MIT 1.3-GHz LTS/HTS NMR Magnet: Post Quench Analysis and New 800-MHz Insert Design

被引:43
作者
Park, Dongkeun [1 ]
Bascunan, Juan [1 ]
Michael, Philip C. [1 ]
Lee, Jiho [1 ]
Choi, Yoon Hyuck [1 ]
Li, Yi [1 ]
Hahn, Seungyong [2 ]
Iwasa, Yukikazu [1 ]
机构
[1] MIT, Plasma Sci & Fus Ctr, Francis Bitter Magnet Lab, 77 Massachusetts Ave, Cambridge, MA 02139 USA
[2] Seoul Natl Univ, Dept Elect & Comp Engn, Seoul 08826, South Korea
基金
美国国家卫生研究院;
关键词
High-temperature superconductors; nuclear magnetic resonance; quench; REBCO; superconducting magnet;
D O I
10.1109/TASC.2019.2901026
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present post-quench analyses of the MIT 800MHz REBCO insert magnet (H800), unexpectedly quenched during operation in March 2018, and design study of a new 800-MHz HTS insert (H800N). The as-wound H800 was supposed to contribute 18.7 T and, with an LTS background magnet (L500), produce 30.5 T corresponding to a proton resonance frequency of 1.3 GHz. The H800 was operated at 4.2 K in liquid helium and, about 5 min after the power supply reached a target operating current of 251.3 A, it experienced a quench. Because the damage in the H800 was more widespread than it first appeared, we decided to design and build a new insert magnet, H800N. In designing H800N, we tried to eliminate unanticipated flaws in our H800 design. H800N has been designed to be more stable and reliable against quench without permanent damage by: 1) adopting a single solenoid structure composed of 40 stacked double pancake coils with improved cross-over sections; 2) enhancing thermal stability; and 3) reducing excessive current margin for quench protection.
引用
收藏
页数:4
相关论文
共 7 条
[1]  
Gupta R., 25 INT C MAG TECHN
[2]  
Leupold M. J., 1978, 6th International Conference on Magnet Technology, P400
[3]   Assembly and Test of a 3-Nested-Coil 800-MHz REBCO Insert (H800) for the MIT 1.3 GHz LTS/HTS NMR Magnet [J].
Michael, Philip C. ;
Park, Dongkeun ;
Choi, Yoon Hyuck ;
Lee, Jiho ;
Li, Yi ;
Bascunan, Juan ;
Noguchi, So ;
Hahn, Seungyong ;
Iwasa, Yukikazu .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2019, 29 (05)
[4]   HTS Shim Coils Energized by a Flux Pump for the MIT 1.3-GHz LTS/HTS NMR Magnet: Design, Construction, and Results of a Proof-of-Concept Prototype [J].
Park, Dongkeun ;
Lee, Jiho ;
Bascunan, Juan ;
Michael, Philip C. ;
Iwasa, Yukikazu .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2018, 28 (03)
[5]   Design of a Tabletop Liquid-Helium-Free 23.5-T Magnet Prototype Toward 1-GHz Microcoil NMR [J].
Park, Dongkeun ;
Choi, Yoon Hyuck ;
Iwasa, Yukikazu .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2019, 29 (05)
[6]   Construction and Test Results of Coils 2 and 3 of a 3-Nested-Coil 800-MHz REBCO Insert for the MIT 1.3-GHz LTS/HTS NMR Magnet [J].
Park, Dongkeun ;
Bascunan, Juan ;
Michael, Philip C. ;
Lee, Jiho ;
Hahn, Seungyong ;
Iwasa, Yukikazu .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2018, 28 (03)
[7]   Overall Thermal Conductance and Thermal Contact Resistance in No-Insulation REBCO Magnet [J].
Seok, Jihoon ;
Kim, Dongmin ;
Kim, Seokho .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2018, 28 (03)