An exploration of the advantages of automated titration testing: low inter-instrument variability and equivalent accuracy for ABO and non-ABO antibody titres relative to tube testing

被引:10
作者
Adkins, Brian D. [1 ]
Egloff, Shanna A. Arnold [1 ,2 ]
Fahey-Ahrndt, Kayla [3 ]
Kjell, Andrea L. [3 ]
Cohn, Claudia S. [4 ]
Young, Pampee P. [1 ,5 ]
机构
[1] Vanderbilt Univ, Med Ctr, Dept Pathol Microbiol & Immunol, Nashville, TN USA
[2] Tennessee Valley Healthcare Syst, Dept Vet Affairs, Nashville, TN USA
[3] Fairview Hosp, M Hlth, Blood Bank Lab, Minneapolis, MN USA
[4] Univ Minnesota, Dept Lab Med & Pathol, Minneapolis, MN 55455 USA
[5] Amer Red Cross, Biomed Serv, Washington, DC 20006 USA
关键词
immunohaematology; RBC antigens and antibodies; serological testing; transfusion medicine (in general); transplantation; O WHOLE-BLOOD; TRAUMA; PLASMA;
D O I
10.1111/vox.12893
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and objectives Obtaining IgM and IgG titres is important in numerous clinical situations, including solid-organ transplant, obstetrics, and for testing of out-of-group plasma-containing components. Tube method is the most prevalent testing modality, though it is both labour-intensive and known for intra- and inter-laboratory variability. The utility of automated gel testing as a method to improve both inter- and intra-laboratory reproducibility is unknown. Materials and methods Two academic centres participated in a study evaluating automated gel titreing. Group O plasma samples were used to measure titres of antibodies against ABO (IgM) with buffered gel cards and 4 minor and minor red-blood-cell antigens (IgG) anti-IgG gel cards. Multiple ORTHO VISION automated analyzers were used to assess inter-instrument variation. A subset of ABO (IgM) samples were compared between laboratories to evaluate inter-laboratory variability. Multiple samples were titred by tube and by automated gel technology to determine similarity of results. Results Testing demonstrated no significant difference between analysers or between sites when performing automated titrations (P >= 0 center dot 99). Non-ABO IgG titres were evaluated and demonstrated little inter-instrument variability. The IgM anti-A and -B titres obtained by automated gel testing were neither consistently higher nor lower than tube titres. Greater than 90% of titre values were within one dilution. Conclusion Based on this study, our data suggest that titreing by automated gel testing is both highly reproducible (IgM and IgG) and does not differ significantly from manual tube testing results of direct agglutination (IgM).
引用
收藏
页码:314 / 322
页数:9
相关论文
共 22 条
[1]  
AuBuchon JP, 2008, ARCH PATHOL LAB MED, V132, P1194, DOI 10.1043/1543-2165(2008)132[1194:RTVIPO]2.0.CO
[2]  
2
[3]   Impact of Uniform Methods on Interlaboratory Antibody Titration Variability Antibody Titration and Uniform Methods [J].
Bachegowda, Lohith S. ;
Cheng, Yan H. ;
Long, Thomas ;
Shaz, Beth H. .
ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2017, 141 (01) :131-138
[4]  
Bajpai Meenu, 2012, Asian J Transfus Sci, V6, P140, DOI 10.4103/0973-6247.98914
[5]   An evaluation of methods for producing low-titer group O whole blood to support military trauma resuscitation [J].
Belin, Tamara R. ;
Yazer, Mark H. ;
Meledeo, Michael A. ;
Fahie, Roland L. ;
Malloy, Wilbur W. ;
Stubbs, James R. ;
Cap, Andrew P. .
JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2017, 82 (06) :S79-S86
[6]  
Bhangale Amit, 2017, Asian J Transfus Sci, V11, P131, DOI 10.4103/0973-6247.214343
[7]   Clinical Guide to ASO-Incompatible Allogeneic Stem Cell Transplantation [J].
Booth, Garrett S. ;
Gehrie, Eric A. ;
Bolan, Charles D. ;
Savani, Bipin N. .
BIOLOGY OF BLOOD AND MARROW TRANSPLANTATION, 2013, 19 (08) :1152-1158
[8]  
Casina TS, 2015, TRANSFUSION, V55, p3A
[9]   Anti-A and anti-B titers in pooled group O platelets are comparable to apheresis platelets [J].
Cooling, Laura L. ;
Downs, Theresa A. ;
Butch, Suzanne H. ;
Davenport, Robertson D. .
TRANSFUSION, 2008, 48 (10) :2106-2113
[10]   Safety of the use of group A plasma in trauma: the STAT study [J].
Dunbar, Nancy M. ;
Yazer, Mark H. .
TRANSFUSION, 2017, 57 (08) :1879-1884