Phase contrast X-ray imaging of large samples using an incoherent laboratory source

被引:36
作者
Kottler, C. [1 ]
Pfeiffer, F.
Bunk, O.
Gruenzweig, C.
Bruder, J.
Kaufmann, R.
Tlustos, L.
Walt, H.
Briod, I.
Weitkamp, T.
David, C.
机构
[1] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
[2] Ctr Suisse Elect Microtech SA, CH-8048 Zurich, Switzerland
[3] CERN, CERN PH, CH-1211 Geneva, Switzerland
[4] Forschungszentrum Karlsruhe, Inst Synchrotonstrahlung, D-76021 Karlsruhe, Germany
来源
PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE | 2007年 / 204卷 / 08期
关键词
D O I
10.1002/pssa.200675659
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
An interferometric method to record quantitative X-ray phase contrast images has been developed that can be used at polychromatic and incoherent X-ray sources such as laboratory tubes. With respect to previously presented results, in this work we report on recent developments and results that have been achieved in view of potential future applications such as in medicine or biology. In particular, due to improvements in the fabrication process large area diffraction gratings with high aspect ratio were achieved. Thereby, the field of view of the interterometer has been drastically increased to 64 x 64 mm(2) and the design value of the photon energy for the gratings could be increased up to 28 keV. Moreover, the use of a Medipix2 single photon-counting pixel detector shows a considerable improvement in image quality and sensitivity over the integrating detector used so far. (c) 2007 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.
引用
收藏
页码:2728 / 2733
页数:6
相关论文
共 7 条
[1]   Differential x-ray phase contrast imaging using a shearing interferometer [J].
David, C ;
Nöhammer, B ;
Solak, HH ;
Ziegler, E .
APPLIED PHYSICS LETTERS, 2002, 81 (17) :3287-3289
[2]  
DAVID C, 2006, UNPUB MICROELECTRON
[3]   Phase-sensitive x-ray imaging [J].
Fitzgerald, R .
PHYSICS TODAY, 2000, 53 (07) :23-26
[4]   Medipix2:: a 64-k pixel readout chip with 55 μm square elements working in single photon counting mode [J].
Llopart, X ;
Campbell, M ;
Dinapoli, R ;
Segundo, DS ;
Pemigotti, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) :2279-2283
[5]   Phase-sensitive imaging and phase tomography using X-ray interferometers [J].
Momose, A .
OPTICS EXPRESS, 2003, 11 (19) :2303-2314
[6]   Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources [J].
Pfeiffer, F ;
Weitkamp, T ;
Bunk, O ;
David, C .
NATURE PHYSICS, 2006, 2 (04) :258-261
[7]   X-ray phase imaging with a grating interferometer [J].
Weitkamp, T ;
Diaz, A ;
David, C ;
Pfeiffer, F ;
Stampanoni, M ;
Cloetens, P ;
Ziegler, E .
OPTICS EXPRESS, 2005, 13 (16) :6296-6304