Variable sac behavior after endovascular repair of abdominal aortic aneurysm: Analysis of core laboratory data

被引:56
作者
Greenberg, RK
Deaton, D
Sullivan, T
Walker, E
Lyden, SP
Srivastava, SD
Ouriel, K
Ivanc, T
Burton, T
Mayo, J
机构
[1] Cleveland Clin Fdn, Dept Vasc Surg, Cleveland, OH 44195 USA
[2] Georgetown Univ Hosp, Washington, DC 20007 USA
[3] Mayo Clin & Mayo Fdn, Rochester, MN 55905 USA
[4] Univ Tennessee, Dept Stat, Knoxville, TN 37996 USA
[5] Cleveland Clin Fdn, Dept Biostat & Epidemiol, Cleveland, OH 44195 USA
关键词
D O I
10.1016/j.jvs.2003.08.009
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: The behavior of the aneurysm sac after endovascular grafting has been the subject of significant speculation. The importance of sac behavior is manifested by the correlation between aneurysm size or size change and risk for rupture, and potentially further extrapolated to define the need for secondary intervention. This study was undertaken to define graft-specific differences and the effect of endoleak on sac remodeling. Methods: Core laboratory data were obtained for three US Phase 11 clinical trials. Patients were included if they met anatomic inclusion criteria and underwent placement of the latest version of a bifurcated endovascular prosthesis. Unsupported Dacron (Ancure), supported Dacron (Zenith), and expanded polytetrafluoroethylene (Excluder) grafts were evaluated. Digitized images were electronically assessed for aneurysm size (area, maximum, minimum diameter) with National Institutes of Health Image software. Two blinded reviewers analyzed each radiographic study to ensure accurate image selection and establish the presence or absence of endoleak. A third reviewer adjudicated discrepancies. X 2 analysis and mixed nonlinear modeling were used to analyze the results. Results: Of 1506 patients evaluated, 723 (227 Ancure, 343 Excluder, 153 Zenith) met inclusion criteria for the study. Mean follow-up was 23.2 months (Ancure, 31.3 months; Excluder, 19.6 months; Zenith, 19.3 months). The incidence of any endoleak was 39.1% (Ancure, 58.1%; Excluder, 34.7%; Zenith, 20.9%; P < .001). Type of prosthesis, presence or absence of endoleak, and baseline size were determinants of rate of aneurysm shrinkage. Reduction in sac size was greatest with the Zenith graft, followed by the Ancure and Excluder grafts. Presence of endoleak had a moderating effect on rate of sac shrinkage with the Zenith and Ancure grafts; however, sac size increased in the presence of endoleak with the Excluder graft. Finally, baseline size was positively correlated with rate of aneurysm shrinkage. Conclusions: The behavior of the aneurysm sac depends on the type of prosthesis, presence or absence of endoleak, and baseline size of the sac. Differential sac behavior must be considered when determining the need for secondary interventions, timing follow-up studies, and assessing success or failure of endovascular repair.
引用
收藏
页码:95 / 101
页数:7
相关论文
共 31 条
[1]   Thrombus distribution and changes in aneurysm size following endovascular aortic aneurysm repair [J].
Armon, MP ;
Yusuf, SW ;
Whitaker, SC ;
Gregson, RHS ;
Wenham, PW ;
Hopkinson, BR .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1998, 16 (06) :472-476
[2]   Abdominal aortic aneurysm size regression after endovascular repair is endograft dependent [J].
Bertges, DJ ;
Chow, K ;
Wyers, MC ;
Landsittel, D ;
Frydrych, AV ;
Stavropoulos, W ;
Tan, WA ;
Rhee, RY ;
Fillinger, MF ;
Fairman, RM ;
Makaroun, MS .
JOURNAL OF VASCULAR SURGERY, 2003, 37 (04) :716-723
[3]   Causes and outcomes of open conversion and aneurysm rupture after endovascular abdominal aortic aneurysm repair: Can type II endoleaks be dangerous? [J].
Buth, J ;
Harris, PL ;
van Marrewijk, C .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2002, 194 (01) :S98-S102
[4]   Identifying and grading factors that modify the outcome of endovascular aortic aneurysm repair [J].
Chaikof, EL ;
Fillinger, MF ;
Matsumura, JS ;
Rutherford, RB ;
White, GH ;
Blankensteijn, JD ;
Bernhard, VM ;
Harris, PL ;
Kent, KC ;
May, J ;
Veith, FJ ;
Zarins, CK .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1061-1066
[5]   Reporting standards for endovascular aortic aneurysm repair [J].
Chaikof, EL ;
Blankensteijn, JD ;
Harris, PL ;
White, GH ;
Zarins, CK ;
Bernhard, VM ;
Matsumura, JS ;
May, J ;
Veith, FJ ;
Fillinger, MF ;
Rutherford, RB ;
Kent, KC .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1048-1060
[6]   Endoleak: Predictive value for aneurysm growth at 3 years [J].
Deaton, DH ;
Makaroun, MS ;
Fairman, RM .
ANNALS OF VASCULAR SURGERY, 2002, 16 (01) :37-42
[7]   Endovascular repair of abdominal aortic aneurysm with the ancure endograft: CT follow-up of perigraft flow and aneurysm size at 6 months [J].
Franco, TJ ;
Zajko, AB ;
Federle, MP ;
Makaroun, MS .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (04) :429-435
[8]   An update of the Zenith endovascular graft for abdominal aortic aneurysms: Initial implantation and mid-term follow-up data [J].
Greenberg, RK ;
Lawrence-Brown, M ;
Bhandari, G ;
Hartley, D ;
Stelter, W ;
Umscheid, T ;
Chuter, T ;
Ivancev, K ;
Green, R ;
Hopkinson, B ;
Semmens, J ;
Ouriel, K .
JOURNAL OF VASCULAR SURGERY, 2001, 33 (02) :S157-S164
[9]  
HARRIS P, 1999, J ENDOVASC SURG
[10]   Incidence and risk factors of late rupture, conversion, and death after endovascular repair of infrarenal aortic aneurysms: The EUROSTAR experience [J].
Harris, PL ;
Vallabhaneni, SR ;
Desgranges, P ;
Bacquemin, JP ;
van Marrewijk, C ;
Laheij, RJF .
JOURNAL OF VASCULAR SURGERY, 2000, 32 (04) :739-749