Key parameters evaluation for hip prosthesis with finite element analysis

被引:0
作者
Guo, Hongqiang [1 ]
Li, Dichen [1 ]
Lian, Qin [1 ]
Li, Xiang [1 ]
Jin, Zhongmin [2 ]
机构
[1] Xi An Jiao Tong Univ, State Key Lab Mfg Syst Engn, Xian 710049, Peoples R China
[2] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, West Yorkshire, England
来源
NUMERICAL ANALYSIS AND APPLIED MATHEMATICS | 2007年 / 936卷
关键词
flip prosthesis; finite element analysis; fatigue analysis;
D O I
暂无
中图分类号
O29 [应用数学];
学科分类号
070104 ;
摘要
Stein length and cross section are two key parameters that influence the stability and longevity of metallic hip prosthesis in the total hip arthroplasty (THA). In order to assess their influence to the stress and fatigue behavior of hip prosthesis, a series model of hip prosthesis with round-shaped or drum-shaped cross section, and with different stem lengths were created. These models were analyzed under both static and dynamic loading conditions with finite element analysis, and dynamic loading represents normal walking was used in the dynamic analysis. The stress on the metallic stem, cement and adjacent bone were got, micromotion on the cement-metal interface were got too. Safety factors for fatigue life of the hip prothesis were calculated based on data obtained from dynamic analysis. Static analysis shows that drum-shaped cross section can decrease the displacement of the stem, that stress on drum-shaped stem focus on the corner of the femoral neck and the distal part of hip prosthesis, whereas the stress on the round-shaped stem distributes evenly over most part of the stem, and maximum stress on stem prosthesis fluctuates with stem length bottoming out at stem length range from 80 mm to 110 mm, that drum-shaped stems with drum height 8 mm generate more stress at the distal part of stem than drum-shaped stems with drum height 10 mm and round stems do. Dynamic and fatigue analysis shows that drum-shaped stem with drum height 10 mm and stem length 90 mm has the greatest safety factor therefore long fatigue life.
引用
收藏
页码:248 / +
页数:2
相关论文
共 17 条
  • [1] HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS
    BERGMANN, G
    GRAICHEN, F
    ROHLMANN, A
    [J]. JOURNAL OF BIOMECHANICS, 1993, 26 (08) : 969 - 990
  • [2] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [3] Comparative in vitro study on the long term performance of cemented hip stems: validation of a protocol to discriminate between "good" and "bad" designs
    Cristofolini, L
    Teutonico, AS
    Monti, L
    Cappello, A
    Toni, A
    [J]. JOURNAL OF BIOMECHANICS, 2003, 36 (11) : 1603 - 1615
  • [4] Changes in strain distribution of loaded proximal femora caused by different types of cementless femoral stems
    Decking, Ralf
    Puhl, Wolfhart
    Simon, Ulrich
    Claes, Lutz E.
    [J]. CLINICAL BIOMECHANICS, 2006, 21 (05) : 495 - 501
  • [5] Finite element simulation of the hip joint during stumbling: a comparison between static and dynamic loading
    El'Sheikh, HF
    MacDonald, BJ
    Hashmi, MSJ
    [J]. JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2003, 143 : 249 - 255
  • [6] Primary stability in cementless femoral stems:: custom-made versus conventional femoral prosthesis
    Götze, C
    Steens, W
    Vieth, V
    Poremba, C
    Claes, L
    Steinbeck, J
    [J]. CLINICAL BIOMECHANICS, 2002, 17 (04) : 267 - 273
  • [7] Differences in stability and bone remodeling between a customized uncemented hydroxyapatite coated and a standard cemented femoral stem - A randomized study with use of radio stereometry and bone densitometry
    Grant, P
    Aamodt, A
    Falch, JA
    Nordsletten, L
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2005, 23 (06) : 1280 - 1285
  • [8] Analysis of a femoral hip prosthesis designed to reduce stress shielding
    Joshi, MG
    Advani, SG
    Miller, F
    Santare, MH
    [J]. JOURNAL OF BIOMECHANICS, 2000, 33 (12) : 1655 - 1662
  • [9] JOSHI MG, 2000, ASME, V53, P1
  • [10] Effects of loading conditions and objective function on three-dimensional shape optimization of femoral components of hip endoprostheses
    Katoozian, H
    Davy, DT
    [J]. MEDICAL ENGINEERING & PHYSICS, 2000, 22 (04) : 243 - 251