Development of rotational digital angiography system - clinical value in acute pulmonary thromboembolism

被引:4
作者
Tajima, H
Kumazaki, T
Kawamata, H
Ichikawa, K
Takano, T
机构
[1] Nippon Med Sch, Dept Radiol, Bunkyo Ku, Tokyo 1138603, Japan
[2] Nippon Med Sch, Dept Internal Med, Bunkyo Ku, Tokyo 1138603, Japan
关键词
angiography; embolism; pulmonary; thrombolysis; interventional procedures; three dimension (3-D);
D O I
10.1016/S0169-2607(01)00144-4
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A rotational digital angiography system was newly developed by authors. During the rotation, a pulse exposure is done at every 1.25 degrees by computer control, so that 288 different projectional images can be acquired following a single injection of contrast medium. Because rotational image data can be easily reconstructed to 3-D images, more useful diagnostic informations are obtained. Rotational digital angiography was performed for 15 patients with acute pulmonary thromboembolism, and massive thromboemboli were revealed in all patients. The diagnostic usefulness of this new system has also proven to be improved in the interventional procedures. At the appropriate angle, the thrombosed vessels were clearly distinguished. Thus, as the tip of the catheter was advanced to the target site, the thrombosed lesion could be exclusively treated by interventional techniques. The device has, therefore, been found to be of great value in the diagnosis and intervention of acute pulmonary thromboembolism. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:111 / 114
页数:4
相关论文
共 7 条
[1]   Possible pulmonary embolus: Evaluation with digital subtraction versus cut-film angiography - Prospect study in 80 patients [J].
Johnson, MS ;
Stine, SB ;
Shah, H ;
Harris, VJ ;
Ambrosius, WT ;
Trerotola, SO .
RADIOLOGY, 1998, 207 (01) :131-138
[2]   Update on diagnostic strategies of pulmonary embolism [J].
Kauczor, HU ;
Heussel, CP ;
Thelen, M .
EUROPEAN RADIOLOGY, 1999, 9 (02) :262-275
[3]   Development of rotational digital angiography and new cone-beam 3D image - Clinical value in vascular lesions [J].
Kumazaki, T .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1998, 57 (1-2) :139-142
[4]   Validity of pulmonary cine arteriography for the diagnosis of pulmonary embolism [J].
Nilsson, T ;
Turén, J ;
Billström, Å ;
Måre, K ;
Carlsson, A ;
Nyman, U .
EUROPEAN RADIOLOGY, 1999, 9 (02) :276-280
[5]  
TAJIMA H, 1990, Nippon Acta Radiologica, V50, P1608
[6]  
Tajima H, 1996, Radiat Med, V14, P49
[7]  
TAJIMA H, UNPUB NIPPON ACTA RA