Comparison of Enzyme Immunoassays for Detection of Antibodies to Hepatitis D Virus in Serum

被引:14
|
作者
Chow, Siu-Kei [1 ,5 ]
Atienza, Ederlyn E. [1 ]
Cook, Linda [1 ]
Prince, Harry [2 ,6 ]
Slev, Patricia [3 ]
Lape-Nixon, Mary [2 ]
Jerome, Keith R. [1 ,4 ]
机构
[1] Univ Washington, Dept Lab Med, Seattle, WA 98195 USA
[2] Focus Diagnost Reference Lab, San Juan Capistrano, CA USA
[3] Univ Utah, Dept Pathol, Salt Lake City, UT USA
[4] Fred Hutchinson Canc Res Ctr, Vaccine & Infect Dis Inst, Seattle, WA 98104 USA
[5] MultiCare Hlth Syst, Tacoma, WA USA
[6] Viracor IBT Labs, Los Angeles, CA USA
关键词
DELTA-ANTIGEN; B-VIRUS; INFECTION; ASSAY; HDV;
D O I
10.1128/CVI.00028-16
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Serology remains critical for diagnosing hepatitis D virus (HDV) infection, which affects 15 to 20 million people worldwide, but the literature on characterizing commercial enzyme immunoassays (EIAs) dates back to 15 years ago. We evaluated 2 commercial EIAs currently available for detecting anti-HDV antibodies. The DiaSorin assay demonstrated 100% sensitivity and specificity. Using a modified cutoff value, the Cusabio assay demonstrated a sensitivity of 81.3% and specificity of 90.9%. Our data show that recently developed EIAs are reliable for anti-HDV antibody detection.