Laser and Electron Beam Additive Manufacturing Methods of Fabricating Titanium Bone Implants

被引:204
作者
Wysocki, Bartlomiej [1 ,2 ]
Maj, Piotr [1 ]
Sitek, Ryszard [1 ]
Buhagiar, Joseph [3 ]
Kurzydlowski, Krzysztof Jan [1 ]
Swieszkowski, Wojciech [1 ]
机构
[1] Warsaw Univ Technol, Fac Mat Sci & Engn, PL-02507 Warsaw, Poland
[2] Materialscare LLC, PL-15333 Bialystok, Poland
[3] Univ Malta, Dept Met & Mat Engn, Fac Engn, MSD-2080 Msida, Malta
来源
APPLIED SCIENCES-BASEL | 2017年 / 7卷 / 07期
关键词
SLM; EBM; Ti-6Al-4V; electron beam melting; selective laser melting; microstructure; MECHANICAL-PROPERTIES; TI-6AL-4V; MICROSTRUCTURE; BEHAVIOR; STRENGTH; ALLOYS; PROTOTYPES; ANISOTROPY; SCAFFOLDS; POROSITY;
D O I
10.3390/app7070657
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Additive Manufacturing (AM) methods are generally used to produce an early sample or near net-shape elements based on three-dimensional geometrical modules. To date, publications on AM of metal implants have mainly focused on knee and hip replacements or bone scaffolds for tissue engineering. The direct fabrication of metallic implants can be achieved by methods, such as Selective Laser Melting (SLM) or Electron Beam Melting (EBM). This work compares the SLM and EBM methods used in the fabrication of titanium bone implants by analyzing the microstructure, mechanical properties and cytotoxicity. The SLM process was conducted in an environmental chamber using 0.4-0.6 vol % of oxygen to enhance the mechanical properties of a Ti-6Al-4V alloy. SLM processed material had high anisotropy of mechanical properties and superior UTS (1246-1421 MPa) when compared to the EBM (972-976 MPa) and the wrought material (933-942 MPa). The microstructure and phase composition depended on the used fabrication method. The AM methods caused the formation of long epitaxial grains of the prior beta phase. The equilibrium phases (alpha + beta) and non-equilibrium alpha' martensite was obtained after EBM and SLM, respectively. Although it was found that the heat transfer that occurs during the layer by layer generation of the component caused aluminum content deviations, neither methods generated any cytotoxic effects. Furthermore, in contrast to SLM, the EBM fabricated material met the ASTMF136 standard for surgical implant applications.
引用
收藏
页数:20
相关论文
共 91 条
  • [61] Sailer R., 1993, TITANIUM HCP SPACE G
  • [62] Microstructure and mechanical behavior of porous Ti-6Al-4V parts obtained by selective laser melting
    Sallica-Leva, E.
    Jardini, A. L.
    Fogagnolo, J. B.
    [J]. JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2013, 26 : 98 - 108
  • [63] INFLUENCE OF PORE DIMENSIONS ON BONE INGROWTH INTO POROUS HYDROXYLAPATITE BLOCKS USED AS BONE-GRAFT SUBSTITUTES - A HISTOMETRIC STUDY
    SCHLIEPHAKE, H
    NEUKAM, FW
    KLOSA, D
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1991, 20 (01) : 52 - 58
  • [64] Overview of Materials Qualification Needs for Metal Additive Manufacturing
    Seifi, Mohsen
    Salem, Ayman
    Beuth, Jack
    Harrysson, Ola
    Lewandowski, John J.
    [J]. JOM, 2016, 68 (03) : 747 - 764
  • [65] Shishkovsky IV, 2011, POWDER METALL MET C+, V50, P275
  • [66] Biocompatibility of Advanced Manufactured Titanium Implants-A Review
    Sidambe, Alfred T.
    [J]. MATERIALS, 2014, 7 (12) : 8168 - 8188
  • [67] Effect of the build orientation on the mechanical properties and fracture modes of SLM Ti-6A1-4V
    Simonelli, M.
    Tse, Y. Y.
    Tuck, C.
    [J]. MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 2014, 616 : 1 - 11
  • [68] Dimensional accuracy of Electron Beam Melting (EBM) additive manufacture with regard to weight optimized truss structures
    Smith, C. J.
    Derguti, F.
    Nava, E. Hernandez
    Thomas, M.
    Tammas-Williams, S.
    Culizia, S.
    Fraser, D.
    Todd, I.
    [J]. JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2016, 229 : 128 - 138
  • [69] Process parameter selection for selective laser melting of Ti6Al4V based on temperature distribution simulation and experimental sintering
    Song, Bo
    Dong, Shujuan
    Liao, Hanlin
    Coddet, Christian
    [J]. INTERNATIONAL JOURNAL OF ADVANCED MANUFACTURING TECHNOLOGY, 2012, 61 (9-12) : 967 - 974
  • [70] Fatigue Performance of Medical Ti6Al4V Alloy after Mechanical Surface Treatments
    Sonntag, Robert
    Reinders, Joern
    Gibmeier, Jens
    Kretzer, J. Philippe
    [J]. PLOS ONE, 2015, 10 (03):