CHARACTERIZATION OF A SMALL FOV PORTABLE GC: MEDIPROBE

被引:3
作者
Acampa, Wanda [1 ]
Capussela, Tiziana [2 ]
Cuocolo, Alberto [1 ]
Di Lillo, Francesca [3 ]
Punzo, Giorgio [4 ]
Quarto, Maria [1 ]
Roberti, Giuseppe [1 ]
Russo, Paolo [3 ]
Serra, Marcello [1 ]
机构
[1] Univ Napoli Federico II, Dipartimento Sci Biomed Avanzate, Via Pansini 5, I-80131 Naples, Italy
[2] AORN Osped Cardarelli, Via Antonio Cardarelli 9, I-80131 Naples, Italy
[3] Univ Napoli Federico II, Dipartimento Fis E Pancini, Via Cinthia, I-80126 Naples, Italy
[4] CNR, Ist Biostrutture & Bioimmagini, Via Tommaso De Amicis 95, I-80145 Naples, Italy
关键词
D O I
10.1093/rpd/ncy229
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In this work, we evaluated the performance of the prototype SFOV MediPROBE developed at the University of Naples Federico II through the protocol proposed by (Bhatia B.S., Bugby S.L., Lees J.E., Perkins A.C. A scheme for assessing the performance characteristics of a small field of-view gamma cameras. Physica. Medica., 31 (1), pp. 98-103. (2015) doi: 10.1016/j.ejmp.2014.08.004). We extensively investigated a new device configuration where the pinhole collimator was placed outside the housing of the probe, in order to increase the system spatial resolution, and the pixel size was doubled, in order to reduce the charge sharing effect. The experimental measurements show that the spatial resolution is enhanced by only about 10%, but the sensitivity decreases strongly. Therefore, the trade-off between these two features does not seem to be advantageous. In addition, our experiments suggest that the charge sharing effect is not completely canceled. Despite these results, the features of this device appear suitable for intraoperative surgical survey. We aim to use this device in the clinical practice for the intraoperative imaging of lymph nodes, breast, thyroid and parathyroid tumors.
引用
收藏
页码:290 / 296
页数:7
相关论文
共 10 条
[1]   Performance evaluation of a hand-held, semiconductor (CdZnTe)-based gamma camera [J].
Abe, A ;
Takahashi, N ;
Lee, J ;
Oka, T ;
Shizukuishi, K ;
Kikuchi, T ;
Inoue, T ;
Jimbo, M ;
Ryuo, H ;
Bickel, C .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (06) :805-811
[2]   A scheme for assessing the performance characteristics of small field-of-view gamma cameras [J].
Bhatia, B. S. ;
Bugby, S. L. ;
Lees, J. E. ;
Perkins, A. C. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (01) :98-103
[3]  
Burak W E Jr, 1996, Surg Technol Int, V5, P259
[4]  
Chapman CS, 2007, J EYE MOVEMENT RES, V1
[5]   Nuclear Probes and Intraoperative Gamma Cameras [J].
Heller, Sherman ;
Zanzonico, Pat .
SEMINARS IN NUCLEAR MEDICINE, 2011, 41 (03) :166-181
[6]   SiC X-ray detectors for harsh environments [J].
Lees, J. E. ;
Barnett, A. M. ;
Bassford, D. J. ;
Stevens, R. C. ;
Horsfall, A. B. .
JOURNAL OF INSTRUMENTATION, 2011, 6
[7]   Resolution- versus sensitivity-effective diameter in pinhole collimation: experimental verification [J].
Metzler, SD ;
Accorsi, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21) :5005-5017
[8]   Direct charge sharing observation in single-photon-counting pixel detector [J].
Pellegrini, G. ;
Maiorino, M. ;
Blanchot, G. ;
Chmeissani, M. ;
Garcia, J. ;
Lozano, M. ;
Martinez, R. ;
Puigdengoles, C. ;
Ullan, M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 573 (1-2) :137-140
[9]   Evaluation of a CdTe semiconductor based compact gamma camera for sentinel lymph node imaging [J].
Russo, Paolo ;
Curion, Assunta S. ;
Mettivier, Giovanni ;
Esposito, Michela ;
Aurilio, Michela ;
Caraco, Corradina ;
Aloj, Luigi ;
Lastoria, Secondo .
MEDICAL PHYSICS, 2011, 38 (03) :1547-1560
[10]   Intraoperative gamma cameras for radioguided surgery: Technical characteristics, performance parameters, and clinical applications [J].
Tsuchimochi, Makoto ;
Hayama, Kazuhide .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (02) :126-138