysis and Non-Destructive Evaluation (NDE). However, the structure design and fabrication of the kerfed ultrasonic array is quite challenging when very high frequency (>= 100 MHz) is required. Objective and method: Here we investigate the effect of kerf depth on the performances of array transducers. A finite element tool, COMSOL, is employed to simulate the properties of acoustic field and to calculate the electrical properties of the arrays, including crosstalk effect and electrical impedance. Furthermore, Inductively Coupled Plasma (ICP) deep etching process is used to etch 36 degrees/Y-cut lithium niobate (LiNbO3) crystals and the limitation of etching aspect ratio is studied. Several arrays with different profiles are realized under optimized processes. At last, arrays with a pitch of 25 mu m and 40 mu m are fabricated and characterized by a network analyzer. Results: Kerf depth plays an important role in the performance of the transducer array. The crosstalk is proportional to kerf depth. When kerf depth is more than 13 mu m, the array with crosstalk less than -20 dB, which is acceptable for the real application, could provide a desired resolution. Compared to beam focusing, kerf depth exhibits more effect on the beam steering/focusing. The lateral pressure distribution is quantitatively summarized for four types of arrays with different kerf depth. The results of halfcut array are similar to those of the full-cut one in both cases of focusing and steering/focusing. The Full-Width-at-Half-Maximum (FWHM) is 55 mu m for the half-cut array, and is 42 mu m for the full-cut one. The 5-mu m-cut array, suffering from severe undesired lobes, demonstrates similar behaviors with the no-cut one. ICP process is used to etch the 36 degrees/Y-cut LiNbO3 film. The aspect ratio of etching profile increases with the kerf width decreasing till it stops by forming a V-shaped groove, and the positive tapered profile angle ranges between 62 degrees and 80 degrees. If the mask selectivity does not limit the process in terms of achievable depth, the aspect ratio is limited to values around 1.3. The measurement shows the electrical impedance and crosstalk are consistent with the numerical calculation. Conclusion: The numerical results indicate that half-cut array is a promising alternative for the fabrication of high-frequency ultrasonic linear arrays. In fact, the minimum pitch that could be obtained is around 25 mu m, equivalent to a pitch of 1.6 lambda, with a kerf depth of 16 mu m under the optimized ICP parameters. Crown Copyright (C) 2011 Published by Elsevier B. V. All rights reserved.
引用
收藏
页码:223 / 229
页数:7
相关论文
共 50 条
  • [1] Investigation of High-Frequency Guided Wave Modes Applicability in Defect Detection Using Phased Array Ultrasonic Transducer
    M. J. Ranjbar Naserabadi
    Russian Journal of Nondestructive Testing, 2021, 57 : 762 - 768
  • [2] Investigation of High-Frequency Guided Wave Modes Applicability in Defect Detection Using Phased Array Ultrasonic Transducer
    Naserabadi, M. J. Ranjbar
    Sodagar, Sina
    RUSSIAN JOURNAL OF NONDESTRUCTIVE TESTING, 2021, 57 (09) : 762 - 768
  • [3] Development of Broadband High-Frequency Piezoelectric Micromachined Ultrasonic Transducer Array
    Wang, Xu-Bo
    He, Le-Ming
    Ma, You-Cao
    Liu, Wen-Juan
    Xu, Wei-Jiang
    Ren, Jun-Yan
    Riaud, Antoine
    Zhou, Jia
    SENSORS, 2021, 21 (05) : 1 - 10
  • [4] An Ultrafast High-Frequency Hardware Beamformer for a Phased Array Endoscope
    Campbell, Nicholas A.
    Samson, Christopher A.
    Brown, Jeremy A.
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1505 - 1508
  • [5] Prototyping of a High Frequency Phased Array Ultrasound Transducer on a Piezoelectric Thick Film
    Buncic, Nikola
    Kabasi, Anton
    Cakaric, Tonci
    Bilusic, Ante
    INFORMACIJE MIDEM-JOURNAL OF MICROELECTRONICS ELECTRONIC COMPONENTS AND MATERIALS, 2019, 49 (01): : 3 - 9
  • [6] High-Frequency Underwater Transducer
    Wang, G.
    Wang, L. K.
    Qin, L.
    FERROELECTRICS, 2010, 408 : 129 - 136
  • [7] Design and Preliminary Experimental Results for a High-Frequency Crossed Electrode Phased Array, Based on a Reconfigurable Fresnel Lens
    Latham, Katherine
    Samson, Christopher A.
    Bezanson, Andre
    Adamson, Rob
    Brown, Jeremy A.
    2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [8] A High-Frequency Phased Array System for Transcranial Ultrasound Delivery in Small Animals
    Rahimi, Saba
    Jones, Ryan Matthew
    Hynynen, Kullervo
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (01) : 127 - 135
  • [9] Design and Fabrication of a High-Frequency Microconvex Array Transducer or Small Animals Imaging
    Chen, Weicen
    Zhang, Qi
    Liu, Jiamei
    Lei, Shuang
    Li, Yongchuan
    Huang, Jiqing
    Guo, Lunhuang
    Zheng, Hairong
    Wu, Dawei
    Ma, Teng
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2022, 69 (06) : 1943 - 1951
  • [10] The Study of Cable Effect on High-Frequency Ultrasound Transducer Performance
    Jian, Xiaohua
    Li, Zhangjian
    Han, Zhile
    Xu, Jie
    Liu, Pengbo
    Liu, Yufei
    Cui, Yaoyao
    Huang, Wenbin
    IEEE SENSORS JOURNAL, 2018, 18 (13) : 5265 - 5271